Adobe Systems Department of Biophysics, Medical Faculty, Masaryk University, Brno 1 Lectures on Medical Biophysics Ultrasound diagnostics Výsledek obrázku pro 4d ultrasound gif jaterní cysta ren-perf2 Adobe Systems 2 Lecture outline ØPhysical properties of ultrasound and acoustic parameters of medium ØUltrasonography •Impulse reflection method •A-mode – one-dimensional •B-mode – two-dimensional •M-mode •Basic characteristics of US images •Interventional sonography •Echocontrast agents •Harmonic imaging •Principle of 3D imaging ØDoppler flow measurement •Principle of Doppler effect •Principle of blood flow measurement •CW Doppler system •Systems with pulsed wave – PW Doppler •Duplex and Triplex methods •Power Doppler method •Tissue Doppler Imaging (TDI) ØElastography ØUltrasonic densitometry ØPatient Safety: reducing Ultrasound ‘Doses’ Adobe Systems 3 Ultrasound diagnostics ØUltrasound diagnostics started to develop as a clinical method in early 50‘ of 20th century. It allows to obtain cross-sectional images of the human body which can also include substantial information about its physiology and pathology. Ø ØUltrasound diagnostics is based mainly on reflection of ultrasound waves at acoustical interfaces ØWe can distinguish: Ø ̶Ultrasonography (A, B and M mode, 3D and 4D imaging) ̶Doppler flow measurement, including Duplex and Triplex methods (Duplex, Colour Doppler, Triplex, Power Doppler) ̶Tissue Doppler imaging ̶Elastography ̶Ultrasound densitometry 4 Physical properties of ultrasound Before we will deal with diagnostic devices, we need to understand what is ultrasound and what are the main acoustical properties of medium. Ultrasound (US) is mechanical oscillations with frequency above 20 kHz which propagate through an elastic medium. In liquids and gases, US propagates as longitudinal waves. In solids, US propagates also as transversal waves. Adobe Systems 5 Interactions of US with Tissue ØReflection (smooth homogeneous interfaces of size greater than beam width or the US wavelength, e.g. organ outlines) Ø ØRayleigh Scatter (small reflector sizes, e.g. blood cells, dominates in non-homogeneous media) Ø ØRefraction (away from normal from less dense to denser medium, note opposite to light, sometimes produces distortion) Ø ØAbsorption (sound to heat) ̶absorption increases with f, note opposite to X-rays ̶absorption high in lungs, less in bone, least in soft tissue, again note opposite to x-rays ̶ ØInterference: ‘speckles’ in US image result of interference between Rayleigh scattered waves. It is an image artefact. Ø ØDiffraction Adobe Systems 6 Acoustic parameters of medium: interactions Interaction of US with medium – reflection and back-scattering, refraction, attenuation (scattering and absorption) 7 Acoustic parameters of medium Speed of US c depends on elasticity and density r of the medium: K - modulus of compression in water and soft tissues c = 1500 - 1600 m·s-1, in bones about 3600 m·s-1 8 Attenuation of US expresses decrease of plane wave amplitude along its trajectory. It depends on frequency Ix = Io·e-2a·x a = a´·f2 Ix – final intensity, Io – initial intensity, 2x – medium layer thickness (reflected wave travels „to and fro“), a - linear attenuation coefficient (increases with frequency). Since a = log10(I0/IX)/2x we can express a in units dB/cm. At 1 MHz: muscle 1.2, liver 0.5, brain 0.9, connective tissue 2.5, bone 8.0 Acoustic parameters of medium Adobe Systems Zápatí prezentace 9 10 We suppose perpendicular incidence of US on an interface between two media with different Z - a portion of waves will pass through, and a portion will be reflected (the larger the difference in Z, the higher reflection). Acoustic parameters of medium: US reflection and transmission on interfaces Coefficient of reflection R – ratio of acoustic pressures of reflected and incident waves Coefficient of transmission D – ratio of acoustic pressures of transmitted and incident waves Acoustic impedance: product of US speed c and medium density r Z = r·c (Pa·s/m) muscles 1.7×106, liver 1.65×106 brain 1.56×106, bone 6.1×106, water 1.48×106 11 ØNear field (Fresnel area) – this part of US beam is cylindrical – there are big pressure differences in beam axis ØFar field (Fraunhofer area) – US beam is divergent – pressure distribution is more homogeneous ØIncrease of frequency of US or smaller probe diameter cause shortening of near field - divergence of far field increases near-far field Acoustic parameters of medium: Near field and far field (optional) Theory and real look Axis of the beam Main lobe Side lobes Transducer centre Adobe Systems 12 Ultrasonography Passive US – low intensity waves which cannot cause substantial changes of medium. In US diagnostics (ultrasonography = sonography = echography) - frequencies used are 2 - 40 MHz with (temporal average, spatial peak) intensity of about 1 kW/m2 Impulse reflection method: a probe with one transducer which is source as well as detector of US impulses. A portion of emitted US energy is reflected on the acoustic interfaces and the same probe then receives reflected signal. After processing, the signal is displayed on a screen. Adobe Systems 13 fkjBez názvu1 Ultrasonography: Impulse reflection method 14 Ultrasonography: Impulse reflection method Main parts of the US apparatus: Common to diagnostics and therapy Øprobe with electroacoustic transducer (transducers) Øgenerator of electric oscillations (continuous, pulsed) Special parts of diagnostic apparatus Øelectronic circuits for processing of reflected signal (today A/D converter and respective software) Ødisplay unit Ørecording unit Ø 15 Ultrasonography: A-mode – one-dimensional ØDistances between reflecting interfaces and the probe are shown. ØReflections from individual interfaces (boundaries of media with different acoustic impedances) are represented by vertical deflections of base line, i.e. the echoes. Echo amplitude is proportional to the intensity of reflected waves (Amplitude modulation) Distance between echoes shown on the screen is approx. proportional to real distance between tissue interfaces. Today used mainly in ophthalmology. 16 Ultrasonography: A-mode – one-dimensional echofig1 A-scan Výsledek obrázku pro ultrazvuk v oftalmologii Contact gel Adobe Systems Zápatí prezentace 17 18 Ultrasonography: M-mode One-dimensional static B-scan shows movement of reflecting tissues. The second dimension is time in this method. Static probe detects reflections from moving structures. The echoes are represented by points moving vertically on the screen, horizontal shifting of the record is given by slowly running time-base. Displayed curves represent movement of tissue structures plice chest wall lungs Adobe Systems 19 Repetitive formation of B-mode images of examined area by fast deflection of US beam mechanically (in the past) or electronically „in real time“ today. Electronic probes consist of many piezoelectric transducers which are gradually activated. Ultrasonography: B-mode - dynamic Adobe Systems 20 sondaUZ Ultrasound probes for dynamic B-mode: electronic and mechanical (history), sector and linear. Ultrasonography: B-mode - dynamic Abdominal cavity is often examined by convex probe – a combination of a sector and linear probe. Adobe Systems 21 Modern ultrasonography - digital processing of image Ø Analogue part – detection system Ø Analogue-digital converters (ADC) Ø Digital processing of signal – possibility of programming (preprocessing, postprocesssing), image storage (CD, flash cards etc.) Ultrasonography: B-mode - dynamic sampling 22 Ultrasonography B-mode - dynamic Výsledek obrázku pro echokardiografie Adobe Systems 23 Degree of reflectivity – echogenicity. The images of cystic (liquid-filled) and solid structures are different. According to the intensity of reflection in the tissue bulk we can distinguish structures: hyperechogenic, izoechogenic, hypoechogenic, anechogenic. ̶ ØSolid structures – acoustic shadow (caused by absorption and reflection of US) Ø ØAir bubbles and other strongly reflecting interfaces cause repeating reflections (reverberation, „comet tail“). Ultrasonography: Basic characteristics of US images Adobe Systems 24 Acoustic shadow caused by absorption and reflection of US by a kidney stone (arrow) Hyperechogenic area below a cyst (low attenuation of US during passage through the cyst compared with the surrounding tissues – arrow) Ultrasonography 25 Limitation! – absorption of US increases with frequency of ultrasound = smaller penetration depth Compromise frequency 3-5 MHz – penetration in depth of about 20 cm penetration-resolution Ultrasonography Spatial resolution of US imaging system is determined by the wavelength of the US. When the object dimension is smaller than this wavelength only scattering occurs. Hence higher spatial resolution requires higher frequencies Adobe Systems 26 ØAxial spatial resolution - it is given by the shortest distance of two distinguishable structures lying in the beam axis – it depends mainly on frequency (at 3.5 MHz about 0.5 mm) ØLateral spatial resolution - it is given by the shortest distance of two distinguishable structures perpendicularly to the beam axis – depends on the beam width ØElevation – ability to distinguish two planes (sections) lying behind or in front of the depicted tomographic plane – it depends on frequency and beam geometry Ultrasonography: Spatial Resolution kinds of resolution Optional! Adobe Systems 27 The best resolving power can be found in the narrowest part of the US beam profile. Focusing – US beam is converged at the examined structure by means of acoustic lenses (shapes of the layer covering the transducer) or electronically. ØThe probes can be universal or specially designed for different purposes with different focuses. ØThe position of focus can be changed in most sector probes). Ultrasonography: Spatial Resolution Optional! Adobe Systems 28 Ultrasonography: Interventional sonography ØInterventional sonography is used mainly for guiding punctures Ødiagnostic – thin needle punctures to take tissue samples for histology Øtherapeutic – for aspiration of a cyst or an abscess content or an exudate etc. Ø ØPuncture can be done by „free hand“ – the probe is next to the puncture site – or the puncture needle is guided by a special probe attachment. https://theultrasoundsite.co.uk/region-specific-ultrasound-guided-injections/ultrasound-guided-inje ctions-wrist-hand/ Adobe Systems 29 Ultrasonography: Echocontrast agents - increase echogenicity of streaming blood Gas microbubbles (mainly air or volatile hydrocarbons) - Free or enclosed in a biopolymer envelope A SEM micrograph of encapsulated echocontrast agent Adobe Systems 30 Ultrasonography: Echocontrast agents - application FNH 18s Echocontrast image of Focal Nodular Hyperplasia of the Liver 18 s after intravenous application of the echocontrast agent CEUS+ CEUS+ technology uses the unique properties of ultrasound contrast agents.  When stimulated with low MI frequencies, the oscillating micro bubbles reflect both basic frequencies and harmonic signals. Enhancement of contrast by microbubbles in liver and kidney http://www.medicalimagingtech.com/us-rs80a 31 An impulse with basic frequency f0 is emitted into the tissue. The receiver, however, does not detect the reflected US with this same frequency but with the second harmonic frequency 2f0. Its source is tissue itself (advantage in patients „difficult to examine“). The method is also used with echocontrast agents – source of the second harmonic are oscillating bubbles. Advantageous when displaying blood supply of some lesions. Conventional (left) and harmonic (right) images of a kidney with a stone. Ultrasonography: Harmonic imaging harmzobr Adobe Systems 32 Ultrasonography: Panoramic imaging * Purpose of this method is a continuous image record of a tissue or organ in desired plane (direction). Panoramic image enables assessment of dimensions and morphology of the whole-body portion. * This method is a supplement to the conventional imaging. játra, žlučník a pravá ledvina DF Panoramic image of epigastrium From left right kidney, right liver lobe, gallbladder, left liver lobe, spleen 33 - The probe is linearly shifted, tilted or rotated. The data about reflected signals in individual planes are stored in memory of a powerful PC which consequently performs mathematical reconstruction of the image. Disadvantages of some 3D imaging systems: relatively long time needed for mathematical processing, price. Ultrasonography Principle of three-dimensional (3D) imaging http://i01.i.aliimg.com/img/pb/363/457/469/469457363_664.jpg Adobe Systems 34 Four-dimensional (4D) image The fourth dimension is time Výsledek obrázku pro 4d ultrasound gif https://giphy.com/gifs/4d-14jU1PuglaN1v2 35 Doppler flow measurement The Doppler effect (frequency shift of waves formed or reflected at a moving object) can be used for detection and measurement of blood flow, as well as, for detection and measurement of movements of some acoustical interfaces inside the body (heart, blood vessel walls) Perceived frequency corresponds with source frequency in rest. Perceived frequency is higher when approaching it. Perceived frequency is lower when moving away it. DOPPLER Christian. A. Doppler (1803-1853), Austrian physicist and mathematician, formulated his theory in 1842 during his stay in Prague. 36 Application of Doppler effect in blood flow velocity measurement Moving reflector (back scatterer) = erythrocytes Doppler effect II Doppler flow measurement: Principle of Doppler effect 37 US Doppler blood flow-meters are based on the difference between the frequency of ultrasound (US) waves emitted by the probe and those reflected (back-scattered) by moving erythrocytes. The frequency of reflected waves is (in comparison with the emitted waves) higher in forward blood flow (towards the probe) lower in back blood flow (away from the probe) The difference between the frequencies of emitted and reflected US waves is proportional to blood flow velocity. Doppler flow measurement: Principle of blood flow measurement 38 Doppler flow measurement: principle of blood flow measurement angle alpha 39 1)Calculation of Doppler frequency change fd 2)Calculation of „reflector“ (erythrocytes) velocity v 1) 2) fv - frequency of emitted US waves α - angle made by axis of emitted US beam and the velocity vector of the reflector c – US speed in the given medium (about 1540 m/s in blood) Doppler flow measurement overestimation Dependence of velocity overestimation on the incidence angle α (if the device is adjusted for a = 0, i.e. cosa = 1) a - angle made by axis of emitted US beam and the velocity vector of the reflector 40 1)Systems with continuous wave – CW. They are used for measurement on superficial blood vessels. High velocities of flow can be measured, but without depth resolution. Used only occasionally. 2) 2)Systems with pulsed wave. It is possible to measure blood flow with accurate depth localisation. Measurement of high velocities in depths is limited. CW and PW method Doppler flow measurement 41 The probe has only one transducer which acts alternately as emitter and receiver. The measurement of velocity and direction of blood flow in the vessel is evaluated in the so-called sampling volume with adjustable size and depth. The pulse duration defines the size of the sampling volume (this volume should involve the whole diameter of the examined blood vessel). Doppler flow measurement Systems with pulsed wave - PW 42 Aliasing – artefact of measurement. At high repetition frequency of pulses the upper part of the spectral curve can appear in negative velocity range. - at velocity above 4m/s aliasing cannot be removed Nyquist limit Doppler methods: Pulse wave (PW) systems – optional! Adobe Systems 43 DUPLEX method is a combination of dynamic B-mode imaging (the morphology of examined area with blood vessels is depicted) and the PW Doppler system (measurement of velocity spectrum of blood flow). It allows to examine blood flow inside heart or in deep blood vessels (flow velocity, direction and character) Doppler methods Adobe Systems 44 Doppler flow measurement Basic spectral curves directional impedance OBR37 Low peripheral impedance: brain arteries, arteries in parenchymatous organs High peripheral impedance: arteries in skeletal muscles syst. = systolic Forward flow Back flow 45 figure2 Doppler methods DUPLEX method Placement of sampling volume (left) and the record of blood flow velocity spectrum in stenotic a. carotis communis (right) Adobe Systems 46 Doppler methods Colour Doppler imaging The image consists of black-white and colour part. The black-white part contains information about reflectivity and structure of tissues. The colour part informs about movements in the examined section. (The colour is derived from average velocity of flow.) The apparatus depicts distribution and direction of flowing blood as a two-dimensional image. BART rule – blue away, red towards. The flow away from the probe is coded by blue colour, the flow towards the probe is coded by red colour. The brightness is proportional to the velocity, turbulences are depicted by green patterns. Adobe Systems 47 Doppler flow measurement TRIPLEX method Combination of duplex method and colour Doppler imaging Normal blood flow in a. carotis communis (left) and in a. renalis dx (right) ACI obr Adobe Systems 48 Doppler methods TRIPLEX method stenosis of a. carotis Cartoid Artery Stenosis Adobe Systems 49 Doppler methods Power Doppler method - the whole energy of the Doppler signal is utilised - mere detection of blood flow only little depends on the so-called Doppler incidence angle - imaging of even very slow flows (blood perfusion of tissues and organs) - flow direction is not shown Power Doppler of Carotid Bifurcation velocity distribution obr Renal perfusion Carotid bifurcation Adobe Systems 50 Tissue Doppler Imaging (TDI) Colour coding of information about velocity and direction of movements of tissues Velocities 1-10 mm/s are depicted. TDI of a. carotis communis during systole Elastography Elastography is an imaging modality analogous to palpation. Basis: pathological changes of tissues can manifest themselves like changed mechanical properties, e.g. rigidity. The tumours are more rigid than healthy tissues in most cases. This method allows to visualise inner structures of some tissues based on the measurement of the response to the tissue compression from the body surface. This response depends, among others, on the microscopic and macroscopic structure of the tissues. Moreover, the tissues have also the so called viscoelastic and poroelastic properties (poroelasticity is a specific elasticity of porous materials which pores are filled by a liquid). 51 Elastography Elastic properties of tissues cannot be evaluated based on a simple sonogram. Hence several ultrasonic elastographic methods were developed: Strain-Stress Elastography where the tissue deformation is caused by the pressure exerted by the ultrasound probe. Acoustic Radiation Forced Impulse Elastography where the pressure is exerted by a strong impulse of radiation force. At present time, the SWE – Shear Wave Elastography is dominant. In this method, instead of the pressure action of the probe, the radiation force of the ultrasound waves is exploited – see the picture. The compression is done by means of relatively long repeated focussed impulses along the imaging axis, which produce the shear (transverse) waves. These waves propagate much slower than the longitudinal waves which speed is proportional to the tissue's elasticity (Young modulus). The particles of medium move (oscillate) with the amplitude of some micrometres, and to visualise this movement is needed a special imaging mode denoted as supersonic imaging – it encompasses ultrafast image processing (5000–20000 frames/s). In comparison with the previous method, the information on tissue elasticity is quantitative, the colour scale is calibrated in kPa. 52 Elastography 53 SWE elastography of a phantom with two areas of different elasticity. In the upper part we can see an elastogram, in the lower part a grayscale ultrasonogram. Adobe Systems 54 Ultrasonic densitometry It is based on both the measurement of speed of ultrasound in bone and the estimation of ultrasound attenuation in bone. In contrast to X-ray methods, ultrasound densitometry also provides information on the structure of bone and its elastic properties. ØThe speed of ultrasound depends on the density and elasticity of the measured medium. The anterior area of the tibia and the posterior area of the calcaneus are frequently used as places of measurement. The speed of ultrasound is given by the quotient of measured distance and the transmission time. Ø ØUltrasound attenuation depends on the physical properties of the given medium and the frequency of the ultrasound applied. For the frequency range 0.1 - 1 MHz the frequency dependence is nearly linear. Attenuation is currently expressed in dB/MHz/cm. ØClinical importance: diagnostics of osteoporosis Adobe Systems 55 Ultrasonic densitometry Ultrasound measurements used to assess bone density at the calcaneus densitometr Adobe Systems Patient Safety: reducing Ultrasound ‘Doses’ (see also the lecture on ultrasound cavitation) Adobe Systems 57 Prudent use of Ultrasound ØUS is non-ionising BUT since many bioeffects of ultrasound have not yet been studied fully, ‘prudent’ use is recommended ØALARA – as low as reasonably achievable (exposure) ØIn practice ‘prudent’ = justification + optimisation Adobe Systems 58 Biological Effects ØPossible bioeffects: inactivation of enzymes, altered cell morphology, internal haemorrhage, free radical formation … ØMechanisms of bioeffects: ̶Mechanical effects Displacement and acceleration of biomolecules Gas bubble cavitation (stable and transient) – see the lecture on biological effects of ultrasound ̶Elevated tissue temperatures (absorption of ultrasound and therefore increase in temperature high in lungs, less in bone, least in soft tissue) ØBioeffects are deterministic with a threshold (cavitation) or without it (heating). Adobe Systems 59 Output Power from Transducer Øvaries from one machine to another Ø ØIncreases as one moves from real-time imaging to colour flow Doppler Ø ØM-mode output intensity is low but dose to tissue is high because beam is stationary Adobe Systems 60 Risk Indicators ØTo avoid potentially dangerous exposures, two indices were introduced. Their values (different for different organs) are often displayed on device screens and should not be exceeded. ØThermal Index (TI): TI = possible tissue temperature rise if transducer is kept stationary ̶TIS: soft tissue path ̶TIB: bone near focus of beam ̶TIC: Cranium (near surface bone) ØMechanical Index (MI): measure of possible mechanical bioeffects Adobe Systems 61 More on the TI and MI Thermal index – device power divided by the power that would increased the temperature by one degree under conditions of minimum heat loss (without perfusion). Mechanical index (for assessment of cavitation-conditioned risk, increased danger when using echocontrast agents): Justification ØNo commercial demos on human subjects ØNo training on students ØNo ‘see baby just for fun’ or excessive screening in obstetrics Adobe Systems 62 Optimisation of ‘Dose’ ØMinimise TI and MI and use appropriate index (TIS, TIB, TIC), care in cases when these underestimated Ø ØCheck acoustic power outputs on manual Ø ØUse high receiver gain when possible as opposed to high transmit power Ø ØStart scan with low transmit power and increase gradually Ø ØAvoid repeat scans and reduce exposure time Ø ØDo not hold transducer stationary Ø ØGreater care when using contrast agents as these increase the possibility of cavitation Ø ØExceptional care must be taken in applying pulsed Doppler in obstetrics Ø ØRegular quality control of the ultrasound device Adobe Systems Authors: Vojtěch Mornstein, Ivo Hrazdira, Pavel Grec Content collaboration and language revision: Carmel J. Caruana Graphical design: Lucie Mornsteinová Last revision: September 2024 http://www.freehotgame.com Bad%20Hair%20Day