Cell Density Monitoring System Operating Instructions 3Cell Density Monitoring System Operating Instructions TABLE OF CONTENTS Hamilton Warranty Please refer to the General Terms of Sales (GTS). Important note Copyright © 2021 Hamilton Bonaduz AG, Bonaduz Switzerland. All rights reserved. The reproduction of any part of this document in any form is forbidden without the express written agreement of Hamilton Bonaduz AG. Contents of this operating instruction can be modified without previous announcement. Technical modifications reserved. Greatest possible care was used on the correctness of the information in this operating instruction. If errors should be discovered nevertheless, Hamilton Bonaduz AG is pleased to be informed about it. Regardless of this, Hamilton Bonaduz AG cannot assume liability for any errors in this operating instruction or for their consequences. Table of Contents 1 Preface..................................................................................................................10 2 General Information.............................................................................................10 2.1 Intended Use.............................................................................................................. 10 2.2 About this Operating Instruction................................................................................ 10 3 Liability.................................................................................................................. 11 4 Safety Precautions and Hazards........................................................................ 11 4.1 General Precautions....................................................................................................11 4.2 Operation of the Cell Density Monitoring System....................................................... 12 4.3 Electrical Safety Precautions...................................................................................... 12 4.4 Chemical, Radioactive or Biological Hazard Precautions.......................................... 14 5 Product Description............................................................................................. 14 5.1 Why Measuring Cell Density?.................................................................................... 14 5.2 The Theory of Permittivity Measurement (Incyte)....................................................... 15 5.2.1 The Dual-Frequency Measurement Mode.................................................... 15 5.2.2 The Frequency Scan Mode........................................................................... 16 5.3 The Theory of Turbidity Measurement........................................................................17 5.4 Hardware Description................................................................................................ 18 5.4.1 Control Unit................................................................................................... 20 5.4.1.1 Arc View Controller..................................................................... 20 5.4.1.2 ComBox..................................................................................... 21 5.4.1.3 Cell Density Integration Kit......................................................... 22 5.4.2 Incyte Sensor Unit......................................................................................... 23 5.4.3 Dencytee Sensor Unit................................................................................... 24 5.5 Outline of the Control Unit Software........................................................................... 25 5.5.1 Run Menu..................................................................................................... 25 5.5.2 Sensor Channel Tab...................................................................................... 28 5.5.3 Main Menu.................................................................................................... 28 5.5.4 Unit Settings Menu........................................................................................ 30 5Cell Density Monitoring System Operating Instructions 5.5.5 Measure Settings Menu................................................................................ 31 5.5.6 Incyte Frequency Scan Settings Menu......................................................... 32 5.5.7 Culture Settings Menu.................................................................................. 33 6 Installation............................................................................................................34 6.1 Install the Control Unit................................................................................................ 34 6.1.1 Install the Arc View Controller........................................................................ 34 6.1.2 Install the ComBox........................................................................................ 35 6.1.3 Install the Cell Density Integration Kit............................................................ 36 6.2 Mount the Sensor in the Bioreactor........................................................................... 37 6.3 Connect the Sensor Unit to the Control Unit.............................................................. 40 7 Operation of Incyte..............................................................................................42 7.1 Choose the Measurement Mode............................................................................... 42 7.2 Addition of Arc Sensors............................................................................................. 42 7.3 Start Recording Measurements in a Culture File ....................................................... 44 7.4 Perform a Product Calibration and Set the Inoculation Time..................................... 45 7.5 Add a Comment to the Culture File............................................................................ 45 7.6 Customize the Main Graph........................................................................................ 46 7.7 Setup a Multi View Graph.......................................................................................... 47 7.8 Stop Recording the Culture File ................................................................................ 48 7.9 Dismount the Sensor Unit ......................................................................................... 48 7.10 Export a Culture File................................................................................................... 49 7.11 Import a Culture File in an Excel Spreadsheet........................................................... 50 7.12 Further Measurement Settings.................................................................................. 51 7.12.1 Set the Cell Factor...................................................................................... 51 7.12.2 Start the Incyte Scan................................................................................. 51 7.12.3 Define the Custom Measurement Mode.................................................... 52 7.12.4 Perform a Manual Product Calibration....................................................... 53 7.13 Use Sensor Cleaning (Probe Cleaning)...................................................................... 53 8 Operation of Dencytee.........................................................................................55 8.1 Choose the Measurement Mode............................................................................... 55 8.2 Addition of Arc Sensors............................................................................................. 55 8.3 Start Recording Measurements in a Culture File........................................................ 58 8.4 Perform Product Calibration and Set the Inoculation Time........................................ 59 8.5 Add a Comment to the Culture File............................................................................ 59 TABLE OF CONTENTS 8.6 Customize the Main Graph ....................................................................................... 60 8.7 Setup a Multi View Graph.......................................................................................... 61 8.8 Stop Recording the Culture File ................................................................................ 62 8.9 Dismount the Sensor Unit ......................................................................................... 62 8.10 Export a Culture File................................................................................................... 62 8.11 Import the Culture File in an Excel Spreadsheet........................................................ 63 8.12 Define the Custom Measurement Mode.................................................................... 64 9 Administrator Menu.............................................................................................65 9.1 Definition of the User Level......................................................................................... 66 9.1.1 Change the Administrator Password............................................................ 66 9.1.2 Define the User Level.................................................................................... 67 9.2 Adjust the General Settings....................................................................................... 67 9.2.1 Touch Screen Calibration.............................................................................. 68 9.2.2 Adjustment of the Decimal Separator in the Culture File............................... 68 9.2.3 Continuous Display of Graphed Data............................................................ 68 9.2.4 Adjustment of Date and Time Settings.......................................................... 68 9.2.5 Initialize the Arc View M12 Connectors......................................................... 69 9.2.6 Enable the Cell Density Integration Kit Option............................................... 69 9.3 Read the Network Settings........................................................................................ 70 9.4 Adjust the Network Applications................................................................................ 71 9.4.1 Activate Share Data Files............................................................................... 72 9.4.2 Activate OPC Communication...................................................................... 72 9.5 Install a License.......................................................................................................... 72 9.6 Transfer Administrator Files........................................................................................ 73 9.6.1 Export the Log Files.......................................................................................74 9.6.2 Erase Temporary Files....................................................................................74 9.6.3 Export Culture Data.......................................................................................74 9.7 Update or Re-Set the Arc View Controller or Cell Density Integration Kit.................. 75 9.7.1 Software Update of the Control Unit............................................................. 75 9.7.1.1 Software Update of the Arc View Controller or Cell Density Integration Kit......................................................... 75 9.7.1.2 Software installation/update of the ComBox.............................. 76 9.7.1.2.1 Procedure for one ComBox per PC.................... 76 9.7.1.2.2 Procedure for two ComBoxes per PC................ 76 9.7.2 Re-Set the Arc View Controller or Cell Density Integration Kit..................... 77 9.7.3 Update Recovery of the Arc View Controller................................................. 77 TABLE OF CONTENTS 7Cell Density Monitoring System Operating Instructions 10 Maintenance.........................................................................................................78 10.1 Daily Maintenance...................................................................................................... 78 10.2 Weekly Maintenance.................................................................................................. 78 10.3 Manual Cleaning Procedure at Process End for Cell Density Sensors...................... 79 10.3.1 Incyte Sensors........................................................................................... 79 10.3.2 Dencytee Sensors...................................................................................... 79 10.4 Yearly Maintenance.................................................................................................... 79 10.5 Sensor Unit Verification.............................................................................................. 80 10.5.1 Verification of an Incyte Sensor Unit........................................................... 80 10.5.2 Verification of a Dencytee Sensor Unit....................................................... 81 10.6 Calibration of a Sensor Unit....................................................................................... 82 10.6.1 Calibrate an Incyte Sensor Unit.................................................................. 83 10.6.2 Calibrate a Dencytee Sensor Unit.............................................................. 86 10.7 Transfer a Calibration Files to the Arc View Controller................................................ 86 10.7.1 Clean and Back Up the Arc View Controller.............................................. 87 10.7.2 Update the Control Unit............................................................................. 87 11 Troubleshooting....................................................................................................88 11.1 Self-Diagnostic of the Sensor Unit............................................................................. 88 11.2 Re-Set the Factory Settings....................................................................................... 89 11.3 Verification and Calibration Error................................................................................ 89 11.3.1 Verification Error......................................................................................... 89 11.3.2 Calibration Error......................................................................................... 89 11.4 Getting Technical Support.........................................................................................90 11.4.1 Create Print Screens..................................................................................90 11.4.2 System Information....................................................................................90 11.4.3 System Data............................................................................................... 91 11.5 Remove Corrupt Files................................................................................................ 91 11.6 Transfer Error-Log-Files from the Arc View Controller................................................ 92 11.7 Returning a System for Repair................................................................................... 93 12 Connection of the Control Unit to the Process Control System......................94 12.1 Connect by Modbus communication......................................................................... 94 12.1.1 Preparation of the Control Unit................................................................... 94 12.1.2 Preparation of the Modbus Connection..................................................... 95 12.1.2.1 Definition of the Return Time...................................................... 95 12.1.2.2 Overview of the Available Data................................................... 96 TABLE OF CONTENTSTABLE OF CONTENTS 12.1.2.3 Definition of 32 Bit Data Type..................................................... 97 12.1.2.4 Definition of Modbus Exceptions................................................ 98 12.1.2.5 Pre-Amplifier Status Codes........................................................ 98 12.1.2.6 Calibration and Validation via Modbus..................................... 100 12.1.2.6.1 General Concepts............................................. 100 12.1.2.6.1.1 Controlling the State Machine via Modbus....... 100 12.1.2.6.1.2 The main state machine.................................... 102 12.1.2.6.1.3 Modbus registers for CDIK, Arc View Controller 265 and PC software....................... 103 12.1.2.6.1.4 Modbus registers for Arc View Controller 465.. 104 12.1.2.6.1.5 General error states.......................................... 105 12.1.2.6.2 Validation Incyte Unit......................................... 106 12.1.2.6.2.1 Table of states................................................... 107 12.1.2.6.2.2 Error states....................................................... 107 12.1.2.6.3 Calibration Incyte Unit....................................... 108 12.1.2.6.3.1 Table of states................................................... 109 12.1.2.6.3.2 Dilution steps for different calibration types.......110 12.1.2.6.3.3 Error states........................................................111 12.1.2.6.4 Validation Dencytee...........................................112 12.1.2.6.4.1 Table of states....................................................113 12.1.2.6.4.2 Error states........................................................113 12.1.2.6.5 Calibration Dencytee..........................................114 12.1.2.6.5.1 Table of states....................................................115 12.1.2.6.5.2 Error states........................................................115 12.1.2.7 Modbus Tables Arc View Controller 265, ComBox and Cell Density Integration Kit........................................................116 12.1.2.7.1 Input Registers Tables (Read only).....................116 12.1.2.7.1.1 Read the Modbus Commands..........................116 12.1.2.7.1.2 Incyte Input Registers Table...............................116 12.1.2.7.1.3 Dencytee Input Registers Table........................ 122 12.1.2.7.2 Holding Registers (read and write).................... 123 12.1.2.7.2.1 Read and Write Function Code ........................ 123 12.1.2.7.2.2 Incyte Holding Registers Table...........................124 12.1.2.7.2.3 Dencytee Holding Registers Table.................... 125 12.1.2.7.2.4 Allowed Values for Holding Registers................127 12.1.2.8 Modbus Tables Arc View Controller 465 (XL) ...........................127 12.1.2.8.1 Input Registers Tables (Read only).....................127 9Cell Density Monitoring System Operating Instructions 12.1.2.8.1.1 Read the Modbus Commands..........................127 12.1.2.8.1.2 Incyte Input Registers Table.............................. 128 12.1.2.8.1.3 Dencytee Input Registers Table........................ 139 12.1.2.8.2 Holding Registers (read and write) ....................142 12.1.2.8.2.1 Read and Write Function Code .........................142 12.1.2.8.2.2 Incyte Holding Registers Table.......................... 143 12.1.2.8.2.3 Dencytee Holding Registers Table.................... 145 12.1.2.8.2.4 Allowed Values for Holding Registers................147 12.2 Connect by OPC Communication........................................................................... 148 12.2.1 Introduction to OPC................................................................................. 148 12.2.2 Configuration of the Control Unit.............................................................. 148 12.2.3 Configuration of the OPC Client............................................................... 150 12.2.4 OPC Classic Demo Client Start Guide..................................................... 150 12.2.4.1 System Overview..................................................................... 150 12.2.4.2 Download and Installation.........................................................151 12.2.4.3 Connecting the OPC Classic Demo Client to the Control Unit....151 12.2.5 Softing OPC Bridge Start Guide.............................................................. 155 12.2.5.1 System Overview..................................................................... 155 12.2.5.2 Download and Installation........................................................ 155 12.3 Connecting the 4-20 mA Output Box...................................................................... 163 12.3.1 Electrical Installation................................................................................. 163 12.3.2 Software Configuration............................................................................ 166 12.4 Connect by Profibus Communication...................................................................... 167 12.4.1 Profibus General Information.................................................................... 167 12.4.1.1 Ground the Modbus-Profibus module..................................... 168 12.4.2 Profibus Interface..................................................................................... 168 12.4.2.1 Overview.................................................................................. 168 12.4.2.2 Input Buffer.............................................................................. 169 12.4.2.3 Output Buffer........................................................................... 169 12.4.2.4 Decoding the Return Values......................................................172 12.4.2.5 Function codes..........................................................................174 12.4.3 Modbus Interface..................................................................................... 182 12.4.3.1 Configuration............................................................................ 182 13 Disposal..............................................................................................................182 14 Software Versions..............................................................................................183 15 Hardware Compatibility.....................................................................................184 16 Ordering Information..........................................................................................184 16.1 Controller................................................................................................................. 184 16.2 Cell Density Integration Kit....................................................................................... 185 16.3 Units......................................................................................................................... 185 16.4 Spare Parts and Accessories................................................................................... 186 16.5 Calibration Tools....................................................................................................... 187 17 Glossary..............................................................................................................187 18 FAQ......................................................................................................................188 18.1 General Questions................................................................................................... 188 18.2 Incyte Questions...................................................................................................... 190 TABLE OF CONTENTSTABLE OF CONTENTS 11Cell Density Monitoring System Operating Instructions LIABILITY SAFETY PRECAUTIONS AND HAZARDS 1 Preface Welcome to the World of Hamilton Precision Instruments. Congratulations on your purchase of a Hamilton Cell Density Monitoring System comprising a Control Unit (Arc View Controller, Cell Density ComBox, Cell Density Integration Kit) and an Incyte and/or Dencytee Unit. The sensors are designed for monitoring permittivity (Incyte) and turbidity (Dencytee). This technology enables the online monitoring of viable and total cell density during biotechnological processes. For the first time you can monitor cells in real-time, obtain actionable data, and automate process adjustment. A standard measuring loop consists of a Sensor Unit (i.e. a sensor with a Pre-amplifier), which is connected to the Control Unit. With the Control Unit, it is also possible to view Arc sensors wirelessly, like pH and DO, on just one display. Proper handling and maintenance of this monitoring technology will increase the lifespan of the system. To learn about proper care and maintenance, please take the time to read this manual, including the warranty information. 2 General Information 2.1 Intended Use The Cell Density Monitoring System is designed to measure permittivity with the Incyte sensor and turbidity with the Dencytee sensor in a liquid medium. These measurements may be used for the control of bioprocesses within the defined specifications (see specifications sheets www.hamiltoncompany.com). The permittivity measurement may be correlated to the viable cell density, and the turbidity measurement may be correlated to the total cell density. In addition to permittivity Incyte also measures conductivity. This measurement should not be used for the control of bioprocesses. 2.2 About this Operating Instruction These Operating Instructions will help users to operate the Cell Density Monitoring System correctly and safely. To achieve that goal, this document describes the different components of the system and how they function. The Operating Instructions describe both the hardware and software of the Cell Density Monitoring System in a depth enabling the user to operate the system. After introducing the various parts, it is shown step by step how to operate the system. After reading the Operating Instructions, users should be capable of installing and operating the Cell Density Monitoring System. Following information are highlighted within this document: PREFACE • GENERAL INFORMATION ATTENTION! Essential information for avoiding personal injury or damage to the equipment. NOTE: Important instructions or interesting information. 3 Liability The liability of Hamilton Bonaduz AG is detailed in the document “General Terms and Conditions of Sale and Delivery,” Hamilton is expressly not liable for direct or indirect losses arising from the use of the sensors. It must in particular be insured in this conjunction that malfunctions can occur on account of the inherently limited useful life of sensors contingent upon their relevant applications. The user is responsible for the calibration, maintenance and regular replacement of the sensors. In the case of critical sensor applications, Hamilton recommends using back-up measuring points in order to avoid consequential damages. The user is responsible for taking suitable precautions in the event of a product failure. The sensor is not intended as a safety device. 4 Safety Precautions and Hazards ATTENTION! Read the following safety instructions carefully before installing and operating the Cell Density Monitoring System. 4.1 General Precautions For safe and correct use of Incyte and Dencytee, it is essential that both operating and service personnel follow generally accepted safety procedures as well as the safety instructions given in this document, the operating instruction of the Cell Density Monitoring System. The specification given (see specifications sheets www.hamiltoncompany.com) as regards temperature; pressure etc. may under no circumstances be exceeded. Inappropriate use or misuse can be dangerous. Cleaning, assembly and maintenance should be performed by personnel trained in such work. 13Cell Density Monitoring System Operating Instructions Before removing the sensor from the measuring setup, always make sure that no process medium can be accidentally spilled. When removing and cleaning the sensor, it is recommended to wear safety goggles and protective gloves. If the system cannot be repaired by the operator, it has to be sent back to Hamilton for inspection. Necessary precautions should be taken when transporting the sensors. For repair or shipment the System should be sent back in the original reusable packaging box. Every Incyte or Dencytee sensor sent back for repair must be decontaminated. If the conditions described in these operating instructions are not adhered to or if there is any inappropriate interference with the equipment, all of our manufacturer’s warranties become obsolete. 4.2 Operation of the Cell Density Monitoring System The Cell Density Monitoring System must be used for the intended applications, and in optimum safety and operational conditions. The specifications (such as temperature or pressure) defined in the section entitled ‘Technical Specification’ (Chapter 15) must not be exceeded under any circumstances. Make sure that the process connections and O-rings are not damaged when screwing a sensor into the process. O-rings are consumable parts which must be exchanged regularly (at least once per year). Even when all required safety measures have been complied with, potential risks still exist with respect to leaks or mechanical damage to the armature. Wherever there are seals or screws, gases or liquids may leak out undetected. Always make sure that no process medium can be accidentally spilled before removing the sensor from its measurement set up. Do not put stress on the system by vibration, bending or torsion. You may leave the pre-amplifier connected to the sensor during cleaning in place (CIP) and sterilization in place (SIP). Stand clear of the pre-amplifier during CIP and SIP procedures as it may become very hot. The pre-amplifier is not autoclavable. Dismount it prior to autoclavation to avoid damages to the electronics. 4.3 Electrical Safety Precautions Only use the power supply provided with the Arc View Controller, or the ComBox. Do not connect it to a power source of any voltage beyond the range stated in the specifications. Failure to do so may lead to malfunction or damage of the system or impair user safety. ATTENTION! Disconnect from power the Arc View Controller and ComBox prior to mounting and dismounting a Sensor Unit. In areas having significant electronic noise, Incyte sensors and preamps may require grounding. Locations having electronic noise may cause interference to your permittivity readings. Grounding the Incyte sensor and preamp should alleviate the interference. • If the headplate of a benchtop bioreactor is not grounded, connect an earth ground wire to it. • For large metallic bioreactors it is not necessary to ground the Incyte sensor or preamp if the bioreactor is already grounded. • For large bioreactors without a ground (e.g. glass lined reactors) connect a ground wire to the fitting the Incyte sensor is mounted to. ATTENTION! Only use the Hamilton M 12 / M 12 (5 pole) cable to connect Control and Sensor Unit. Connect the headplate to ground. The sensor and preamp will be grounded via its stainless steel PG13.5 thread. The stainless steel PG13.5 thread connects the sensor and preamp to ground. Connect the fitting to ground. The sensor and preamp will be grounded via its stainless steel PG13.5 thread. The Tank is grounded. The Tank is not grounded. Figure 1: Grounding SAFETY PRECAUTIONS AND HAZARDSSAFETY PRECAUTIONS AND HAZARDS 15Cell Density Monitoring System Operating Instructions 4.4 Chemical, Radioactive or Biological Hazard Precautions Selection of the appropriate biological safety level and implementation of the required biosafety measures for working with the Cell Density Monitoring System is the sole responsibility of the user. If working with hazardous liquids observe and carry out the maintenance procedures, paying attention to cleaning and decontamination. If parts or the complete Cell Density Monitoring System become contaminated with biohazardous, radioactive or chemical material, it should be cleaned. Failure to observe and carry out the maintenance procedures may impair the reliability and correction functioning of the system. Avoid damaging the power cord. Do not bend it excessively, step on it, or place heavy objects on it. A damaged cord can easily become a shock or fire hazard. Never use a power cord after it has become damaged. 5 Product Description 5.1 Why Measuring Cell Density? Continuous monitoring of bioprocesses is required for both process control and optimization. The control of the environmental conditions including pH or dissolved oxygen is well established but does not provide information on the cell physiology. Parameters relating to cell physiology are usually monitored offline after daily sampling of the culture. This method is time-consuming and provides only discrete information on the bioprocess. Incyte and Dencytee sensors offer an alternative for directly monitoring viable and total cell density in real time. The Incyte sensor enables real-time, and online measurement of permittivity, which correlates with the viable cell density. The measurement is not influenced by changes in the media, or by the presence of microcarriers, dead cells, and cellular debris. The Incyte sensor has been especially designed for monitoring the culture of mammalian and insect cells. It may also be used for controlling yeast culture and high-density bacterial fermentation. Online monitoring of permittivity with Incyte enables the early detection of process deviations without sampling and supports timely process adjustment. The Dencytee sensor allows the real-time, and online measurement of turbidity, which correlates with the total cell density. The turbidity of a cell suspension refers to the optical density of the cell suspension measured at a wavelength of 880 nm. Monitoring total cell density with Dencytee PRODUCT DESCRIPTIONPRODUCT DESCRIPTION is particularly suitable after inoculation when cells are expanding quickly but concentrations are low, making capacitance-based readings less reliable. The turbidity measurement is performed in the infra-red range and is therefore insensitive to changes in media color. Used in combination Incyte and Dencytee provide a complete and real-time characterization of the cell population for better bioprocess control. 5.2 The Theory of Permittivity Measurement (Incyte) 5.2.1 The Dual-Frequency Measurement Mode In an alternating electrical field viable cells behave like small capacitors, but not dead cells or cellular debris (Figure 1A). The charge of these small capacitors is measured by the Incyte sensor and reported as permittivity in pF/cm. The permittivity of viable cells is measured at a frequency specific of the cell type (fmes): usually 1 MHz for mammalian cells and bacteria, and 2 MHz for yeast (Figure 1B). It is continuously and automatically corrected for the background permittivity measured at high frequency (fhigh). This is the standard Dual-Frequency Measurement Mode. The permittivity measured by Incyte is specific to viable cells and correlates with the viable cell density, especially during the exponential growth phase. ELECTRODES ELECTRIC FIELD VIABLE CELLS POLARIZE DEAD CELLS HAVE A DAMAGED MEMBRANE AND DO NOT POLARIZE A + + + + + + + + + - - - - - - - - Figure 1: Incyte measurement principle. A: Viable but not dead cells exhibit a permittivity signal. 17Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION B: In the dual measurement mode the permittivity of viable cells is measured at a frequency specific to the cell type (fmes) and corrected by the background permittivity measured at high frequency (fhigh). 5.2.2 The Frequency Scan Mode The polarization behavior of cells varies strongly at different frequencies, as shown in Figure 1B. Cells fully polarize at low frequency whereas they hardly polarize at high frequency. The Incyte Scan measures the permittivity signal at 17 different frequencies between 0.3 and 10 MHz (Figure 2) and provides additional information especially on the cells physiology. permittivity[pF/cm] 0.3 1 20 Frequency [MHz] Medium Permittivity High frequency plateau Figure 2: The ideal frequency spectrum of mammalian cells, called beta-dispersion. permittivity[pF/cm] 0.1 1 20fmeas fhigh Frequency [MHz] 2 3 2 3 3 Medium Permittivity 1 OpOptitimamall frequency for correlation of permittivity and viable cell density Low frequency plateau Permittivity of cells and medium High frequency plateau 3 = - 3 1 B The analysis of the Incyte Scan provides additional information on the cell physiology1 . During the Incyte Scan the most relevant parameters - Δεmax, fc, and α - are automatically calculated and displayed on the Arc View Controller (Figure 3). The characteristic frequency, fc, provides an indication of the average cell diameter. If fc decreases, it indicates that the cell diameter increases during the culture. On the contrary, a shift of fc towards the higher frequency range indicates that the cell size reduces. The height of the low frequency plateau, Δεmax, correlates with the viable cell density. It increases as the cells grow. The slope (α) of the beta-dispersion at the characteristic frequency fc provides an indication of the distribution of the cell diameter. A steep slope, i.e. a large α indicates an homogenous culture. permittivity[pF/cm] 0.3 1.0 20 fcFF frequency [MHz] 3 Figure 3: Data Interpretation of the beta-dispersion, gained from an Incyte Scan. 5.3 The Theory of Turbidity Measurement Turbidity is commonly used to estimate the concentration of a cell suspension, especially in microbiology. The Dencytee sensor provides online measurement of the turbidity of a cell suspension, based on optical density (Figure 4). This measurement correlates with the total cell density in a bioprocess. The sensor emits light through a 5 mm window onto a light detector. Cells in suspension absorb and scatter light so less light is read by the detector. The measurement is made at NIR (near infra-red) wavelengths so it is insensitive to changes in media color. All particles that scatter light at 880 nm will be detected, including living and dead cells. 1 Ansorge S, Esteban G, Schmid G. On-line monitoring of infected Sf-9 insect cell cultures by scanning permittivity measurements and comparison with off- line biovolume measurements. Cytotechnology. 2007;55(2-3):115-124. doi:10.1007/s10616-007-9093-0. 19Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION Figure 4: Measurement principle of Dencytee sensors. 5.4 Hardware Description A standard measuring loop consists of a Sensor Unit connected to the Control Unit (Figure 5). It is recommended to measure in parallel turbidity with Dencytee and permittivity with Incyte in order to collect information on both total and viable cell density. LIGHT SOURCE, EMITS AT 880 NM DETECTOR 5 MM WINDOW Figure 5: Overview of the Cell Density Monitoring System (A: Arc View Controller, B: ComBox, C: Cell Density Integration Kit), connected to a process control system (PCS). PCS PC 4-20 mA OUTPUT BOX MODBUS DENCYTEE UNIT INCYTE UNIT DN 12 B OPC* COMBOX PCS INDUSTRIAL PC 4-20 mA OUTPUT BOX MODBUS DENCYTEE UNIT INCYTE UNIT DN 12 C OPC COMBOX CELL DENSITY INTEGRATION KITSKID CONTROLLER *By Modbus-Profibus DP Module, only for Arc View 465 (XL) *In order to enable a OPC connection between a Combox and the PCS via OPC only one Box can be installed per PC. VCD TCD PCS ARC VIEW CONTROLLER 4-20 mA OUTPUT BOX MODBUS / PROFIBUS* DENCYTEE UNIT INCYTE UNIT DN 12 A OPC 21Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 5.4.1 Control Unit 5.4.1.1 Arc View Controller The Arc View Controller (Figure 6) allows recording and plotting the measurements of up to four connected Sensor Units. It stores the measurements in Culture Files and the calibration data of all Sensor Units that have been connected. Hamilton Arc sensors, especially pH and dissolved oxygen sensors, may be visualized and recorded in parallel via a wireless connection. Most connection ports are available on the right hand side panel (Figure 6). The Arc View Controller 265 holds two M12 plugs on the side panel for the connection of two Sensor Units. They are labeled as Channel. The Arc View Controller 465 is a four channel Controller and has two more M12 plugs for connecting additional Sensor Units. The AUX and Modbus M12 connectors, and the Ethernet plug may be used to connect the Arc View Controller to the process control system. The USB ports may be used to connect a USB stick and export the recorded log files. It also serves to connect the Arc Wireless Converter MB (Ref 243498) or BT (Ref 243499) required to log Arc sensors. Figure 6: Arc View Controller 265, front and side panel. VCD TCD ON/OFF Channel 1 Channel 2 ModbusAUX DC 24V MOUNTING HOLES MAIN DISPLAY ON / OFF KEY SWITCH POWER SUPPLY M12 CONNECTORS TO SENSOR M12 CONNECTOR TO ANALOG BOX M12 CONNECTOR TO MODBUS RTU AND MODBUSPROFIBUS DP MODULE USB PORT RJ45 ETHERNET CONNECTOR ATTENTION! Instead of using the key switch it is recommended to plug/unplug the Arc View Controller, if the serial number is below 1500/5500. 5.4.1.2 ComBox The ComBox (Figure 7) allows recording and plotting the measurements of up to two connected Sensor Units on a personal computer. The measurements in Culture Files and the calibration data of all Sensor Units that have been connected are recorded on the PC. Hamilton Arc sensors, especially pH and dissolved oxygen sensors, may be visualized and recorded in parallel via a wireless connection, the Arc Wireless Converter MB (Ref 243498) or BT (Ref 243499) has to be connected to an USB Port. The connection ports are available on top of the ComBox (Figure 7). It holds two M12 plugs for the connection of two Sensor Units. They are labeled as Channel. The AUX and Modbus M12 connectors may be used to connect the ComBox to the process control system. Figure 7: ComBox, top panel ModbusAUXChannel 2Channel 1 MOUNTINGHOLES (onlyCellDensityIntegrationKit) POWER SUPPLY (not for Cell Density Integration Kit) M12 CONNECTORS TO SENSOR MOUNTINGHOLES (onlyCellDensityIntegrationKit) M12 CONNECTOR TO ANALOG BOX M12 CONNECTOR TO MODBUS RTU AND MODBUS-PROFIBUS DP MODULE 23Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 5.4.2 Incyte Sensor Unit An Incyte Sensor Unit consists of a Pre-amplifier and a Sensor (Figure 9). The Pre-amplifier and the Sensor have been factory calibrated as a pair. They are connected together via a VP11 connector. A status LED, indicating whether the sensor is ready for measurement is available on top of the Incyte pre-amplifier. The Incyte Sensor Unit is available with a shaft diameter of 12 mm (DN12) or 25 mm (DN25). The measurement tip is made of PEEK and contains four platinum electrodes for permittivity measurement. All wetted parts are USP Class VI certified. Figure 9: Incyte Sensor Unit assembled. STATUS LED  M12 CONNECTOR VP11 CONNECTOR PEEK TIP WITH PLATINUM ELECTRODES PRE-AMPLIFIERSENSOR INCYTESENSORUNIT 5.4.1.3 Cell Density Integration Kit The Cell Density Integration Kit (Figure 8) allows the complete integration of Incyte and Dencytee sensors into the process control system either by Modbus, OPC or 4-20 mA. The kit comprises an industrial PC, to run the required Software and a ComBox, to communicate to the Sensor Units and the process control system. The connection ports are available on top of the ComBox (Figure 7). It holds two M12 plugs for the connection of two Sensor Units. They are labeled as Channel. The AUX and Modbus M12 connectors may be used to connect the ComBox to the process control system. Figure 8: Cell Density Integration Kit comprising the industrial PC and ComBox NOTE: The connector is specified for 8 A and accepts a cable cross-section upto 1,5 mm2 . 12 34 LAN 1 LAN 2 MOUNTINGHOLES POWER SUPPLY USB PORT RJ45 ETHERNET CONNECTOR DVI-I PORT P-S (POWER-STATE EXIT) PC_ON (PC_ON INLET) GND (OV) 24V + (POWER SUPPLY) 25Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 5.5 Outline of the Control Unit Software 5.5.1 Run Menu The Run Menu (Figure 11) appears automatically upon starting the Arc View Controller or ComBox. It provides all the relevant information on a sensor connected to one of the Control Unit Channels. The sensor status is shown as a status LED in the Channel Tab: The Main Graph shows the measurements of the sensor connected to the selected channel over time. The measurement values are also indicated on top of the Main Graph. The Run Menu provides access to additional menus. • The Main Menu is accessible with the Menu Button located on the bottom right hand side of the screen. • Four additional Settings Menus (Unit, Measure, F-Scan and Culture) are available using the buttons located on the left hand side of the screen. • The Graph Arc Settings Menu is used to start recording additional Arc sensors usually pH or dissolved oxygen. • The Multi-View Tab enables displaying simultaneously sensors connected to different channels. It is also possible to overlay a running culture with an existing Culture File. NOTE: A light blue tab in the run menu indicates that the function is activated, to deactivate press the tab again. It turns to dark blue. 5.4.3 Dencytee Sensor Unit A Dencytee Sensor Unit consists of a Pre-amplifier and a Sensor (Figure 10). They are linked together via a connecting cable with a VP8 connector. The measurement window has an optical path length of 5 mm. The seal-less, sapphire window design assures the highest level of sterility, cleanability and sensor integrity. All wetted parts are USP Class VI certified. Figure 10: Picture of the assembled Dencytee Unit. M12 CONNECTOR CONNECTING CABLE VP8 CONNECTOR MEASUREMENT WINDOW PRE-AMPLIFIERSENSOR DENCYTEESENSORUNIT 27Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION Figure 11: Run Menu overview. CHANNEL TAB SENSOR STATUS LED MULTI-VIEW TAB MEASUREMENT VALUES MAIN GRAPHUNIT SETTINGS MENU MEASURE SETTINGS MENU FREQUENCY SCAN SETTINGS MENU CULTURE SETTINGS MENU GRAPH ARC MENU MAIN MENU 29Cell Density Monitoring System Operating Instructions PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 5.5.2 Sensor Channel Tab The Channel Tab (Figure 12) is used to view the measurements of the Sensor Unit connected to the selected channel. The status LED indicates whether the Sensor Unit is ready for measurement. The floppy icon indicates whether the measurements are currently recorded in a Culture File. Recording is indicated by the floppy icon without the red mark (x). FLOPPY ICON STATUS LED Figure 12: Sensor Channel Tab. 5.5.3 Main Menu The Main Menu (Figure 13) provides access to functions that may not be required on a daily basis such as calibration, user level settings, export of culture files, or connection to a process control system. The Explorer is used to export recorded Culture Files to a USB stick. The Administrator Menu enables following functionality: • Password control: this functionality allows modifying both the user level and the administrator level password. (see Chapter 9.1) • General settings: in this menu, it is possible to manage the general settings of the system such as touch screen calibration and setting of time and date (see Chapter 9.2). • Network applications: this functionality manages network applications such as remote access; OPC server and file sharing system (see Chapter 9.4). • Manage licenses: this interface allows the installation of new licenses (see Chapter 9.5). • Administrator files: this functionality allows the export of the Controller’s data files (see Chapter 9.6). • Update system: this enables the update of the Arc View Controller (see Chapter 9.7). NOTE: Update system functionality is not available for ComBox. The About Menu contains the hardware and firmware version of the complete measurement loop including Control Unit and Sensor Units. The Calibration Menu contains the Sensor Unit calibration and verification (called Validate / Check Unit Calibration) procedures. In the Advanced Settings Menu, all adjustments to connect the Control Unit to a process control system may be defined. Figure 13: Structure of the Main Menu. EXPLORER ADMINISTRATOR MENU ABOUT MENU RUN MENU CALIBRATION MENU ADVANCED SETTINGS MENU 31Cell Density Monitoring System Operating Instructions 5.5.4 Unit Settings Menu The Unit Setting Menu (Figure 14) contains all relevant information of the Sensor Unit connected to the selected channel, including • Sensor Unit type • Pre-amplifier information, notably the calibration lifetime • Sensor information, e.g. the number of sterilization cycles and the calibration lifetime Figure 14: Unit Settings Menu. 5.5.5 Measure Settings Menu In the Measure Settings Menu (Figure 15) all relevant adjustments for the bioprocess may be set. In Measure Settings Menu it is possible to • Run a product calibration using the Mark Zero Button • Select the measure mode • Set the Cell Density Unit and Cell Density Factor, which characterize the correlation of the permittivity measurement with an offline cell density measurement (for Incyte only). Figure 15: Measure Settings Menu of Incyte (A) and Dencytee (B). A B PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 33Cell Density Monitoring System Operating Instructions 5.5.6 Incyte Frequency Scan Settings Menu The Incyte Frequency Scan may be activated in the Frequency Scan Settings Menu (Figure 16). The product calibration of the Frequency Scan may also be performed in this Menu using the Mark Zero Button. It consists of a product calibration of the permittivity measurement at every frequency used in the Frequency Scan. Figure 16: Scan display. 5.5.7 Culture Settings Menu The Culture Settings Menu (Figure 17) controls all steps required to create a Culture Files: • Start recording measurements with the Record Button • Set the inoculation time to log the inoculation time in the Culture File • Add comments to log manually any event in the Culture File • End recording to stop logging the measurements in a Culture File If a recording has been started, the Culture Settings Menu also contains key information relative to the culture started and the Culture File in which the measurements are logged. Figure 17: Culture Settings Menu. PRODUCT DESCRIPTIONPRODUCT DESCRIPTION 35Cell Density Monitoring System Operating Instructions 6 Installation 6.1 Install the Control Unit 6.1.1 Install the Arc View Controller 1. Inspect the workplace environment for: - Compliance to technical specification - Providing enough space to perform the next steps 2. Unpack the Arc View Controller from the packaging Figure 18: Open packaging of the Arc View Controller. 3. Inspect the Controller and accessories for shipping damage or missing parts: The packaging should contain the Arc View Controller, one test report, the operating instructions, power supply, with 5 country specific cables. NOTE: The ordered licenses are installed on the Arc View Controller. 4. The Arc View Controller may be installed on a bench besides the bioreactor, or mounted on a wall. Please refer to the hardware specifications (Chapter 16.1.1) to choose the appropriate mounting elements to fix the controller. 5. Mount the power supply and connect it to the Arc View Controller (Figure 6). The Controller will automatically start up. 6. Switch off the Arc View Controller using the Key Switch (Figure 6). NOTE: It is recommended to perform a Installation Qualification. NOTE: It is recommended to connect the Arc View Controller to an Uninterruptible Power Supply (UPS) to ensure continuous recording of data, nevertheless the Controller restarts after power shutdown automatically at the last setting and continues recording. After a power shut down the Sensor Unit may require some time for new equilibration. 6.1.2 Install the ComBox 1. Inspect the workplace environment for: - Compliance to technical specification - Providing enough space to perform the next steps 2. Unpack the ComBox from the packaging Figure 19: Open packaging of the ComBox. INSTALLATIONINSTALLATION 37Cell Density Monitoring System Operating Instructions INSTALLATIONINSTALLATION 3. Inspect the ComBox and accessories for shipping damage or missing parts: The packaging should contain the ComBox, a power supply, with 5 country specific cables, the operating instructions and a USB stick with the Software and ordered licenses 4. The ComBox has to be installed on a bench besides the bioreactor 5. Connect the power supply to the ComBox (Figure 7) 6. Switch on the PC and, install the Software (see chapter 9.7.1.2) and connect the ComBox by USB, start the Software 7. Initialize the M12 connectors (see chapter 9.2.5) 8. Re-start the software and install the licenses (see chapter 9.5) NOTE: It is recommended to perform a Installation Qualification. 6.1.3 Install the Cell Density Integration Kit 1. Inspect the workplace environment for: - Compliance to technical specification - Providing enough space to perform the next steps 2. Unpack the ComBox from the packaging Figure 20: Open packaging of the Cell Density Integration Kit. 3. Inspect the ComBox and accessories for shipping damage or missing parts: The packaging should contain the ComBox, industrial PC (Software and licenses are pre-installed) and the operating instructions 4. The Cell Density Integration kit has to be installed inside a skid cabinet 5. Connect the industrial PC via the 4-pole connector with the power supply of the cabinet (Figure 8) 6. Connect the USB cable of the ComBox to the industrial PC 7. The ComBox has to be powered with 24 V via the Modbus Connector (Pin 1 and 3, see chapter 12.1.1) 8. Ground the ComBox using the thread on the top NOTE: It is recommended to perform a Installation Qualification. 6.2 Mount the Sensor in the Bioreactor 1. Unpack your sensor and Pre-amplifier from the protective box 2. Inspect the unit for shipping damage or missing parts. The packaging should contain the sensor, the Pre-amplifier and the calibration certificate Figure 21: Incyte Sensor Unit. 39Cell Density Monitoring System Operating Instructions 3. Select the most appropriate port for mounting the sensor (Figure 22). Choose a position where gas bubbles may not interfere with the measurement, for example in ports opposite to the sparger or at a significant distance. Gas bubbles interfere with the measurement and may cause reading errors. The electrodes of the Incyte sensor have to be kept at least 1 cm away from any solid parts, especially metallic ones. The electric field may be subject to interference and the measurement may be affected. AGITATION SPARGER 1 cm1 cm 1cm1cm A B Figure 22: Mounting restrictions, A: Space requirements / B: Reactor mounting restrictions. 4. Insert the sensor in the port. Do not scratch the platinum electrodes while inserting them. 5. Verify the sensor orientation to make sure that the Pre-amplifier can be mounted conveniently. The orientation notch of the sensor head is oriented opposite to the Pre-amplifier cable (Figure 23). NOTE: For mounting from a side-port ensure that the notch of the sensor head is mounted in a horizontal position facing upwards. Any angle of inclination may lead to an influence of the measurement. YES NO NOTCH OF THE SENSOR HEAD NOTCH OF THE SENSOR HEAD SIDE PORT INSTALLATION BENCH TOP INSTALLATION Figure 23: Mounting position of the Pre-amplifier. NOTE: The vessel should be pressure-less and cold when removing the sensor. The wear of safety goggles and protective gloves is prescribed. 6. Tighten the connection nut by hand. NOTE: Various housings are available to custom fit your sensor in your installation. INSTALLATIONINSTALLATION 41Cell Density Monitoring System Operating Instructions 6.3 Connect the Sensor Unit to the Control Unit 1. Switch off the Arc View Controller. 2. Verify that the sensor’s male VP connector and the Pre-amplifier’s female VP connector are dry and clean. 3. Align the sensor VP head and the Pre-amplifier connector and insert the Sensor in the Pre-amplifier. 4. Hand-tighten the connector ring of the Pre-amplifier to ensure a robust and waterproof seal (Figure 25). 5. Align the pins of the M12 connector and connect the Arc View Controller to the Pre-amplifier using the Sensor Unit Cable. ATTENTION! Make sure that the connector ring is tightened by hand. Never use a wrench or another tool to tighten or loosen the ring. ATTENTION! Do not use the Pre-amplifier as a lever to screw or unscrew the sensor. This may cause mechanical damage in the preamplifier. Always use the PG13.5 thread instead (Figure 20). ATTENTION! Please ensure a stable position of the Dencytee Pre-amplifier in a way that the cable is not bend. 6. Switch on the Arc View Controller. Wait at least 30 sec. until the Status LED on the main screen turns green (Figure 11) and an automated firmware update starts. The status LED of the Incyte Pre-amplifier (Figure 9) flashes green during this procedure. NOTE: When receiving a new Sensor Unit, make sure to first connect it to the Arc View Controller in order to transfer the calibration data. Do not use the Pre-amplifier with another sensor before the calibration data has been transferred. If you do so the calibration data will be lost. 1 2 3 Figure 24: Connection of a Sensor Unit. INSTALLATIONINSTALLATION 43Cell Density Monitoring System Operating Instructions 7 Operation of Incyte 7.1 Choose the Measurement Mode 1. Go to the Measure Settings Menu (Figure 15). Choose the measurement mode, from the predefined settings: • Cell culture > fmes = 1000kHz, fhigh = 10MHz, signal integration = HIGH • Yeast/fungi > fmes = 2000kHz, fhigh = 10MHz, signal integration = HIGH • Bacteria > fmes = 1000kHz, fhigh = 10MHz, signal integration = HIGH 2. Enter the User password if one has been defined. Press OK. 3. Please refer to Chapter 7.12.3 to create your own measurement mode. 7.2 Addition of Arc Sensors It is possible to record and plot Arc sensor measurements in parallel to cell density (Figure 25). That means in addition to the wired Incyte or Dencytee Unit, each Channel can also concurrently support two Arc Sensors. Cell density is monitored on Chart A. Another measurement may be shown additionally on Chart A (e.g. Arc pH, Arc DO or the conductivity measurement of Incyte). The remaining measurements will appear on Chart B. Chart A is shown as the Main Graph per default. NOTE: The measurement values transmitted over wireless communication are not intended to be used for process control. CHART A CHART B EXCHANGE CHART BUTTON A B Figure 25: Run Menu when Arc DO (Chart A) and Arc pH (Chart B) sensors are monitored in parallel to Incyte (A). The Frequency Scan has been activated (B). NOTE: The Graph Arc requires Arc Sensors, the Arc Wireless Converter MB (Ref 243498) or BT (Ref 243499) and Arc Wi Sensor Adapter 1G MB (Ref 242170), Arc Wi Sensor Adapter 2G MB (Ref 243030), Arc Wi Sensor Adapter 1G BT (Ref 243460) or Arc Wi Sensor Adapter 2G BT (Ref 243470). 1. Mount the Arc Wi Sensor Adapter between the Arc sensor and the Arc Cable VP 8 (Figure 26). Figure 26: Mounting an Arc Sensor. OPERATION OF INCYTEOPERATION OF INCYTE ARC SENSOR ARC WI SENSOR ADAPTER FOR WIRELESS FUNCTIONALITY VP8 CONNECTOR 45Cell Density Monitoring System Operating Instructions 2. Plug the Arc Wireless Converter MB or BT in one of the USB Ports on the side panel of the Arc View Controller. 3. Press the button Graph Arc on the bottom left hand corner of the Run Menu (Figure 11). The Arc View Controller will automatically scan for sensors. The Arc Wireless Converter MB or BT is flashing in red during this process. The Graph Arc display opens (Figure 27). The cell density measurement is set on Chart A per default. 4. Choose the second parameter to be shown on Chart A and press OK. The remaining parameters may be shown on Chart B. Figure 27: Graph Arc Menu. 5. Chart A is displayed per default on the main graph (Figure 23). Chart B is shown in the lower left corner. Press the Exchange Chart Button (Figure 23) to switch the display of the two charts. Chart B is not visible if the Incyte Scan mode is activated. It may be exchanged with Chart A on the Main Graph using the Exchange Chart Button. 7.3 Start Recording Measurements in a Culture File 1. Verify that the Status Indicator on the Run Menu of the Arc View Controller (Figure 11) and the status LED of the Pre-amplifier are green (Figure 9). 2. Go to the Culture Settings Menu and press the Record Button (Figure 17). 3. Enter the User password if one has been defined. Press OK. 4. Follow the guided procedure to start recording the Culture file. Define the directory to save the file, per default it is the home directory. Select the acquisition time, enter the optional name of the Culture file. Add possible comments and the user name. 5. The Record Button is now disabled, and the information relative to the Culture and the Culture File appears in the Culture Settings Menu. The Floppy Icon on the Channel Tab (Figure 12) also confirms that the measurements are being recorded. 6. Start the Frequency scan if required (Chapter 7.12.2). 7.4 Perform a Product Calibration and Set the Inoculation Time NOTE: Even though the Dual Frequency Measurement Mode reduces the influence of medium and medium changes on the measurement, it is usual to do a product calibration, i.e. a zero-adjustment before inoculation. The product calibration may be done automatically, which is the preferred setting. If the background signal is known the product calibration may be done manually or adjusted over time (see Chapter 7.12.4). 1. Equilibrate the Incyte Sensor in culture medium, for at least 30 minutes prior to product calibration and inoculation. 2. Verify on the Main Graph that the cell density measurement is stable. 3. Go to the Measure Settings Menu and press the Mark Zero Button to perform a product calibration (Figure 15). 4. Enter the User password if one has been defined. Press OK. 5. Select automatic. The cell density measurement is now compensated for an offset shown in the measurement values above the Main Graph on the Run Menu. 6. To assign the inoculation time, go to the Culture Settings Menu and press the Inoculate Button (Figure 17). The time on the Main Graph will be reset to zero. A new time scale is added to the Culture File to track the real culture time. 7. Enter the User password if one has been defined. Press OK. 7.5 Add a Comment to the Culture File During recording, a comment may be added at any time to the Culture File. This functionality may be used to track offline samplings. 1. Go to the Culture Settings Menu and press the Comments Button (Figure 17). OPERATION OF INCYTEOPERATION OF INCYTE 47Cell Density Monitoring System Operating Instructions 2. Enter the User password if one has been defined. Press OK. 3. Enter the title and the description of the comment. Both information are recorded in the Culture File but only the title is shown on the Main Graph. 7.6 Customize the Main Graph In the Run Menu it is possible to adjust the visualization settings of the Main Graph. • Press the Auto Scale Button to set an automated scaling. • Press the Follow Button, to track the current cell density measurements. • To display a certain part of the graph, move the graph to the left or the right using the Move Right and Move Left Buttons. • Click on the Zoom In Button to enlarge a detail in the graph and use the Zoom Out Button to set back. • Click on the X Scale or Y Scale Buttons to scale a specific area of the Main Graph between Xmin and Xmax respectively Ymin and Ymax. Both Y axes may be scaled independently by selecting axis 2. • Clear the graph at the end of the Culture by pressing the Clear Button. Figure 28: Visualization Settings of the Main Graph. 7.7 Setup a Multi View Graph In the Multi-View Tab (Figure 29), it is possible to plot the measurements of several channels on the same graph. In addition the current measurements may be plotted in parallel to a previously recorded Culture File. NOTE: In the Multi-View Graph you can display two of five measurements at the same time. 1. Go to the Multi-View Tab and select Graph 1. 2. Tick the Display Graph Check Box and enter a Title. 3. Select the Channel(s) to be displayed. 4. Select a recorded Culture File to be viewed. 5. Select the Measurements to be compared, e.g. cell density, Incyte Scan or conductivity (σ). If using the Incyte scan the following values are also available fc, Δε, and α. 6. After definition of the Multi-View, the chosen Measurements are plotted on the Multi-View Graph. All current measurement values are displayed on the upper right area. Figure 29: Multi View setup. Y-SCALE BUTTON CLEAR BUTTON X-SCALE BUTTON AUTO SCALE BUTTON FOLLOW BUTTON MOVE LEFT BUTTON MOVE RIGHT BUTTON ZOOM IN BUTTON ZOOM OUT BUTTON OPERATION OF INCYTEOPERATION OF INCYTE 49Cell Density Monitoring System Operating Instructions 7.8 Stop Recording the Culture File 1. Go to the Culture Settings Menu and press the End Recording Button (Figure 17). 2. Enter the User password if one has been defined. Press OK. 3. Export the culture file (Chapter 7.10). 4. Dismount the Sensor (Chapter 7.9). 7.9 Dismount the Sensor Unit 1. Switch off the Arc View Controller using the Key Switch. 2. Disconnect the Sensor Unit Cable from the Pre-amplifier. 3. Loosen the connector ring of the Pre-amplifier by hand and remove the Pre-Amplifier from the Sensor (Figure 24). 4. Loosen the thread of the sensor manually and remove the sensor from the bioreactor. ATTENTION: Make sure that the connector ring is loosen by hand. Never use a wrench or another tool to tighten or loosened the ring. ATTENTION: Do not use the Pre-amplifier as a lever to screw or unscrew the sensor. This may cause mechanical damage in the preamplifier. Always use the PG13.5 thread instead (Figure 24). 7.10 Export a Culture File 1. Go to the Main Menu and press the Explorer Button (Figure 13). The Explorer (Figure 30) displays simultaneously two directories; the Arc View Controller Home per default. Select USB for the second directory. 2. Enter the User password if one has been defined. Press OK. 3. Plug a USB Stick on the side panel of the Arc View Controller (Figure 7) or PC, if using the ComBox (Figure 8). Click on the refresh, select the USB stick and press open dir. 4. Select the Culture File in the home directory and press the Copy Button. 5. Refresh the USB directory, and verify that the Culture File has been transferred successfully. 6. Delete the Culture File copied to free storage space. 7. If needed, the Culture File may be renamed, by selecting the file and pressing the Rename Button. 8. It is possible to create a new directory by pressing the Create Directory Button. NOTE: Screenshots (see Chapter 11.4.1) are also stored in the Arc View Controller Home directory. Figure 30: Explorer Menu. OPERATION OF INCYTEOPERATION OF INCYTE 51Cell Density Monitoring System Operating Instructions 7.11 Import a Culture File in an Excel Spreadsheet 1. Open Microsoft Excel and select the Data Tab. 2. In the area Get External Data press the From Text Icon. 3. Choose the Culture file and click open. 4. Mark the column Date and use a User-defined Format to display the time recordings: yyyy. mm.dd.hh.mm.ss. 5. Select the columns Record and Culture time and define the format as Numbers. The file (Figure 31) is composed of three parts. The Header contains general information on the Culture. The Events list contains all significant events and modifications that happened while the culture file was recorded. The Measurement table with the following measurement values: • Date • Record time (h): time in hours since the start of recording • Culture time (h): time in hours since the inoculation • Comment: contains start recording, inoculation, comments and stop recording • Raw permittivity measurements • Cell density value correlated to the product calibration and cell factor HEADER EVENTS MEASUREMENT Figure 31: Culture File. 7.12 Further Measurement Settings 7.12.1 Set the Cell Factor The measured permittivity may be correlated to offline measurements, by defining a cell factor. 1. Go to the Measure Tab (Figure 15) and press the Cell Density Unit / Factor Button. 2. Enter the User password if one has been defined. Press OK. 3. Choose a UnitFactor from the drop-down list and press edit. Insert the Cell Factor. To learn how to calculate the cell factor refer to Chapter 18.2. 4. The Cell Factor is displayed as measurement value on top of the Main Graph (Figure 11). Per default the cell factor is 1. NOTE: The Permittivity Factor 1.00 pF/cm is the factory value, which cannot be changed. This is the factor the measurement is relying on. The UnitFactors are the correlated customer factors. 7.12.2 Start the Incyte Scan NOTE: The Scan License is required to use the Incyte Scan option. 1. Go to the Scan Settings Menu (Figure 6) and activate the Frequency Scan. 2. Enter the User password if one has been defined. Press OK. 3. The curve of the Frequency Scan is displayed in the Run Menu, below the Main Graph. Δεmax, fc and α are calculated and shown as well (Figure 32). 4. Equilibrate the sensor in the culture medium for at least 30 minutes. 5. Go to the Scan Settings Menu (Figure 6) and press the Mark Zero Button to perform the product calibration of the Frequency Scan. OPERATION OF INCYTEOPERATION OF INCYTE 53Cell Density Monitoring System Operating Instructions DISPLAY OF INCYTE FREQUENCY SCAN Figure 32: Run Menu with activated Incyte Frequency Scan. 7.12.3 Define the Custom Measurement Mode The measurement frequencies, integration time and pulse filter may be adjusted to define a custom Measurement Mode. 1. Go to the Measure Tab (Figure 15) and select MeasureMode1 in the drop down menu. Press the Edit Button. 2. Enter the User password if one has been defined. Press OK. 3. Enter a new name for the measure mode. Define the measurement frequency (fmes) and the background frequency (fhigh) (Figure 33). 4. Select the integration time, which creates a moving average over a defined number of measurements: • NONE: 1 measurement - Integration time : 6 sec • LOW: 3 measurement3 - Integration time : 18 sec • MEDIUM: 22 measurements - Integration time : 2.2 min • HIGH: 111 measurements : Integration time : 11.1 min (default) • SUPER HIGH: 333 measurements - Integration time : 33.3 min • Custom, whereas 6 sec is the minimum time 5. If required activate the Pulse filter. The Pulse filter is only necessary for media with a high content of solid particles. 6. Press the Save Button. Figure 33: Custom Measurement Mode. 7.12.4 Perform a Manual Product Calibration 1. Go to the Measure Settings Menu and press the Mark Zero Button to perform a product calibration (Figure 15). 2. Enter the User password if one has been defined. Press OK. 3. Select manual and compensate the cell density measurement manually. Enter the known permittivity of the background. The cell density measurement is now compensated for an offset shown in the measurement values above the Main Graph on the Run Menu. 7.13 Use Sensor Cleaning (Probe Cleaning) The Sensor Cleaning is deactivated per default. It is only required if cell adhesion at the sensor tip is noticed. The Cleaning function reduces the attachment of cells and may be required in few processes, i.e. in long-term cell culture or fermentation of filamentous fungi. Use the cleaning mode with caution and only if adhesion of cells at the platinum electrodes is noticed or suspected. Start with SHORT cleaning time and a long Auto-Cleaning Period, at least 12 h. Increase cleaning time / decrease the Auto-Cleaning Period only if no improvement is observed. 1. Enter the Advanced Settings Menu and select Probe cleaning. 2. Enter the User password if one has been defined. Press OK. OPERATION OF INCYTEOPERATION OF INCYTE 55Cell Density Monitoring System Operating Instructions Figure 34: Sensor Cleaning. 3. Choose between the Manual or Auto Cleaning. The auto cleaning consists of periodic cleaning cycles, which start automatically after defined period whereas in the manual cleaning mode the user starts a cleaning cycle manually. 4. Go to the Run Menu and press the Measure Button. 5. Select the duration of the Cleaning: short (cleaning time 30 s) or long (cleaning time 2 min). 6. Enter the User password if one has been defined. Press OK. 7. Activate the Auto-Cleaning (Figure 35) and define the Auto-Cleaning Period, i.e. the time between two Auto-Cleaning cycles. 8. Alternatively, press Clean now to start a Cleaning cycle manually directly. 9. Confirm the password request, without entering any code (see Chapter 9.1). NOTE: Measurements are unavailable during a cleaning cycle. The status LED on the Pre-amplifier (Figure 9) is flashing green during this procedure. The sensor may need time for equilibration after the cleaning cycle. Figure 35: Selection of the cleaning mode. 8 Operation of Dencytee 8.1 Choose the Measurement Mode 1. Go to the Measure Settings Menu (Figure 15). Choose the measurement mode, from the predefined settings of Measure Mode 1, 2 or 3, which are a combination of pulse filter and signal integration. 2. Enter the User password if one has been defined. Press OK. 3. To create an own measurement mode see Chapter 8.12 8.2 Addition of Arc Sensors It is possible to record and plot Arc sensor measurements, in parallel to cell density (Figure 36). That means in addition to the wired Incyte or Dencytee Unit, each channel can also concurrently support two Arc Sensors. Cell density is monitored on Chart A. Another measurement may be shown additionally on Chart A (e.g. Arc pH, Arc DO). The remaining measurements will appear on Chart B. Chart A is shown as the Main Graph per default. OPERATION OF DENCYTEEOPERATION OF INCYTE 57Cell Density Monitoring System Operating Instructions NOTE: The measurement values transmitted over wireless communication are not intended to be used for process control. CHART A CHART B EXCHANGE CHART BUTTON Figure 36: Cell Density Graph with Arc DO, and Arc pH records. NOTE: The Graph Arc requires Arc Sensors, the Arc Wireless Converter MB (Ref 243498) or BT (Ref 243499) and Arc Wi Sensor Adapter 1G MB (Ref 242170), Arc Wi Sensor Adapter 2G MB (Ref 243030), Arc Wi Sensor Adapter 1G BT (Ref 243460) or Arc Wi Sensor Adapter 2G BT (Ref 243470). OPERATION OF DENCYTEEOPERATION OF DENCYTEE 1. Mount the Arc Wi Sensor Adapter between the Arc sensor and the Arc Cable VP 8 (Figure 37). ARC SENSOR ARC WI SENSOR ADAPTER FOR WIRELESS FUNCTIONALITY VP8 CONNECTOR Figure 37: Mounting an Arc Sensor. 2. Plug the Arc Wireless Converter MB or BT in one of the USB Ports on the side panel of the Arc View Controller. 3. Press the button Graph Arc on the bottom left hand corner of the Run Menu (Figure 11). The Arc View Controller will automatically scan for sensors. The Arc Wireless Converter MB or BT is flashing in red during this process. The Graph Arc display opens (Figure 38). The cell density measurement Dencytee is set on Chart A per default. 4. Choose the second parameter to be shown on Chart A and press OK. The remaining parameters will be shown on Chart B. 59Cell Density Monitoring System Operating Instructions Figure 38: Graph Arc Menu. 5. Chart A is displayed per default on the main graph (Figure 36). Chart B is shown in the lower left corner. Press the Exchange Chart Button (Figure 36) to switch the display of the both charts. 8.3 Start Recording Measurements in a Culture File 1. Verify that the Status Indicator on the Run Menu of the Arc View Controller (Figure 11) is green. 2. Go to the Culture Settings Menu and press the Record Button (Figure 17). 3. Enter the User password if one has been defined. Press OK. 4. Follow the guided procedure to start recording the Culture file. Define the directory to save the file, per default it is the home directory. Select the acquisition time, enter the optional name of the Culture file. Add possible comments and the user name. 5. The Record Button is now disabled, and the information relative to the Culture and the Culture File appears in the Culture Settings Menu. The Floppy Icon on the Channel Tab (Figure 12) also confirms that the measurements are being recorded. 8.4 Perform Product Calibration and Set the Inoculation Time NOTE: Even though our measurement principle is not influenced by the media color, it is usual to do a product calibration, i.e. a zero-adjustment before inoculation. 1. Equilibrate the Incyte Sensor in culture medium, for at least 30 minutes prior to product calibration and inoculation. 2. Verify on the Main Graph that the cell density measurement is stable. 3. Go to the Measure Settings Menu and press the Mark Zero Button to perform a product calibration (Figure 15). 4. Enter the User password if one has been defined. Press OK. 5. Select automatic. The cell density measurement is now compensated for an offset shown in the measurement values above the Main Graph on the Run Menu. 6. To assign the inoculation time, go to the Culture Settings Menu and press the Inoculate Button (Figure 17). The time on the Main Graph will be reset to zero. A new time scale is added to the Culture File to track the real culture time. 8.5 Add a Comment to the Culture File During recording, a comment may be added any time to the Culture File. This functionality may be used to track offline samplings. 1. Go to the Culture Settings Menu and press the Comments Button (Figure 17). 2. Enter the User password if one has been defined. Press OK. 3. Enter the title and the description of the comment. Both information are recorded in the Culture File but only the title is shown on the Main Graph. OPERATION OF DENCYTEEOPERATION OF DENCYTEE 61Cell Density Monitoring System Operating Instructions 8.6 Customize the Main Graph In the Run Menu it is possible to adjust the visualization settings of the Main Graph. • Press the Auto Scale Button to set an automated scaling. • Press the Follow Button, to track the current cell density measurements. • To display a certain part of the graph, move the graph to the left or the right using the Move Right and Move Left Buttons. • Click on the Zoom In Button to enlarge a detail in the graph and use the Zoom Out Button to set back. • Click on the X Scale or Y Scale Buttons to scale a specific area of the Main Graph between Xmin and Xmax respectively Ymin and Ymax. Both Y axes may be scaled independently by selecting axis 2. • Clear the graph at the end of the Culture by pressing the Clear Button. Figure 39: Adjustments on the Main Graph. 8.7 Setup a Multi View Graph In the Multi-View Tab (Figure 40), it is possible to plot Incyte and Dencytee measurements in the same graph. In addition the current data may be plotted in parallel to the graph of a previous recorded culture file NOTE: When setting up of the Multi-View, be aware that only 2 types of curves can be displayed on a single graph. 1. Go to the Multi-View Tab and select Graph 1 (Figure 40), enter a title to label the Multi-View graph. 2. Tick the display graph button and select the channel to be displayed. 3. Select a recorded culture file in the File selection. 4. Define the measurement type, e.g. cell density. 5. After definition of the Multi-View, all measurements are plotted in the defined Multi-View Graph. Figure 40: Multi View setup. Y-SCALE BUTTON CLEAR BUTTON X-SCALE BUTTON AUTO SCALE BUTTON FOLLOW BUTTON MOVE LEFT BUTTON MOVE RIGHT BUTTON ZOOM IN BUTTON ZOOM OUT BUTTON OPERATION OF DENCYTEEOPERATION OF DENCYTEE 63Cell Density Monitoring System Operating Instructions 8.8 Stop Recording the Culture File 1. Go to the Culture Settings Menu and press the End Recording Button (Figure 17). 2. Enter the User password if one has been defined. Press OK. 3. Export the culture file (Chapter 8.10). 4. Dismount the Sensor (Chapter 8.9). 8.9 Dismount the Sensor Unit 1. Switch off the Arc View Controller using the Key Switch. 2. Disconnect the Sensor Unit Cable from the Pre-amplifier. 3. Loosen the connector ring of the Pre-amplifier by hand and remove the Pre-Amplifier from the Sensor (Figure 24). 4. Loosen the thread of the sensor manually and remove the sensor from the bioreactor. ATTENTION! Make sure that the connector ring is loosen by hand. Never use a wrench or another tool to tighten or loosen the ring (Figure 24). 8.10 Export a Culture File 1. Go to the Main Menu and press the Explorer Button (Figure 13). The Explorer (Figure 41) displays simultaneously two directories; the Arc View Controller Home per default. Select USB for the second directory. 2. Enter the User password if one has been defined. Press OK. 3. Plug a USB Stick on the side panel of the Arc View Controller (Figure 7) or PC, if using the ComBox (Figure 8). Click on the refresh, select the USB stick and press open dir. 4. Select the Culture file in the home directory and press the Copy Button. 5. Refresh the USB directory, and verify that the Culture file has been transferred successfully. 6. Delete the Culture file copied to free storage space. 7. If needed, the Culture file may be renamed, by selecting the file and pressing the Rename Button. 8. It is possible to create a new directory by pressing the Create Directory Button. NOTE: Screenshots (see Chapter 11.4.1) are also stored in the Arc View Controller Home directory. Figure 41: Explorer Menu. 8.11 Import the Culture File in an Excel Spreadsheet 1. Open Microsoft Excel and select the Data Tab. 2. In the area Get External Data press the From Text Icon. 3. Choose the Culture file and click open. 4. Mark the column Date and use a User-defined Format to display the time recordings: yyyy. mm.dd.hh.mm.ss. 5. Select the columns Record and Culture time and define the format as Numbers. OPERATION OF DENCYTEEOPERATION OF DENCYTEE 65Cell Density Monitoring System Operating Instructions The file (Figure 42) is composed of three parts. The Header contains general information on the Culture. The Events list contains all significant events and modifications that happened while the culture file was recorded. The Measurement table with the following measurement values: • Date • Record time (h): time in hours since the start of Recording • Culture time (h): time in hours since the inoculation • Comment: contains start recording, inoculation, comments and stop recording • Raw turbidity measurements • Cell density value correlated to the product calibration and cell factor a HEADER EVENTS MEASUREMENT Figure 42: Culture File 8.12 Define the Custom Measurement Mode The measurement, integration time and pulse filter may be adjusted to define the custom Measurement Mode. 1. Go to the Measure Tab (Figure 15) and select MeasureMode1 in the drop down menu. Press the Edit Button. 2. Enter the User password if one has been defined. Press OK. 3. Enter a new name for the measure mode (Figure 43). 4. Select the integration time, this creates a moving average over a defined number of measurements • NONE: 1 measurement - Integration time : 6 sec • LOW: 3 measurement3 - Integration time : 18 sec • MEDIUM: 22 measurements - Integration time : 2.2 min • HIGH: 111 measurements : Integration time : 11.1 min (default) • SUPER HIGH: 333 measurements - Integration time : 33.3 min • Custom, whereas 6 sec is the minimum time 5. Activate the Pulse filter. The pulse filter is only necessary for media with a high content of solid particles. 6. Press the Save Button. Figure 43: Create new measure mode. 9 Administrator Menu From the Administrator Menu all system settings are accessible, including password control, network settings and applications, general settings, like the touch screen calibration (only Arc View Controller), the possibility to install licenses or update the system and transfer the Administrator Files. NOTE: To enter the Administrator Menu, a password is requested, enter the code 14147125 and press ok. ADMINISTRATOR MENUOPERATION OF DENCYTEE 67Cell Density Monitoring System Operating Instructions 9.1 Definition of the User Level To limit the access to the Arc View Controller, following User Level may be defined (Figure 44): • Viewer: permission to see the Run Menu, no password needed • User: permission to change the Run Settings, a User password may be defined • Administrator: complete access to the Arc View Controller, default code 14147125 NOTE: As soon as a User Level password is defined, there is no possibility to delete it. Figure 44: Definition of the User Level. 9.1.1 Change the Administrator Password 1. Enter the Administrator Menu. 2. Press the Password Control button. 3. Insert the existing password in the old password-box of the Administrator level and define a new one, confirm it and press validate. NOTE: Please note the new administrator password. ADMINISTRATOR MENUADMINISTRATOR MENU 9.1.2 Define the User Level 1. Enter the Administrator Menu. 2. Press the Password Control button. 3. Leave the old password-box of the User level empty and define a new one, confirm it and press validate. NOTE: Please note the new user password. 9.2 Adjust the General Settings In the General Settings (Figure 45) it is possible to perform a touch screen calibration, adjust the decimal separator in the Culture file, activate the continuous plotting of the cell density measurement (without recording), adjust the time and date settings. Additionally it is possible to initialize the M12 connectors and activate the Cell Density Integration Kit option. Figure 45: Adjustment of the general settings. 69Cell Density Monitoring System Operating Instructions 9.2.1 Touch Screen Calibration 1. Enter the Administrator Menu and press general settings button. 2. Select touch screen calibration and follow the instructions on the Main Display. NOTE: Available for the Arc View Controller only. 9.2.2 Adjustment of the Decimal Separator in the Culture File If the decimal separate output is not compatible to the output software, e.g. Windows Excel, the record file may be adjusted. 1. Enter the Administrator Menu and press general settings button. 2. Select the decimal separator to be dot or comma, per default international is selected. 9.2.3 Continuous Display of Graphed Data Always display the graph may be activated, to plot the current measurements without recording. 1. Enter the Administrator Menu and press general settings button. 2. Activate the Chart options and the graph is plotted continuously. 9.2.4 Adjustment of Date and Time Settings NOTE: Available only for the Arc View Controller and Cell Density Integration Kit (if screen connected). Time and date may be adjusted if displayed incorrect. 1. Enter the Administrator Menu and press general settings button. 2. Press set time and adjust date, time and time-zone (offset to UTC). 3. Press OK, the Control Unit requests for a re-start to change the date and time settings. 9.2.5 Initialize the Arc View M12 Connectors NOTE: You can use 1 of the combinations below for Channel assignment (including Aux and Modbus channels): 1. M12 cable plus Pre-amplifier with sensor attached 2. M12 cable plus Pre-amplifier with Cal Simulator attached 3. M12 cable plus Pre-amplifier only Initialize the M12 connectors after a Software update (see Chapter 9.7.1), or the re-set of the factory settings (see Chapter 9.7.2), This ensures that the Sensor Units are correctly detected. 1. Enter the Administrator Menu and press general settings button. 2. Press Initialization of the Control Unit and follow the instructions on the main display. 9.2.6 Enable the Cell Density Integration Kit Option NOTE: This procedure enables the verification and calibration by Modbus, but disables this possibility on the screen. It is only required if the procedure should run via Modbus, e.g. if the system is integrated in a process control system. 1. Enter the Administrator Menu and press the general settings button. 2. Activate the Cell Density Integration Kit option ADMINISTRATOR MENUADMINISTRATOR MENU 71Cell Density Monitoring System Operating Instructions ADMINISTRATOR MENUADMINISTRATOR MENU 9.3 Read the Network Settings NOTE: Available only for the Arc View Controller and Cell Density Integration Kit (if screen connected). In the Network Settings it is possible to read the Serial Number (Device Name) of the Arc View Controller and Cell Density Integration Kit, as well as the max address, IP address, subnet mask and gateway address (Figure 42). This information is relevant for the connection of the Arc View Controller and Cell Density Integration Kit to the process control system via OPC or the integration of the Controller or Kit in a network. Figure 46: Network Settings display. The Network Settings page lets you configure the network configuration of your device. In order to use the networking features of the device, an IP address must be assigned. Depending on the configuration of the network where you intend to install the device, one of the following cases applies: • Manual IP addressing (checkbox not selected) • Automatic IP addressing (checkbox selected) Manual IP addressing means that the IP configuration is given by the network administrator. The relevant information (IP address of your device, subnet mask, gateway) must be entered in the corresponding fields. Depending on your company's policy, you should not take an unassigned IP address randomly for your device since it may result in networking problems. Dynamic addressing means that the IP configuration is given by a service, called DHCP, running on the network. This is the default configuration and is suitable in most cases. NOTE: Check, when troubleshooting dynamic IP addressing: • Cabling • DHCP service is running on the physical network where your Control Unit is connected. • Check with your system administrator if the policy requires a manual entry of the device's MAC address in the DHCP configuration. The MAC address is written on the small sticker with barcode located on the main board. In addition the MAC address is shown in the Network Settings View. 9.4 Adjust the Network Applications In the Network Applications it is possible to activate share data files for the integration in a Windows network. Additionally the OPC communication (see Chapter 12.2) may be enabled (Figure 47). Figure 47: Activate the Network Applications. 73Cell Density Monitoring System Operating Instructions 9.4.1 Activate Share Data Files NOTE: Available only for the Arc View Controller and Cell Density Integration Kit (if screen connected). 1. Enter the Administrator Menu and press the Network Applications button. 2. Activate the share data files to enable the integration of the Control Unit into a Windows Network. 3. Set Up the integration in the local network, using the Network Settings (see Chapter 9.3). 9.4.2 Activate OPC Communication 1. Enter the Administrator Menu and press the Network Applications button. 2. Activate OPC to allow OPC communication to the process control system (see Chapter 12.2). 9.5 Install a License NOTE: The ordered licenses are installed on the Arc View Controller. The ordered licenses are stored on the USB stick provided with the ComBox. 1. Enter the Administrator Menu and press manage license. 2. Connect the USB stick with the new license on one of the USB ports at the right panel (Figure 6) of the Arc View Controller, or a PC where the ComBox is connected. 3. Click the load license button and select the license and press ok (Figure 48). NOTE: After installation of the OPC license a re-start is requested. Figure 48: License Menu. 9.6 Transfer Administrator Files The Administrator files (Figure 49) may be transferred for storage in an external hard drive or used for trouble shooting purposes. Figure 49: Transfer the administrator files. ADMINISTRATOR MENUADMINISTRATOR MENU 75Cell Density Monitoring System Operating Instructions 9.6.1 Export the Log Files NOTE: Available only for the Arc View Controller and Cell Density Integration Kit (if screen connected). 1. Enter the Administrator Menu and press Administrator Files. 2. Connect a USB stick on one of the USB ports at the right panel (Figure 6) of the Arc View Controller, or the industrial PC of the Cell Density Integration Kit. 3. Follow the instructions on the Main Display. The log files may be needed for trouble shooting the system. 9.6.2 Erase Temporary Files NOTE: Available only for the Arc View Controller and Cell Density Integration Kit (if screen connected). 1. Enter the Administrator Menu and press Administrator Files. 2. Press Erase temporary files to clean the memory of the Arc View Controller, or the industrial PC of the Cell Density Integration Kit. 3. Follow the instruction on the main display of the Control Unit. 9.6.3 Export Culture Data NOTE: This function is available starting with Software Version 3.8. 1. Enter the Administrator Menu and press Administrator Files. 2. Select the Channel, to export the corresponding data, and enter the initial and final date for the export. 3. Connect a USB stick on one of the USB ports at the right panel (Figure 6) of the Arc View Controller or the PC, where the ComBox is Installed and follow the instruction on the Main Display of the Control Unit. 9.7 Update or Re-Set the Arc View Controller or Cell Density Integration Kit In the Update System Menu (Figure 50), it is possible to re-set the Control Unit to the factory Software, or update the system. Figure 50: Update System Setting. NOTE: The settings are reset to factory when updating the Control Unit. 9.7.1 Software Update of the Control Unit 9.7.1.1 Software Update of the Arc View Controller or Cell Density Integration Kit 1. Please check our Homepage www.hamiltoncompany.com/arc_view_update for new updates. Save the file on an USB stick. 2. Enter the Administrator Menu and press Update system. 3. Connect the USB stick to the Arc View Controller and select the update file. Activate the system update. ADMINISTRATOR MENUADMINISTRATOR MENU 77Cell Density Monitoring System Operating Instructions 4. Follow the instructions of the Main Display. 5. After the update is finished initialize the M12 Connectors (see Chapter 9.2.5). 9.7.1.2 Software installation/update of the ComBox NOTE: Prior to the update, save the current data. 9.7.1.2.1 Procedure for one ComBox per PC 1. Please check our Homepage www.hamiltoncompany.com/arc_view_update for the newest Software version. 2. Download. 3. Execute the "Combox_Setup.exe". 4. Follow the instructions on the screen, until the Software is installed. 5. Connect the ComBox and close the Installer Software. 6. Start the Software and initialize the M12 connectors (see Chapter 9.2.5), close the Software. 7. Start the Software and install the licenses (see Chapter 9.5). 9.7.1.2.2 Procedure for two ComBoxes per PC 1. Follow steps 1 and 5 in Chapter 9.7.1.2.1. 2. Duplicate folder "C:\Program Files (x86)\Hamilton\CellDensityComBox". 3. Adjust the folder name, while adding the number of the box, e.g. "...ComBox_2". 4. Create a shortcut for the "interface_Biomas_rack_V2.exe" in the new folder. 5. Save shortcut on the desktop under the defined name (in this case ...2). 7. Connect ComBox 2 and start Software 2. 8. Switch to the first Software, initialize the Channels (see Chapter 9.2.5) and install the licenses (see Chapter 9.5). 9. Switch to the second Software, initialize the Channels (see Chapter 9.2.5) and install the licenses (see Chapter 9.5). ADMINISTRATOR MENUADMINISTRATOR MENU NOTE: If you have more than two ComBoxes, repeat the procedure for the amount of ComBoxes available. 9.7.2 Re-Set the Arc View Controller or Cell Density Integration Kit In case of Software failure, it is possible to re-set the factory settings. 1. Enter the Administrator Menu and press Update system. 2. Connect a USB Keyboard and a USB Mouse to the USB connections of the Arc View Controller (Figure 6), or industrial PC of the Cell Density Integration Kit (Figure 8). 3. Press factory version, to re-set the factory setting of the Control Unit. 4. Follow the instructions on the Main Display. 5. After the system is re-set initialize the M12 Connectors (see Chapter 9.2.5). 9.7.3 Update Recovery of the Arc View Controller In case of script update, it is possible to perform a recovery update process instead of a complete software update. NOTE: This functionality is only available for Arc View Controller. 1. Connect the USB stick with the scripts (provided by Hamilton) to the Arc View Controller. 2. Enter the Administrator Menu select the .zip file, which contains the recovery scripts and press Update recovery. The software now checks for completeness of the data and copies the files to the recovery partition. 3. Upon completion the Arc View Controller requires two re-starts, which which have to be confirmed after pressing Update recovery. NOTE: If a wrong file is selected a instruction for the correct procedure will be displayed on the Controller screen. 79Cell Density Monitoring System Operating Instructions 10 Maintenance 10.1 Daily Maintenance As a daily routine, at least prior to every run, the Status Indication on the Arc View Controller may be checked (Figure 12). The Indicator should be green, if not please refer to Chapter 11.1. 10.2 Weekly Maintenance The Weekly Maintenance should be performed on a weekly basis or at the end of a Culture. 1. Check the Status Indicator on the Arc View Controller (Figure 12). The Indicator should be green. 2. Decontaminate the sensor by Sterilization in Place, Cleaning in Place or Autoclavation. 3. Press the increment sterilization button (Figure 14) in the Unit Tab after autoclavation. 4. If needed perform the Manual Cleaning Procedure (see Chapter 10.3). 5. Download the recent Culture Files if not yet done. NOTE: Incyte and Dencytee Sensors need to be replaced at least after 100 Autoclavation, Sterilization-in place, or Cleaning-in-place cycles. The Status Indicator on the Arc View Controller turns orange and a request to replace the sensor appears. Failure to do so may lead to measurement errors. ATTENTION! The pre-Amplifier is not autoclavable. Dismount it prior to autoclavation to avoid damages to the electronics Please refer to Chapter 7.9 and 8.9 for instructions. Protect the sensor’s male connector with aluminum foil to prevent condensed vapors from reaching the electrical contacts during sterilization. 10.3 Manual Cleaning Procedure at Process End for Cell Density Sensors Manual cleaning may be needed, when a thick biofilm sticks on the sensor. 10.3.1 Incyte Sensors 1. After decontamination, immerse the PEEK sensor tip in Sodium Sulfite (Na2 SO3), 15 g/L. 2. Connect the pre-amplifier (see Chapter 6.3) and perform 2 cycles of short cleaning (see Chapter 7.13). 3. Remove the Sensor from the solution and rinse the residual sodium sulfite with deionised water. 4. Perform the Sensor Unit Verification procedure (see Chapter 10.5). 10.3.2 Dencytee Sensors 1. After decontamination, emerse the Dencytee Sensor for 30 minutes in 1% (v/V) acetic acid and afterwards 30 min in 1% (w/v) NaOH at room temperature. 2. Rinse the Sensor with deionized water. 3. Perform the Sensor Unit Verification procedure (see Chapter 10.5). 10.4 Yearly Maintenance 1. Perform the Weekly Maintenance. 2. Perform the Sensor Verification procedure. 3. Export the temporary data (see Chapter 9.6.3). 4. Delete the temporary files (see Chapter 9.6.2). 5. Update the Control Unit with the latest Software Update (see Chapter 9.7.1). MAINTENANCEMAINTENANCE 81Cell Density Monitoring System Operating Instructions MAINTENANCEMAINTENANCE 10.5 Sensor Unit Verification NOTE: Verification just available on screen if Cell Density Integration Kit option not activated. A report of the verification is automatically created and stored (explorer or local hard disk). All tolerances and acceptance criteria are shown in the report. 10.5.1 Verification of an Incyte Sensor Unit NOTE: The Capacitance Simulator (Ref 243743) and Hamilton 12880 μS/cm Conductivity Standard (Ref 238988), formerly called Solution A, as well as a thermometer are needed to perform the Sensor Unit Verification. 1. Connect the Incyte Unit, power on the Control Unit and equilibrate for at least 30 minutes prior to starting the Verification procedure. 2. Go to the Main Menu and enter the Calibration Menu (Figure 13). Select the Verify / Check Unit Calibration to start the Verification procedure. 3. Select the channel. The Sensor type is automatically detected. 4. Enter the user name and press start. 5. Connect the Capacitance Simulator and press Test Pre-amplifier Calibration to run the Verification procedure (Figure 51). The successful Verification is indicated. If this procedure fails it is recommended to perform a calibration. Figure 51: Testing the pre-amplifier calibration. 6. Connect the Incyte Sensor to the Pre-amplifier. 7. Select the Manual temperature correction mode. The Automatic mode is disabled. 8. Pour Hamilton 12880 μS/cm Conductivity Standard in a clean beaker. Make sure the sensor mounting instructions are followed (Figure 22). 9. Measure the temperature of the Standard with an external thermometer. Enter the conductivity value according to the temperature correction table on the bottle of the Conductivity Standard. 10. Press Test Sensor Calibration to carry on with the Verification (Figure 52). The successful Verification is indicated. If this procedure fails it is recommended to perform a calibration (see Chapter 10.6.1). Figure 52: Checking the Sensor calibration. 10.5.2 Verification of a Dencytee Sensor Unit ATTENTION! Make sure that the Val/Cal Solution (Ref 243886) is not expired and avoid contamination or dilution. Mix well prior to usage. 1. Connect a Dencytee Sensor Unit and power on the Control Unit. Equilibrate the Val/Cal Solution (Ref 243886) at 23°C +/- 1°C for at least 30 min prior to starting the Verification procedure. 83Cell Density Monitoring System Operating Instructions 2. Go to the Main Menu and enter the Calibration Menu (Figure 13). Select the Verify / Check Unit Calibration to start the Verification procedure. 3. Select the channel. The Sensor type is automatically detected. 4. Enter the user name and press start. 5. Make sure that the light path is clean, dust-free and placed in the dark. Press Measure in air to run the Verification procedure (Figure 53). 6. Insert the sensor directly in the bottle of the Val/Cal solution. Make sure that the light path is well immersed and placed in the dark. Press Measure in solution (Figure 53). The successful Verification is indicated. If this procedure fails it is recommended to perform a calibration (see Chapter 10.6.2). Figure 53: Testing the pre-amplifier calibration. 10.6 Calibration of a Sensor Unit NOTE: Calibration just available on screen if Cell Density Integration Kit option not activated. Calibration of a Sensor Unit is required if the Sensor Unit Verification did not passed success- fully. MAINTENANCEMAINTENANCE 10.6.1 Calibrate an Incyte Sensor Unit The Calibration procedure of an Incyte Sensor Unit is a five-step guided procedure. It takes about an hour to perform this procedure. The Val/Cal Kit for Standard Sensors (Ref 243740), Low Conductivity "LC" (Ref 243713) or High Conductivity "HC" (Ref 243714) is needed to perform a Incyte Sensor Unit Calibration. ATTENTION! Do not disconnect the Unit from the Arc View Controller, or the Sensor from the pre-amplifier during the Calibration procedure. NOTE: Make sure to prepare following equipment in addition to the Val/Cal Kit: stirrer-plate, thermometer 1. Connect the Incyte Unit, power on the Control Unit and equilibrate for at least 30 minutes prior to starting the Calibration procedure. 2. Go to the Main Menu and enter the Calibration Menu (Figure 13). Select Calibrate Unit to run the Calibration procedure. 3. Select the channel. The Sensor type is automatically detected. 4. Enter the user name and press start (Figure 54). Figure 54: Start of the Sensor Unit calibration. 85Cell Density Monitoring System Operating Instructions 5. Connect the Capacitance Simulator and press Test Pre-amplifier Calibration to proceed with the Calibration procedure. The successful test is indicated. If this procedure fails it is recommended to contact the local representative. 6. Replace the Simulator with the Incyte Sensor and press Measure in air to calibrate at zero conductivity. If this procedure fails it is recommended to contact the local representative. 7. Select Manual temperature correction mode. The Automatic mode is disabled. 8. Pour Hamilton 12880 μS/cm Conductivity Standard in a clean beaker. Make sure the sensor mounting instructions are followed (Figure 22). 9. Measure the temperature of the Standard with an external thermometer. Enter the conductivity value according to the temperature correction table on the bottle of the Conductivity Standard. 10. Press Calibrate sensor constant to start the Calibration (Figure 55). The successful Calibration is indicated. If this procedure fails it is recommended to contact the local representative. Figure 55: Calibration of the Sensor constant. 11. Rinse the Sensor with deionized water. 12. Pour 150 mL deionized water in a clean beaker. In the following steps 1 to 30 mL of Solution B is added gradually to the deionized water. Press start (Figure 56). 13. Add the required amount of Solution B. Press Start. MAINTENANCEMAINTENANCE Figure 56: Incyte Sensor calibration. 15. The successful Calibration is indicated. If this procedure fails it is recommended to contact the local representative. 16. Choose the directory, where the Calibration Report should be stored and press Generate Calibration Report (Figure 57). The valid calibration is automatically saved in the memory of the Pre-amplifier. Figure 57: Generation of a Calibration Report. 87Cell Density Monitoring System Operating Instructions MAINTENANCEMAINTENANCE 10.6.2 Calibrate a Dencytee Sensor Unit ATTENTION! Make sure that the Val/Cal Solution (Ref 243886) is not expired and avoid contamination or dilution. Mix well prior to usage. 1. Connect a Dencytee Sensor Unit and power on the Control Unit. Equilibrate the Val/Cal Solution (Ref 243886) at 23°C +/- 1°C for at least 30 min prior to starting the Calibration procedure. 2. Go to the Main Menu and enter the Calibration Menu (Figure 13). Select Calibrate Unit to run the Calibration procedure. 3. Select the channel. The Sensor type is automatically detected. 4. Enter the user name. 5. Make sure that the light path is clean, dust-free and placed in the dark. Press Measure in air to proceed with the Calibration procedure (Figure 53). 6. Insert the sensor directly in the bottle of the Val/Cal solution. Make sure that the light path is well immersed and placed in the dark. Press Measure in solution (Figure 53). 7. The successful Calibration is indicated. If this procedure fails it is recommended to contact the local representative. 8. Choose the directory, where the Calibration Report should be stored and press Generate Calibration Report (Figure 57). The valid calibration is automatically saved in the memory of the Pre-amplifier. 10.7 Transfer a Calibration Files to the Arc View Controller Upon request the calibration of a new Sensor Unit may be done at Hamilton. This new Unit is shipped with the calibration data on an USB stick. It is possible to transfer this Calibration File to the Arc View Controller without connecting the Sensor Unit. 1. Go to the Main Menu and enter the Calibration Menu (Figure 13). Select Load Calibration Files (Figure 58). 2. Connect the USB stick to the Control Unit (Figure 6, 7, 8) and choose the directory. Press Load file. Figure 58: Upload of calibration files. 10.7.1 Clean and Back Up the Arc View Controller To keep the storage space of the Control Unit free, the temporary data may be transferred to an USB stick and deleted. 1. Export the temporary data (see Chapter 9.6.3). 2. Delete the temporary files (see Chapter 9.6.2). 10.7.2 Update the Control Unit 1. Please check our Homepage www.hamiltoncompany.com/arc_view_update for recent updates at least once a year. 2. Update the Control Unit (see Chapter 9.7.1 and 9.7.2). 89Cell Density Monitoring System Operating Instructions 11 Troubleshooting 11.1 Self-Diagnostic of the Sensor Unit The Sensor Unit is running a self-diagnosis continuously. The Status Indicator on the Control Unit (Figure 12). and the Incyte Status LED on the Incyte Pre-amplifier (Figure 9) report direct feedback. The feedback output is similar to a traffic light. A green Status Indicator means everything works as excepted. An orange Status Indicator shows a warning, that the measurement may be not correct. A red indication gives an error that the system has stopped working (Table 1). Table 1: Status Table. 11.2 Re-Set the Factory Settings It is possible to re-set the factory settings, in case of Software failure. See Chapter 9.7.2. 11.3 Verification and Calibration Error 11.3.1 Verification Error It is recommended to perform a Calibration (Chapter 10.6), if the Verification of an Incyte or Dencytee Unit fails (see Chapter 10.5). 11.3.2 Calibration Error If a calibration of a Dencytee Unit fails (see Chapter 10.6.2) re-start the Control Unit and perform a Calibration again. Please contact your local representative, if the Calibration fails again after re-start of the Controller. Please contact your local representative, if the Calibration of an Incyte Unit fails. TROUBLESHOOTINGTROUBLESHOOTING Notification on the Status Status LED Description Control Unit Indicator (Incyte) SENSOR: ok Green Green The Sensor Unit is working as expected. Simulator connected Green Green The permittivity simulator is connected to the Pre-amplifier. SENSOR: cleaning state Green Flashing green The Pre-amplifier is applying a cleaning cycle, no measurement available. AMPLIFIER: temperature Green Flashing green The Pre-amplifier temperature is security activated too high, the security fuse is activated, no measurement available. SENSOR: Out of range Green Flashing green The measurement is out of range. measurement AMPLIFIER: not found Gray Green No communication between Sensor Unit and Arc View Controller, please contact the local representative. SENSOR: sterilization Orange Orange The Sensor has been sterilized more count overflow than 100 times. Please contact your local representative. AMPLIFIER : total Orange Orange The Pre-amplifier has been used duration of use overflow more than 6 months continuously. Please perform a verification. SENSOR: invalid Orange Orange The current sensor has not been sensor calibration calibrated with the Pre-amplifier. Notification on the Status Status LED Description Control Unit Indicator (Incyte) SENSOR: invalid Orange Orange The sensor serial number is not valid. calibration SENSOR: total Orange Orange The sensor has been used more than duration overflow 6 months continuously. Please perform a verification. SENSOR: out of Orange Orange The measurement is out of the calibration range calibration range. SENSOR: not found Red Red No Sensor connected to the Pre-amplifier. BOX: not connected Grey Grey ComBox is not connected to the PC. 91Cell Density Monitoring System Operating Instructions 11.4 Getting Technical Support If any troubles are observed, which you cannot solve based on this operation instruction, please contact Hamilton Technical Support for further help: techsupport.pa.ch@hamilton.ch or the local Hamilton representative. For diagnosis of the measurement loop further data may be needed, it is explained in chapters 11.4.1 to 11.4.3 to get the relevant information. 11.4.1 Create Print Screens NOTE: Integrated print screen functionality only available with Arc View Controller. It is possible to create print screens on certain unexpected events. 1. Press the Print Screen Button in the Run Menu (Figure 11) The image is automatically recorded in the home directory. 2. Transfer the Screen Shoot to a USB stick (see Chapter 7.10 and 8.10). 11.4.2 System Information The System information gives information about the current Soft- and Hardware version of the complete Measuring Loop. 1. Press the About Button in the Main Manu (Figure 13) and press print screen (Figure 59). 2. Transfer the Screen Shoot to a USB stick (see Chapter 7.10 and 8.10). TROUBLESHOOTINGTROUBLESHOOTING Figure 59: General information menu of the Arc View Controller. 11.4.3 System Data The log files contain the relevant information about the measurement loop, including error codes. Please send the log files to the technical support for trouble shooting. See Chapter 9.6.1 11.5 Remove Corrupt Files In case of Software failure the deletion of corrupt files may be required. NOTE: Connect a keyboard to the USB connector of the Arc View Controller prior to system start. NOTE: A USB stick is required to remove the files from the Arc View Controller and store them for safety reasons, to ensure minimal loss of data. 93Cell Density Monitoring System Operating Instructions ATTENTION! Remove the files stepwise and run a system start after each step to confirm if the Software is working stable: • move temporary files (stored culture files and calibration/verification reports) • move log files (recent/running culture files) • move configuration files (all configuration data and settings) - this requires the new set up of channel initialization, the OPC or 4- 20 mA communication 1. Switch on the Arc View Controller 2. Select Evo Recovery with the arrow key, as soon as the possibility is displayed, the software will now boot in the recovery mode and shows a selection menu. 3. Select possibility «4. File recovery» by typing 4 and confirming with the enter key 4. Enter the administrator password and confirm with enter. A new selection menu will be shown. 5. Select the possibilities stepwise and confirm with enter, by skipping step 4 as, as this is only required to transfer error-logs (please see chapter 11.6). 6. Wait until the procedure is finished and the system has re-started 7. In case the Software is not running stabile perform the next step NOTE: The data is saved on the USB stick in the folder: Arc-View_backup 11.6 Transfer Error-Log-Files from the Arc View Controller If case the Software fails a tool will create Error-Log-Files. These files are relevant information for the Technical Support. NOTE: Connect a keyboard to the USB connector of the Arc View Controller prior to system start. TROUBLESHOOTINGTROUBLESHOOTING NOTE: A USB stick is required to remove the files from the Arc View Controller. 1. Switch on the Arc View Controller. 2. Select Evo Recovery with the arrow key, as soon as the possibility is displayed, the software will now boot in the recovery mode and shows a selection menu. 3. Select possibility «4. File recovery» by typing 4 and confirming with the enter key. 4. Enter the administrator password and confirm with enter. A new selection menu will be shown. 5. Select «4: Move crash reports» by typing 4 and confirming with the enter key. 6. Wait until the procedure is finished and the system has re-started. 7. Remove the USB drive. The error report is now located on the USB drive and is a file ending with «.dmp», like «interface_Biomass_rack_V2.exe.X.dmp», whereas X may be a 4 or 5 digit number. There may be up to two files like this. 11.7 Returning a System for Repair Please contact Hamilton Technical Support for return of the Instrument. You have to assign a RGA-number clearly written on the packaging. Please be aware that we are not responsible for any damages during transportation, if the material is not shipped in original wrapping and suitable protected. NOTE: Please ensure that the material is properly decontaminated prior to wrapping. 95Cell Density Monitoring System Operating Instructions 12 Connection of the Control Unit to the Process Control System In this chapter it is explained, how to connect the Control Unit to the process control system by: • Modbus communication • OPC communication • 4-20 mA Output signal 12.1 Connect by Modbus communication 12.1.1 Preparation of the Control Unit NOTE: The Control Unit acts as Modbus slave (Modbus server), all cell density measurements can be read by a Modbus master (client). The present implementation of the Modbus protocol uses the RTU mode over a RS-485 serial line. 1. Enter the Advanced Settings and select the Modbus Configuration Figure 60: Menu for Modbus Configuration. 2. Connect an open-end M12 cable to the Modbus Connector (Figure 61) on the right side of the Controller and configure the Modbus parameters (Figure 60). Pin Function 1 +24 V 2 RS-485 A 3 GND 4 RS-485 B 5 n.c. 2 4 31 5 Figure 61: Female M12 connector description. 12.1.2 Preparation of the Modbus Connection 12.1.2.1 Definition of the Return Time The return time depends on the type of Modbus requests. Modbus request Reading Writing one register or multiple Writing the same register or registers once (next write must the same multiple registers be on different registers) twice in a row Return time 200ms 200ms 1s Table 2: Modbus return time. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 97Cell Density Monitoring System Operating Instructions 12.1.2.2 Overview of the Available Data The Control Unit provides for each channel three sets of information on Modbus. Standard cell density data: • Sensor SN • Amplifier SN • Amplifier Status (integer type format) • Conductivity • Cell density • Measure index • Amplifier status (float type format) Incyte Scan data: • Delta Epsilon • Fc • Alpha • Fscan Epsilon (not available with standard licenses) • Spectro index Capacitance at all frequencies of the Incyte Scan: • Fscan - Index • Scan frequencies list (frequency) • Scan C(f) list (permitivity) Each one of the three data sets (standard cell density, Incyte Scan data, permittivity at all frequencies of the Incyte Scan) has its own index which is incremented at each valid measurement. An index increment stopped for a long time means there is no communication between the Arc View Controller and amplifier, e.g. if the pre-amplifier is not connected. To preserve data integrity, the Modbus master must request the entire data set in only one frame. The system guarantees the data set index tallies with only one measurement. NOTE: If the master reads data separately with an undefined time between requests, the Arc View Controller could change registers during this time and so the index could not tally with only a single measurement. 12.1.2.3 Definition of 32 Bit Data Type The Modbus protocol defines registers of 16 bit words. The registers are transmitted in big-endian format. The float values and unsigned long values are 32 bit sized; therefore they are mapped into two registers. These 32 bit values are transmitted low word (16 bit) first whereas the bytes inside the words are sent with big-endian format. For example: • Floating point representation: 97.18467 • Hexadecimal representation 0x42C25E8D (IEEE 754 Standard) • The float value in the register table: Registers Value 86 0x5E8D 87 0x42C2 Table 3: Register Values. Modbus answer frame (slave address = 1): • 0x01 | 0x04 | 0x5E | 0x8D | 0x42 | 0xC2 | 0xC9 | 0x76 Activate the byte swap option in the Modbus configuration screen allow to swap the byte in the word. The answer frame with this option is: • 0x01 | 0x04 | 0x8D | 0x5E | 0xC2 | 0x42 | 0xC9 | 0x76 CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 99Cell Density Monitoring System Operating Instructions 12.1.2.4 Definition of Modbus Exceptions If the slave cannot serve a request, it returns an exception with the following error codes: Error code Name Description 0x01 Illegal Function The function code received in the query is not an allowable action for the slave. 0x02 Illegal Data Address The data address received in the query is not an allowable address for the slave. 0x03 Illegal Data Value A value contained in the query data field is not an allowable value for the slave. 0x06 Slave Device Busy The slave is engaged in processing a long-duration program command and transmits the message later when the slave is free. Table 4: Definitions of Modbus Exceptions. 12.1.2.5 Pre-Amplifier Status Codes The following table lists the possible status code for Modbus registers amplifier status. Table 5: Pre-amplifier Status Code. NOTE: Some status codes are only available for Incyte Sensors. Status Code Amplifier led Description indicator SENSOR: not found 0 Red No sensor connected to the Pre-amplifier. SENSOR: sterilization 1 Orange The sensor has been sterilized more count overflow than 100 times. Please contact the local representative. AMPLIFIER : total duration 4 Orange The preamplifier has been used more of use overflow than 1 year continuously. Please perform a Pre-amplifier check. PROBE: invalid sensor 5 Orange The current sensor has not been calibration calibrated with the Pre-amplifier. SENSOR: invalid calibration 6 Orange The current sensor has not been calibrated with the Pre-amplifier. The sensor serial number is not valid. SENSOR: total duration 7 Orange The sensor has been used more than overflow 6 months continuously. Please perform a verification. PROBE: out of calibration 8 Orange The measurement is out of the range calibration range. PROBE: cleaning state 9 Flashing green The Pre-amplifier is applying specific voltage to clean the sensor, no measurement available. AMPLIFIER: temperature 10 Flashing green The Pre-amplifier temperature is too security activated high, the security fuse is activated, no measurement available. PROBE: Out of range 11 Flashing green The measurement is out of range. measurement PROBE: ok 12 Green The unit is working as expected. Biomass simulator connected 13 Green The Simulator is connected to the Pre-amplifier. SENSOR: total sterilization 14 Orange The automatic sterilization counter has time overflow reached the maximum value (100). SENSOR: overtime 15 Orange The sensor has been used more than reached-calib/check needed 1 year since the last calibration. Please perform a calibration. AMPLIFIER: overtime 16 Orange The Pre-Amplifier has been used more reached/check needed than 1 year since the last calibration Please perform a calibration. AMPLIFIER: not found 17 Off No Pre-Amplifier is connected to the channel. bootloader 18 Red flashing Pre-Amplifier firmware updating. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 101Cell Density Monitoring System Operating Instructions 12.1.2.6 Calibration and Validation via Modbus Dencytee and Incyte units can be calibrated and validated via Modbus if the Cell Density Integration Kit mode is enabled (see Chapter 9.2.6). Since calibration and validation are both procedures with multiple steps involved, an individual state machine that can be controlled via Modbus is used for the four use cases: • Calibration of Incyte Unit • Validation of Incyte Unit • Calibration of Dencytee Unit • Validation of Dencytee Unit 12.1.2.6.1 General Concepts 12.1.2.6.1.1 Controlling the State Machine via Modbus The state machine of one measurement channel can be controlled using four Modbus registers. • Parameter: Register for passing an input argument to the state machine (e.g. nominal conductivity of currently used calibration standard) • Command: Register for command codes that trigger state transitions in the state machine. • Acknowledge: Status register controlled by the software in order to signal the state of processing the command. • State: Register that returns the current state of the state machine A state transition involves the following typical steps: • Modbus Master writes the parameter (if required) • Modbus Master writes the command; the software confirms reception of the command by setting Acknowledge to 0. • Modbus Master resets the command to 0 in order to start execution of the command in the software. • Once execution of a command is finished, the software writes the new state to the State register and resets the Acknowledge register to 1 in order to indicate readiness for a new command. Parameter Modbus Master Command Modbus Master 0 Acknoledge Control Unit 0 State Control Unit COMMAND EXECUTION IN BIOMASS NEW VALUE P NEW VALUE S NEW VALUE A Default value for P is 0 / Default value for A is 1 NEW VALUE C 1 CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 103Cell Density Monitoring System Operating Instructions The main state machine for every single measurement channel looks as depicted in this chapter. If CDIK mode is switched off when starting the software, the state machine will transition automatically to state 8 “not supported”. If CDIK (Cell Density Integration Kit) mode is switched on when starting the software, the state machine will be in state 1 “Initial”. At any time only one calibration or validation process on one channel can be active. As soon as a calibration or validation on any measurement channel is started, the state machines of the other channels will automatically transition to state 9 “Busy with another channel”. Every calibration / validation has a specific sub-state machine that will be described in chapters 12.1.2.6.2 – 12.1.2.6.5. When the calibration or validation is completed the state machine of the corresponding measurement channel will be in state 2 “Final”. After resetting the state machine to state 1 “Initial” the state machines of the other measurement channels will transition automatically from state 9 “Busy” to 1 “Initial”. State Command Next State Description 1 1 10 Start a calibration or validation. The command code is always 1, the sensor type and the action (validation or calibration) is specified by selecting the corresponding Modbus register (see Chapter 12.1.2.6.1.3). Make sure the sensor is connected to the Pre-amplifier 2 2 1 Reset the state machine to the initial state in order to resume normal operation 8 – Not Supported: CDIK mode is switched off 9 – Busy: A calibration/validation is active on another measurement channel 12.1.2.6.1.3 Modbus registers for CDIK, Arc View Controller 265 and PC software The following table defines the Modbus registers used for controlling the state machines for the two measurement channels depending on the type of sensor and action (calibration or validation). Note that Parameter and Command have to be read and written as Modbus Holding Registers (Modbus function codes 0x03 and 0x10). Acknowledge and State have to be read as Modbus Input Registers (Modbus function code 0x04). 12.1.2.6.1.2 The main state machine State 8 CDIK not supported CDIK not supported CDIKsupported State 1 Initial Statemachine «Calibration Incyte» Statemachine «Validation Incyte» Statemachine «Calibration Dencytee» Statemachine «Validation Dencytee» Command1 Command1 Command1 Command1 Command2 State 2 Final State 9 Busy with another channel Any other channel is not in State Initial All other channels are in State Initial CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 105Cell Density Monitoring System Operating Instructions Parameter Command Acknowledge State (Holding Reg.) (Holding Reg.) (Input Reg.) (Input Reg.) Channel 3 Validation Incyte 44 43 130 131 Calibration Incyte 46 45 132 133 Validation Dencytee 114 113 824 825 Calibration Dencytee 116 115 826 827 Channel 4 Validation Incyte 60 59 176 177 Calibration Incyte 62 61 178 179 Validation Dencytee 130 129 871 872 Calibration Dencytee 132 131 873 874 12.1.2.6.1.5 General error states During calibration and validation errors will be indicated by a transition to an error state. Error states have a state number of 900 and above. With command 3 a transition to state 2 “Final” is possible. In any state if an invalid command is invoked the state machine transitions to state 999 “Invalid command”. State Command Next State Description 9xx 3 2 States numbered from 900 to 999 are error states. The command 3 “Cancel” transitions from any error state to the state 2 “Final”. Some error states allow repeating the step that has led to the error. This will be mentioned in the corresponding error state description 900 3 2 Wrong sensor type 997 3 2 Error while resetting the sensor or preamplifier 998 3 2 Generic error while communicating with the sensor / preamplifier 999 3 2 Invalid command Parameter Command Acknowledge State (Holding Reg.) (Holding Reg.) (Input Reg.) (Input Reg.) Channel 1 Validation Incyte 12 11 38 39 Calibration Incyte 14 13 40 41 Validation Dencytee 44 43 388 389 Calibration Dencytee 46 45 390 391 Channel 2 Validation Incyte 28 27 84 85 Calibration Incyte 30 29 86 87 Validation Dencytee 60 59 435 436 Calibration Dencytee 62 61 437 438 12.1.2.6.1.4 Modbus registers for Arc View Controller 465 The following table defines the Modbus registers used for controlling the state machines for the four measurement channels depending on the type of sensor and action (calibration or validation). Note that Parameter and Command have to be read and written as Modbus Holding Registers (Modbus function codes 0x03 and 0x10). Acknowledge and State have to be read as Modbus Input Registers (Modbus function code 0x04). Parameter Command Acknowledge State (Holding Reg.) (Holding Reg.) (Input Reg.) (Input Reg.) Channel 1 Validation Incyte 12 11 38 39 Calibration Incyte 14 13 40 41 Validation Dencytee 82 81 730 731 Calibration Dencytee 84 83 732 733 Channel 2 Validation Incyte 28 27 84 85 Calibration Incyte 30 29 86 87 Validation Dencytee 98 97 777 778 Calibration Dencytee 100 99 779 780 CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 107Cell Density Monitoring System Operating Instructions 12.1.2.6.2.1 Table of states State Command Next State Description 10 10 20 Temperature stabilization of the preamplifier 20 User has to connect the Incyte simulator to the preamplifer 20 20 30 Calibration check of the preamplifier 30 User disconnects the Incyte simulator and connects the Incyte sensor. The sensor is immersed in solution A. The conductivity of solution A at the current temperature has to be determined using the table on the bottle 30 30 2 Check sensor calibration Parameter: conductivity of solution A at current temperature The conductivity has to be written as a 16 Bit integer value in the unit µS/cm to the corresponding Modbus register 2 The user removes the sensor from Solution A 12.1.2.6.2.2 Error states State Command Next State Description 901 3 2 Error while reading the preamplifier’s temperature 902 3 2 Preamplifier temperature is out of specified range 903 3 2 Preamplifier temperature is not stable 904 3 2 Preamplifier calibration is invalid 905 3 2 Sensor/Preamplifier is not calibrated 920 3 2 Preamplifier/sensor calibration is invalid 12.1.2.6.2 Validation Incyte Unit No Incyte sensor connected State 10 Ready for temperature stabilization Failure Execute preamplifier reset Execute temperature stabilization Error states according table Failure State 20 Ready for check preamplifier calibration User connects the simulator Preamplifier calibration check is performed Failure State 30 Ready for check sensor calibration Sensor calibration check is performed User disconnects the simulator User connects the sensor and immerses it in solution A Failure Failure Execute preamplifier reset CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 109Cell Density Monitoring System Operating Instructions 12.1.2.6.3.1 Table of states State Command Next State Description 10 10 20 Temperature stabilization of the preamplifier. Acknowledge: The type of calibration is returned 1 = Standard 2 = High conductivity 3 = Low conductivity 20 20 30 Initialize Preamplifier 30 User has to connect the Incyte simulator to the preamplifer 30 30 40 Calibration check of the preamplifier 40 User disconnects the Incyte simulator and connects the Incyte sensor 40 40 50 Calibrate zero conductivity in air 50 The sensor is immersed in solution A. The conductivity of solution A at the current temperature has to be determined using the table on the bottle 50 50 110 Calibration of cell constant Parameter: conductivity of solution A at current temperature The conductivity has to be written as a 16 Bit integer value in the unit µS/cm to the corresponding Modbus register 110 The user immerses the sensor in de-ionized water and adds solution B according the calibration type (see 12.1.2.6.3.2) 110 110 120 Calibration of dilution step 1. If this step fails, it can be repeated (see 12.1.2.6.3.3) 120 User adds solution B according the calibration type (see 12.1.2.6.3.2) 120 120 130 Calibration of dilution step 2. If this step fails, it can be repeated (see 12.1.2.6.3.3) 130 User adds solution B according the calibration type (see 12.1.2.6.3.2) … 180 User adds solution B according the calibration type (see 12.1.2.6.3.2) 180 180 200 Calibration of dilution step 8. If this step fails, it can be repeated (see 12.1.2.6.3.3) 200 The user removes the sensor from Solution B 200 200 2 Calibration file is created 12.1.2.6.3 Calibration Incyte Unit No Incyte sensor connected State 10 Ready for temperature stabilization Failure Execute preamplifier reset Execute temperature stabilization Error states according table Failure State 20 Ready for preamplifier initialization Failure State 30 Ready for preamplifier calibration check Preamplifier calibration check is performed User connects the simulator Failure Failure Create calibration file State 40 Ready for zero conductivity calibration Zero conductivity calibration is performed User disconnects the simulator and connects the sensor State 50 Ready for cell constant calibration Cell constant calibration is performed Failure State 1n0 Ready for dilution step n User immerses the sensor in solution A User immerses the sensor in dilution n Failure State 200 Ready for creating calibration file n<=8 n=n+1 Failure Execute preamplifier reset Failure Cancel Cancel States 91n Failure in dilution step n CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 111Cell Density Monitoring System Operating Instructions 12.1.2.6.3.3 Error states State Command Next State Description 901 3 2 Error while reading the preamplifier’s temperature 902 3 2 Preamplifier temperature is out of specified range 903 3 2 Preamplifier temperature is not stable 904 3 2 Preamplifier is not calibrated or calibration is invalid 905 3 2 Error while initializing sensor 906 3 2 Error while calibrating zero conductivity 907 3 2 Conductivity out of range 908 3 2 Calibration in conductivity standard failed 909 3 2 Failure in dilution step 1. Incyte calibration can be cancelled with command 3. Dilution step can be 4 110 repeated with command 4 910 3 2 Failure in dilution step 2. Incyte calibration can be cancelled with command 3. Dilution step can be 4 120 repeated with command 4 911 3 2 Failure in dilution step 3. Incyte calibration can be cancelled with command 3. Dilution step can be 4 130 repeated with command 4 912 3 2 Failure in dilution step 4. Incyte calibration can be cancelled with command 3. Dilution step can be 4 140 repeated with command 4 913 3 2 Failure in dilution step 5. Incyte calibration can be cancelled with command 3. Dilution step can be 4 150 repeated with command 4 914 3 2 Failure in dilution step 6. Incyte calibration can be cancelled with command 3. Dilution step can be 4 160 repeated with command 4 915 3 2 Failure in dilution step 7. Incyte calibration can be cancelled with command 3. Dilution step can be 4 170 repeated with command 4 916 3 2 Failure in dilution step 8. Incyte calibration can be cancelled with command 3. Dilution step can be 4 180 repeated with command 4 917 3 2 Conductivity is not stable 918 3 2 Error while writing calibration file 12.1.2.6.3.2 Dilution steps for different calibration types Part of the Incyte calibration is a series of measurements in different conductivity dilutions. Solution B is added to a beaker with 150ml de-ionized water in 8 steps according the following table. The dose is specific for the calibration type of the Incyte unit (can be read in state 10, see Chapter 12.1.2.6.3.1) Addition of Solution B per dilution step [ml] Standard High conductivity Low conductivity calibration calibration calibration Dilution step 1 1 1 0.5 Dilution step 2 1 2 1 Dilution step 3 2 2 1.5 Dilution step 4 5 5 2 Dilution step 5 10 15 5 Dilution step 6 20 20 10 Dilution step 7 20 30 10 Dilution step 8 30 40 20 CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 113Cell Density Monitoring System Operating Instructions 12.1.2.6.4.1 Table of states State Command Next State Description 10 10 20 Temperature stabilization of the preamplifier 20 Make sure the light path is clean, dust-free and dark 20 20 30 Measurement in air 30 User immerses the sensor in the Val/Cal Solution OD 30 30 2 Measurement in Val/Cal Solution 2 User removes the sensor from Val/Cal Solution 12.1.2.6.4.2 Error states State Command Next State Description 920 3 2 Error while reading the preamplifier’s temperature 921 3 2 Preamplifier temperature is out of specified range 922 3 2 Preamplifier temperature is not stable 923 3 2 External interference light detected 928 3 2 Error while measuring in air 929 3 2 Error while measuring in Val/Cal solution 930 3 2 Preamplifier/sensor calibration is invalid 12.1.2.6.4 Validation Dencytee No Dencytee sensor connected State 10 Ready for temperature stabilization Failure Execute preamplifier reset Execute temperature stabilization Error states according table Failure State 20 Ready for measurement in air Measurement in air is performed Failure State 30 Ready for measurement in Val/Cal Solution Measurement in Val/Cal Solution is performed Failure Failure Execute preamplifier reset Cancel User immerses the sensor in Val/Cal Solution CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 115Cell Density Monitoring System Operating Instructions 12.1.2.6.5.1 Table of states State Command Next State Description 10 10 20 Temperature stabilization of the preamplifier 20 20 30 Initialize Preamplifier 30 Make sure the light path is clean, dust-free and dark 30 30 40 Calibration in air 40 User immerses the sensor in the Val/Cal Solution OD 40 40 50 Calibration in Val/Cal Solution 50 User removes the sensor from Val/Cal Solution 50 50 2 Calibration file is created 12.1.2.6.5.2 Error states State Command Next State Description 920 3 2 Error while reading the preamplifier’s temperature 921 3 2 Preamplifier temperature is out of specified range 922 3 2 Preamplifier temperature is not stable 923 3 2 External interference light detected 924 3 2 Error while initializing sensor 925 3 2 Error while calibrating in air 926 3 2 Error while calibrating in Val/Cal Solution 927 3 2 Error while writing calibration file 12.1.2.6.5 Calibration Dencytee State 40 Ready for calibration in Val/Cal solution No Dencytee sensor connected State 10 Ready for temperature stabilization Failure Execute preamplifier reset Execute temperature stabilization Error states according table Failure State 20 Ready for preamplifier initialization Failure State 30 Ready for calibration in air Calibration in air is performed Failure Failure Execute preamplifier reset Calibration in Val/Cal solution is performed State 50 Ready for creating calibration file Failure Create calibration file Failure User immerses the sensor in Val/Cal Solution CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 117Cell Density Monitoring System Operating Instructions Function Register Unit Data type Data Size address (registers) Gateway serial number 001   string 6 Measures - Ch 1         Sensor serial number 007   string 6 Amplifier serial number 013   string 6 Amplifier status 019   word 16bits 1 Conductivity 020 mS/cm float32 2 Permittivity 022 pF/cm float32 2 Measure Index 024   unsigned long 2 Amplifier Status - float format 026   float32 2 Reserved 028 float32 2 Reserved 030 float32 2 Reserved 032 float32 2 Reserved 034 float32 2 Reserved 036 float32 2 Acknowledge Val. Incyte 038   word 16bits 1 State Val. Incyte 039   word 16bits 1 Acknowledge Cal. Incyte 040   word 16bits 1 State Cal. Incyte 041   word 16bits 1 Spectro - Ch 1         DeltaEpsilon 042   float32 2 Fc 044   float32 2 Alpha 046   float32 2 FscanEpsilon 048   float32 2 Spectro index 050   unsigned long 2 Reserved 052 Word 16bits 1 Measures - Ch 2         Sensor serial number 053   string 6 Amplifier serial number 059   string 6 Amplifier status 065   word 16bits 1 Conductivity 066 mS/cm float32 2 Permittivity 068 pF/cm float32 2 Measure Index 070   unsigned long 2 Amplifier Status - float format 072   float32 2 Reserved 074 float32 2 Reserved 076 float32 2 Reserved 078 float32 2 Reserved 080 float32 2 Reserved 082 float32 2 Acknowledge Val. Incyte 084   word 16bits 1 State Val. Incyte 085   word 16bits 1 Acknowledge Cal. Incyte 086   word 16bits 1 State Cal. Incyte 087   word 16bits 1 Spectro - Ch 2         DeltaEpsilon 088   float32 2 Fc 090   float32 2 Alpha 092   float32 2 FscanEpsilon 094   float32 2 Spectro index 096   float32 / ul 2 Reserved 098   word 16bits 1 Info-Fscan-Ch1         Reserved 99   word 16bits 1 Function Register Unit Data type Data Size address (registers) 12.1.2.7 Modbus Tables Arc View Controller 265, ComBox and Cell Density Integration Kit 12.1.2.7.1 Input Registers Tables (Read only) 12.1.2.7.1.1 Read the Modbus Commands All measurements and readable information which comes from the sensors are memorized in the Modbus input registers table. The Modbus function code 0x04 is used to read this table. It allows reading 1 to 125 contiguous input registers from the Control Unit. The request specifies the starting address and the number of registers to be read. In the Modbus frame, registers are addressed starting at zero. For example, input registers numbered 1-16 are addressed as 0-15 inside the frame. 12.1.2.7.1.2 Incyte Input Registers Table CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 119Cell Density Monitoring System Operating Instructions Reserved 100   word 16bits 1 Reserved 101   word 16bits 1 index-Fscan-ch1 102   unsigned long 2 Freq-FSCAN-Ch1(30 points)         Scan-Freq-Pt0 104 Hz float 32 bits 2 Scan-Freq-Pt1 106 Hz float 32 bits 2 Scan-Freq-Pt2 108 Hz float 32 bits 2 Scan-Freq-Pt3 110 Hz float 32 bits 2 Scan-Freq-Pt4 112 Hz float 32 bits 2 Scan-Freq-Pt5 114 Hz float 32 bits 2 Scan-Freq-Pt6 116 Hz float 32 bits 2 Scan-Freq-Pt7 118 Hz float 32 bits 2 Scan-Freq-Pt8 120 Hz float 32 bits 2 Scan-Freq-Pt9 122 Hz float 32 bits 2 Scan-Freq-Pt10 124 Hz float 32 bits 2 Scan-Freq-Pt11 126 Hz float 32 bits 2 Scan-Freq-Pt12 128 Hz float 32 bits 2 Scan-Freq-Pt13 130 Hz float 32 bits 2 Scan-Freq-Pt14 132 Hz float 32 bits 2 Scan-Freq-Pt15 134 Hz float 32 bits 2 Scan-Freq-Pt16 136 Hz float 32 bits 2 Scan-Freq-Pt17 138 Hz float 32 bits 2 Scan-Freq-Pt18 140 Hz float 32 bits 2 Scan-Freq-Pt19 142 Hz float 32 bits 2 Scan-Freq-Pt20 144 Hz float 32 bits 2 Scan-Freq-Pt21 146 Hz float 32 bits 2 Scan-Freq-Pt22 148 Hz float 32 bits 2 Scan-Freq-Pt23 150 Hz float 32 bits 2 Scan-Freq-Pt24 152 Hz float 32 bits 2 Scan-Freq-Pt25 154 Hz float 32 bits 2 Scan-Freq-Pt26 156 Hz float 32 bits 2 Scan-Freq-Pt27 158 Hz float 32 bits 2 Scan-Freq-Pt28 160 Hz float 32 bits 2 Scan-Freq-Pt29 162 Hz float 32 bits 2 C(f)-FSCAN-Ch1(30 points)         Scan-C(f)-Pt0 164 pF/cm float 32 bits 2 Scan-C(f)-Pt1 166 pF/cm float 32 bits 2 Scan-C(f)-Pt2 168 pF/cm float 32 bits 2 Scan-C(f)-Pt3 170 pF/cm float 32 bits 2 Scan-C(f)-Pt4 172 pF/cm float 32 bits 2 Scan-C(f)-Pt5 174 pF/cm float 32 bits 2 Scan-C(f)-Pt6 176 pF/cm float 32 bits 2 Scan-C(f)-Pt7 178 pF/cm float 32 bits 2 Scan-C(f)-Pt8 180 pF/cm float 32 bits 2 Scan-C(f)-Pt9 182 pF/cm float 32 bits 2 Scan-C(f)-Pt10 184 pF/cm float 32 bits 2 Scan-C(f)-Pt11 186 pF/cm float 32 bits 2 Scan-C(f)-Pt12 188 pF/cm float 32 bits 2 Scan-C(f)-Pt13 190 pF/cm float 32 bits 2 Scan-C(f)-Pt14 192 pF/cm float 32 bits 2 Scan-C(f)-Pt15 194 pF/cm float 32 bits 2 Scan-C(f)-Pt16 196 pF/cm float 32 bits 2 Scan-C(f)-Pt17 198 pF/cm float 32 bits 2 Scan-C(f)-Pt18 200 pF/cm float 32 bits 2 Scan-C(f)-Pt19 202 pF/cm float 32 bits 2 Scan-C(f)-Pt20 204 pF/cm float 32 bits 2 Scan-C(f)-Pt21 206 pF/cm float 32 bits 2 Scan-C(f)-Pt22 208 pF/cm float 32 bits 2 Scan-C(f)-Pt23 210 pF/cm float 32 bits 2 Scan-C(f)-Pt24 212 pF/cm float 32 bits 2 Scan-C(f)-Pt25 214 pF/cm float 32 bits 2 Scan-C(f)-Pt26 216 pF/cm float 32 bits 2 Scan-C(f)-Pt27 218 pF/cm float 32 bits 2 Scan-C(f)-Pt28 220 pF/cm float 32 bits 2 Scan-C(f)-Pt29 222 pF/cm float 32 bits 2 Info-Fscan-Ch2       Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 121Cell Density Monitoring System Operating Instructions Reserved 224   word 16bits 1 Reserved 225   word 16bits 1 Reserved 226   word 16bits 1 index-Fscan-ch2 227   unsigned long 2 Freq-FSCAN-Ch2(30 points)         Scan-Freq-Pt0 229 Hz float 32 bits 2 Scan-Freq-Pt1 231 Hz float 32 bits 2 Scan-Freq-Pt2 233 Hz float 32 bits 2 Scan-Freq-Pt3 235 Hz float 32 bits 2 Scan-Freq-Pt4 237 Hz float 32 bits 2 Scan-Freq-Pt5 239 Hz float 32 bits 2 Scan-Freq-Pt6 241 Hz float 32 bits 2 Scan-Freq-Pt7 243 Hz float 32 bits 2 Scan-Freq-Pt8 245 Hz float 32 bits 2 Scan-Freq-Pt9 247 Hz float 32 bits 2 Scan-Freq-Pt10 249 Hz float 32 bits 2 Scan-Freq-Pt11 251 Hz float 32 bits 2 Scan-Freq-Pt12 253 Hz float 32 bits 2 Scan-Freq-Pt13 255 Hz float 32 bits 2 Scan-Freq-Pt14 257 Hz float 32 bits 2 Scan-Freq-Pt15 259 Hz float 32 bits 2 Scan-Freq-Pt16 261 Hz float 32 bits 2 Scan-Freq-Pt17 263 Hz float 32 bits 2 Scan-Freq-Pt18 265 Hz float 32 bits 2 Scan-Freq-Pt19 267 Hz float 32 bits 2 Scan-Freq-Pt20 269 Hz float 32 bits 2 Scan-Freq-Pt21 271 Hz float 32 bits 2 Scan-Freq-Pt22 273 Hz float 32 bits 2 Scan-Freq-Pt23 275 Hz float 32 bits 2 Scan-Freq-Pt24 277 Hz float 32 bits 2 Scan-Freq-Pt25 279 Hz float 32 bits 2 Scan-Freq-Pt26 281 Hz float 32 bits 2 Scan-Freq-Pt27 283 Hz float 32 bits 2 Scan-Freq-Pt28 285 Hz float 32 bits 2 Scan-Freq-Pt29 287 Hz float 32 bits 2 C(f)-FSCAN-Ch2(30 points)         Scan-C(f)-Pt0 289 pF/cm float 32 bits 2 Scan-C(f)-Pt1 291 pF/cm float 32 bits 2 Scan-C(f)-Pt2 293 pF/cm float 32 bits 2 Scan-C(f)-Pt3 295 pF/cm float 32 bits 2 Scan-C(f)-Pt4 297 pF/cm float 32 bits 2 Scan-C(f)-Pt5 299 pF/cm float 32 bits 2 Scan-C(f)-Pt6 301 pF/cm float 32 bits 2 Scan-C(f)-Pt7 303 pF/cm float 32 bits 2 Scan-C(f)-Pt8 305 pF/cm float 32 bits 2 Scan-C(f)-Pt9 307 pF/cm float 32 bits 2 Scan-C(f)-Pt10 309 pF/cm float 32 bits 2 Scan-C(f)-Pt11 311 pF/cm float 32 bits 2 Scan-C(f)-Pt12 313 pF/cm float 32 bits 2 Scan-C(f)-Pt13 315 pF/cm float 32 bits 2 Scan-C(f)-Pt14 317 pF/cm float 32 bits 2 Scan-C(f)-Pt15 319 pF/cm float 32 bits 2 Scan-C(f)-Pt16 321 pF/cm float 32 bits 2 Scan-C(f)-Pt17 323 pF/cm float 32 bits 2 Scan-C(f)-Pt18 325 pF/cm float 32 bits 2 Scan-C(f)-Pt19 327 pF/cm float 32 bits 2 Scan-C(f)-Pt20 329 pF/cm float 32 bits 2 Scan-C(f)-Pt21 331 pF/cm float 32 bits 2 Scan-C(f)-Pt22 333 pF/cm float 32 bits 2 Scan-C(f)-Pt23 335 pF/cm float 32 bits 2 Scan-C(f)-Pt24 337 pF/cm float 32 bits 2 Scan-C(f)-Pt25 339 pF/cm float 32 bits 2 Scan-C(f)-Pt26 341 pF/cm float 32 bits 2 Scan-C(f)-Pt27 343 pF/cm float 32 bits 2 Scan-C(f)-Pt28 345 pF/cm float 32 bits 2 Scan-C(f)-Pt29 347 pF/cm float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) Table 6: Incyte Input Registers. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 123Cell Density Monitoring System Operating Instructions Gateway serial number 351   string 6 Measures - Ch 1         Sensor serial number 357   string 6 Amplifier serial number 363 string 6 Amplifier status 369 word 16 bits 1 Reserved 370 float 32 2 Biomass 372 float 32 2 Measure Index 374 unsigned long 2 Amplifier Status – float format 376 float 32 2 Reserved 378 float 32 2 Reserved 380 float 32 2 Reserved 382 float 32 2 Reserved 384 float 32 2 Reserved 386 float 32 2 Acknowledge Val. Dencytee 388 word 16 bits 1 State Val. Dencytee 389 word 16 bits 1 Acknowledge Cal. Dencytee 390 word 16 bits 1 State Cal. Dencytee 391 word 16 bits 1 Reserved 392 float32 2 Reserved 394 float 32 2 Reserved 396 float 32 2 Reserved 398 float 32 2 Reserved 400 float 32 2 Reserved 402 float 32 2 Measures - Ch 2         Sensor serial number 404   string 6 Amplifier serial number 410 string 6 Amplifier status 416 word 16 bits 1 Reserved 417 float 32 2 Biomass 419 float 32 2 Measure Index 421 unsigned long 2 Amplifier Status – float format 423 float 32 2 Reserved 425 float 32 2 Reserved 427 float 32 2 Reserved 429 float 32 2 Reserved 431 float 32 2 Reserved 433 float 32 2 Acknowledge Val. Dencytee 435 word 16 bits 1 State Val. Dencytee 436 word 16 bits 1 Acknowledge Cal. Dencytee 437 word 16 bits 1 State Cal. Dencytee 438 word 16 bits 1 Reserved 439 float32 2 Reserved 441 float 32 2 Reserved 443 float 32 2 Reserved 445 float 32 2 Reserved 447 float 32 2 Reserved 449 float 32 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) 12.1.2.7.1.3 Dencytee Input Registers Table Table 7: Dencytee input register. 12.1.2.7.2 Holding Registers (read and write) 12.1.2.7.2.1 Read and Write Function Code Writing holding registers allow changing configuration and performs several actions. • Use 0x06 standard function code to write single holding register. This function code is used to write a single holding register in the Control Unit. The Request PDU specifies the address of the register to be written. Registers are addressed starting at zero. Therefore register numbered 1 is addressed as 0. The normal response is an echo of the request, returned after the register contents have been written. • Use 0x10 standard function code to write holding registers. This function code is used to write a block of contiguous registers (1 to 123 registers) in the CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 125Cell Density Monitoring System Operating Instructions Channel 1 Mark Zero auto 1 word 16bits 1 Clear Zero auto 2 word 16bits 1 Recepy number Unit Factor 3 word 16bits 1 Recepy number Measure Mode 4 word 16bits 1 Activate Manual Cleaning 5 word 16bits 1 Activate auto Cleaning 6 word 16bits 1 Recepy number Cleaning 7 word 16bits 1 Reserved 8 word 16bits 1 Offset zero 9 pF/cm Float 32bits 2 Command Val. Incyte 11 word 16bits 1 Parameter Val. Incyte 12 word 16bits 1 Command Cal. Incyte 13 word 16bits 1 Parameter Cal. Incyte 14 word 16bits 1 Reserved 15 word 16bits 1 Reserved 16 word 16bits 1 Channel 2 Mark Zero auto 17 word 16bits 1 Clear Zero auto 18 word 16bits 1 Recepy number Unit Factor 19 word 16bits 1 Recepy number Measure Mode 20 word 16bits 1 Activate Manual Cleaning 21 word 16bits 1 Activate auto Cleaning 22 word 16bits 1 Recepy number Cleaning 23 word 16bits 1 Reserved 24 word 16bits 1 Offset zero 25 pF/cm Float 32bits 2 Command Val. Incyte 27 word 16bits 1 Parameter Val. Incyte 28 word 16bits 1 Command Cal. Incyte 29 word 16bits 1 Parameter Cal. Incyte 30 word 16bits 1 Reserved 31 word 16bits 1 Reserved 32 word 16bits 1 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) Control Unit. The data to be written is specified in the request data field. Data is packed as two bytes per register. The normal response returns the function code, starting address, and quantity of registers written. • Use 0x03 standard function code to read Holding Registers This function code is used to read the contents of a contiguous block of holding registers in the Control Unit. The Request PDU specifies the starting register address and the number of registers. In the PDU registers are addressed starting at zero. Therefore registers numbered 1-16 are addressed as 0-15. The register data in the response message are packed as two bytes per register, with the binary contents right justified within each byte. 12.1.2.7.2.2 Incyte Holding Registers Table Table 8: Incyte holding registers table. 12.1.2.7.2.3 Dencytee Holding Registers Table Channel 1 Mark Zero auto 33 word 16bits 1 Clear Zero auto 34 word 16bits 1 Recepy number Unit Factor 35 word 16bits 1 Recepy number Measure Mode 36 word 16bits 1 Reserved 37 word 16bits 1 Reserved 38 word 16bits 1 Reserved 39 word 16bits 1 Reserved 40 word 16bits 1 Offset zero 41 pF/cm Float 32bits 2 Command Val. Dencytee 43 word 16bits 1 Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 127Cell Density Monitoring System Operating Instructions Parameter Val. Dencytee 44 word 16bits 1 Command Cal. Dencytee 45 word 16bits 1 Parameter Cal. Dencytee 46 word 16bits 1 Reserved 47 word 16bits 1 Reserved 48 word 16bits 1 Channel 2 Mark Zero auto 49 word 16bits 1 Clear Zero auto 50 word 16bits 1 Recepy number Unit Factor 51 word 16bits 1 Recepy number Measure Mode 52 word 16bits 1 Reserved 53 word 16bits 1 Reserved 54 word 16bits 1 Reserved 55 word 16bits 1 Reserved 56 word 16bits 1 Offset zero 57 pF/cm Float 32bits 2 Command Val. Dencytee 59 word 16bits 1 Parameter Val. Dencytee 60 word 16bits 1 Command Cal. Dencytee 61 word 16bits 1 Parameter Cal. Dencytee 62 word 16bits 1 Reserved 63 word 16bits 1 Reserved 64 word 16bits 1 Function Register Unit Data type Data Size address (registers) Table 9: Dencytee holding registers table. 12.1.2.7.2.4 Allowed Values for Holding Registers Only the following values may be written in the holding registers: Register Name Value(s) Mark Zero 1 Clear Zero 1 Activate Auto Cleaning 0 to 1 Recipe Number Unit Factor 0 to 8 Recipe Number Measure Mode 0 to 3 Recipe Number Cleaning 0 to 1 Table 10: Allowed values for holding registers. 12.1.2.8 Modbus Tables Arc View Controller 465 (XL) 12.1.2.8.1 Input Registers Tables (Read only) 12.1.2.8.1.1 Read the Modbus Commands All measurements and readable information which comes from the sensors are memorized in the Modbus input registers table. The Modbus function code 0x04 is used to read this table. It allows reading 1 to 125 contiguous input registers from the Arc View Controller. The request specifies the starting address and the number of registers to be read. In the Modbus frame, registers are addressed starting at zero. For example, input registers numbered 1-16 are addressed as 0-15 inside the frame. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 129Cell Density Monitoring System Operating Instructions Gateway serial number 001   string 6 Measures - Ch 1         Sensor serial number 007   string 6 Amplifier serial number 013   string 6 Amplifier status 019   word 16bits 1 Conductivity 020 mS/cm float32 2 Permittivity 022 pF/cm float32 2 Measure Index 024   unsigned long 2 Amplifier Status - float format 026   float32 2 Reserved 028 float32 2 Reserved 030 float32 2 Reserved 032 float32 2 Reserved 034 float32 2 Reserved 036 float32 2 Acknowledge Val. Incyte 038   word 16bits 1 State Val. Incyte 039   word 16bits 1 Acknowledge Cal. Incyte 040   word 16bits 1 State Cal. Incyte 041   word 16bits 1 Spectro - Ch 1         DeltaEpsilon 042   float32 2 Fc 044   float32 2 Alpha 046   float32 2 FscanEpsilon 048   float32 2 Spectro index 050   unsigned long 2 Reserved 052 Word 16bits 1 Measures - Ch 2         Sensor serial number 053   string 6 Amplifier serial number 059   string 6 Amplifier status 065   word 16bits 1 Conductivity 066 mS/cm float32 2 Permittivity 068 pF/cm float32 2 Measure Index 070   unsigned long 2 Amplifier Status - float format 072   float32 2 Reserved 074 float32 2 Reserved 076 float32 2 Reserved 078 float32 2 Reserved 080 float32 2 Reserved 082 float32 2 Acknowledge Val. Incyte 084   word 16bits 1 State Val. Incyte 085   word 16bits 1 Acknowledge Cal. Incyte 086   word 16bits 1 State Cal. Incyte 087   word 16bits 1 Spectro - Ch 2         DeltaEpsilon 088   float32 2 Fc 090   float32 2 Alpha 092   float32 2 FscanEpsilon 094   float32 2 Spectro index 096   float32 / ul 2 Reserved 098   word 16bits 1 Measures - Ch 3         Sensor serial number 099   string 6 Amplifier serial number 105   string 6 Amplifier status 111   word 16bits 1 Conductivity 112 mS/cm float32 2 Permittivity 114 pF/cm float32 2 Measure Index 116   unsigned long 2 Amplifier Status - float format 118   float32 2 Reserved 120 float32 2 Reserved 122 float32 2 Reserved 124 float32 2 Reserved 126 float32 2 Reserved 128 float32 2 Acknowledge Val. Incyte 130   word 16bits 1 State Val. Incyte 131   word 16bits 1 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) 12.1.2.8.1.2 Incyte Input Registers Table CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 131Cell Density Monitoring System Operating Instructions Acknowledge Cal. Incyte 132   word 16bits 1 State Cal. Incyte 133   word 16bits 1 Spectro - Ch 3       DeltaEpsilon 134   float32 2 Fc 136   float32 2 Alpha 138   float32 2 FscanEpsilon 140   float32 2 Spectro index 142   unsigned long 2 Reserved 144 Word 16bits 1 Measures - Ch 4         Sensor serial number 145   string 6 Amplifier serial number 151   string 6 Amplifier status 157   word 16bits 1 Conductivity 158 mS/cm float32 2 Permittivity 160 pF/cm float32 2 Measure Index 162   unsigned long 2 Amplifier Status - float format 164   float32 2 Reserved 166 float32 2 Reserved 168 float32 2 Reserved 170 float32 2 Reserved 172 float32 2 Reserved 174 float32 2 Acknowledge Val. Incyte 176   word 16bits 1 State Val. Incyte 177   word 16bits 1 Acknowledge Cal. Incyte 178   word 16bits 1 State Cal. Incyte 179   word 16bits 1 Spectro - Ch 4       DeltaEpsilon 180   float32 2 Fc 182   float32 2 Alpha 184   float32 2 FscanEpsilon 186   float32 2 Spectro index 188   unsigned long 2 Reserved 190 Word 16bits 1 Info-Fscan-Ch1         Reserved 191   word 16bits 1 Reserved 192   word 16bits 1 Reserved 193   word 16bits 1 index-Fscan-ch1 194   unsigned long 2 Freq-FSCAN-Ch1(30 points)       Scan-Freq-Pt0 196 Hz float 32 bits 2 Scan-Freq-Pt1 198 Hz float 32 bits 2 Scan-Freq-Pt2 200 Hz float 32 bits 2 Scan-Freq-Pt3 202 Hz float 32 bits 2 Scan-Freq-Pt4 204 Hz float 32 bits 2 Scan-Freq-Pt5 206 Hz float 32 bits 2 Scan-Freq-Pt6 208 Hz float 32 bits 2 Scan-Freq-Pt7 210 Hz float 32 bits 2 Scan-Freq-Pt8 212 Hz float 32 bits 2 Scan-Freq-Pt9 214 Hz float 32 bits 2 Scan-Freq-Pt10 216 Hz float 32 bits 2 Scan-Freq-Pt11 218 Hz float 32 bits 2 Scan-Freq-Pt12 220 Hz float 32 bits 2 Scan-Freq-Pt13 222 Hz float 32 bits 2 Scan-Freq-Pt14 224 Hz float 32 bits 2 Scan-Freq-Pt15 226 Hz float 32 bits 2 Scan-Freq-Pt16 228 Hz float 32 bits 2 Scan-Freq-Pt17 230 Hz float 32 bits 2 Scan-Freq-Pt18 232 Hz float 32 bits 2 Scan-Freq-Pt19 234 Hz float 32 bits 2 Scan-Freq-Pt20 236 Hz float 32 bits 2 Scan-Freq-Pt21 238 Hz float 32 bits 2 Scan-Freq-Pt22 240 Hz float 32 bits 2 Scan-Freq-Pt23 242 Hz float 32 bits 2 Scan-Freq-Pt24 244 Hz float 32 bits 2 Scan-Freq-Pt25 246 Hz float 32 bits 2 Scan-Freq-Pt26 248 Hz float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 133Cell Density Monitoring System Operating Instructions Scan-Freq-Pt27 250 Hz float 32 bits 2 Scan-Freq-Pt28 252 Hz float 32 bits 2 Scan-Freq-Pt29 254 Hz float 32 bits 2 C(f)-FSCAN-Ch1(30 points)       Scan-C(f)-Pt0 256 pF/cm float 32 bits 2 Scan-C(f)-Pt1 258 pF/cm float 32 bits 2 Scan-C(f)-Pt2 260 pF/cm float 32 bits 2 Scan-C(f)-Pt3 262 pF/cm float 32 bits 2 Scan-C(f)-Pt4 264 pF/cm float 32 bits 2 Scan-C(f)-Pt5 266 pF/cm float 32 bits 2 Scan-C(f)-Pt6 268 pF/cm float 32 bits 2 Scan-C(f)-Pt7 270 pF/cm float 32 bits 2 Scan-C(f)-Pt8 272 pF/cm float 32 bits 2 Scan-C(f)-Pt9 274 pF/cm float 32 bits 2 Scan-C(f)-Pt10 276 pF/cm float 32 bits 2 Scan-C(f)-Pt11 278 pF/cm float 32 bits 2 Scan-C(f)-Pt12 280 pF/cm float 32 bits 2 Scan-C(f)-Pt13 282 pF/cm float 32 bits 2 Scan-C(f)-Pt14 284 pF/cm float 32 bits 2 Scan-C(f)-Pt15 286 pF/cm float 32 bits 2 Scan-C(f)-Pt16 288 pF/cm float 32 bits 2 Scan-C(f)-Pt17 290 pF/cm float 32 bits 2 Scan-C(f)-Pt18 292 pF/cm float 32 bits 2 Scan-C(f)-Pt19 294 pF/cm float 32 bits 2 Scan-C(f)-Pt20 296 pF/cm float 32 bits 2 Scan-C(f)-Pt21 298 pF/cm float 32 bits 2 Scan-C(f)-Pt22 300 pF/cm float 32 bits 2 Scan-C(f)-Pt23 302 pF/cm float 32 bits 2 Scan-C(f)-Pt24 304 pF/cm float 32 bits 2 Scan-C(f)-Pt25 306 pF/cm float 32 bits 2 Scan-C(f)-Pt26 308 pF/cm float 32 bits 2 Scan-C(f)-Pt27 310 pF/cm float 32 bits 2 Scan-C(f)-Pt28 312 pF/cm float 32 bits 2 Scan-C(f)-Pt29 314 pF/cm float 32 bits 2 Info-Fscan-Ch2     Reserved 316   word 16bits 1 Reserved 317   word 16bits 1 Reserved 318   word 16bits 1 index-Fscan-ch2 319   unsigned long 2 Freq-FSCAN-Ch2(30 points)       Scan-Freq-Pt0 321 Hz float 32 bits 2 Scan-Freq-Pt1 323 Hz float 32 bits 2 Scan-Freq-Pt2 325 Hz float 32 bits 2 Scan-Freq-Pt3 327 Hz float 32 bits 2 Scan-Freq-Pt4 329 Hz float 32 bits 2 Scan-Freq-Pt5 331 Hz float 32 bits 2 Scan-Freq-Pt6 333 Hz float 32 bits 2 Scan-Freq-Pt7 335 Hz float 32 bits 2 Scan-Freq-Pt8 337 Hz float 32 bits 2 Scan-Freq-Pt9 339 Hz float 32 bits 2 Scan-Freq-Pt10 341 Hz float 32 bits 2 Scan-Freq-Pt11 343 Hz float 32 bits 2 Scan-Freq-Pt12 345 Hz float 32 bits 2 Scan-Freq-Pt13 347 Hz float 32 bits 2 Scan-Freq-Pt14 349 Hz float 32 bits 2 Scan-Freq-Pt15 351 Hz float 32 bits 2 Scan-Freq-Pt16 353 Hz float 32 bits 2 Scan-Freq-Pt17 355 Hz float 32 bits 2 Scan-Freq-Pt18 357 Hz float 32 bits 2 Scan-Freq-Pt19 359 Hz float 32 bits 2 Scan-Freq-Pt20 361 Hz float 32 bits 2 Scan-Freq-Pt21 363 Hz float 32 bits 2 Scan-Freq-Pt22 365 Hz float 32 bits 2 Scan-Freq-Pt23 367 Hz float 32 bits 2 Scan-Freq-Pt24 369 Hz float 32 bits 2 Scan-Freq-Pt25 371 Hz float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 135Cell Density Monitoring System Operating Instructions Scan-Freq-Pt26 373 Hz float 32 bits 2 Scan-Freq-Pt27 375 Hz float 32 bits 2 Scan-Freq-Pt28 377 Hz float 32 bits 2 Scan-Freq-Pt29 379 Hz float 32 bits 2 C(f)-FSCAN-Ch2(30 points)       Scan-C(f)-Pt0 381 pF/cm float 32 bits 2 Scan-C(f)-Pt1 383 pF/cm float 32 bits 2 Scan-C(f)-Pt2 385 pF/cm float 32 bits 2 Scan-C(f)-Pt3 387 pF/cm float 32 bits 2 Scan-C(f)-Pt4 389 pF/cm float 32 bits 2 Scan-C(f)-Pt5 391 pF/cm float 32 bits 2 Scan-C(f)-Pt6 393 pF/cm float 32 bits 2 Scan-C(f)-Pt7 395 pF/cm float 32 bits 2 Scan-C(f)-Pt8 397 pF/cm float 32 bits 2 Scan-C(f)-Pt9 399 pF/cm float 32 bits 2 Scan-C(f)-Pt10 401 pF/cm float 32 bits 2 Scan-C(f)-Pt11 403 pF/cm float 32 bits 2 Scan-C(f)-Pt12 405 pF/cm float 32 bits 2 Scan-C(f)-Pt13 407 pF/cm float 32 bits 2 Scan-C(f)-Pt14 409 pF/cm float 32 bits 2 Scan-C(f)-Pt15 411 pF/cm float 32 bits 2 Scan-C(f)-Pt16 413 pF/cm float 32 bits 2 Scan-C(f)-Pt17 415 pF/cm float 32 bits 2 Scan-C(f)-Pt18 417 pF/cm float 32 bits 2 Scan-C(f)-Pt19 419 pF/cm float 32 bits 2 Scan-C(f)-Pt20 421 pF/cm float 32 bits 2 Scan-C(f)-Pt21 423 pF/cm float 32 bits 2 Scan-C(f)-Pt22 425 pF/cm float 32 bits 2 Scan-C(f)-Pt23 427 pF/cm float 32 bits 2 Scan-C(f)-Pt24 429 pF/cm float 32 bits 2 Scan-C(f)-Pt25 431 pF/cm float 32 bits 2 Scan-C(f)-Pt26 433 pF/cm float 32 bits 2 Scan-C(f)-Pt27 435 pF/cm float 32 bits 2 Scan-C(f)-Pt28 437 pF/cm float 32 bits 2 Scan-C(f)-Pt29 439 pF/cm float 32 bits 2 Info-Fscan-Ch3       Reserved 441   word 16bits 1 Reserved 442   word 16bits 1 Reserved 443   word 16bits 1 index-Fscan-ch1 444   unsigned long 2 Freq-FSCAN-Ch3(30 points)       Scan-Freq-Pt0 446 Hz float 32 bits 2 Scan-Freq-Pt1 448 Hz float 32 bits 2 Scan-Freq-Pt2 450 Hz float 32 bits 2 Scan-Freq-Pt3 452 Hz float 32 bits 2 Scan-Freq-Pt4 454 Hz float 32 bits 2 Scan-Freq-Pt5 456 Hz float 32 bits 2 Scan-Freq-Pt6 458 Hz float 32 bits 2 Scan-Freq-Pt7 460 Hz float 32 bits 2 Scan-Freq-Pt8 462 Hz float 32 bits 2 Scan-Freq-Pt9 464 Hz float 32 bits 2 Scan-Freq-Pt10 466 Hz float 32 bits 2 Scan-Freq-Pt11 468 Hz float 32 bits 2 Scan-Freq-Pt12 470 Hz float 32 bits 2 Scan-Freq-Pt13 472 Hz float 32 bits 2 Scan-Freq-Pt14 474 Hz float 32 bits 2 Scan-Freq-Pt15 476 Hz float 32 bits 2 Scan-Freq-Pt16 478 Hz float 32 bits 2 Scan-Freq-Pt17 480 Hz float 32 bits 2 Scan-Freq-Pt18 482 Hz float 32 bits 2 Scan-Freq-Pt19 484 Hz float 32 bits 2 Scan-Freq-Pt20 486 Hz float 32 bits 2 Scan-Freq-Pt21 488 Hz float 32 bits 2 Scan-Freq-Pt22 490 Hz float 32 bits 2 Scan-Freq-Pt23 492 Hz float 32 bits 2 Scan-Freq-Pt24 494 Hz float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 137Cell Density Monitoring System Operating Instructions Scan-Freq-Pt25 496 Hz float 32 bits 2 Scan-Freq-Pt26 498 Hz float 32 bits 2 Scan-Freq-Pt27 500 Hz float 32 bits 2 Scan-Freq-Pt28 502 Hz float 32 bits 2 Scan-Freq-Pt29 504 Hz float 32 bits 2 C(f)-FSCAN-Ch3(30 points)       Scan-C(f)-Pt0 506 pF/cm float 32 bits 2 Scan-C(f)-Pt1 508 pF/cm float 32 bits 2 Scan-C(f)-Pt2 510 pF/cm float 32 bits 2 Scan-C(f)-Pt3 512 pF/cm float 32 bits 2 Scan-C(f)-Pt4 514 pF/cm float 32 bits 2 Scan-C(f)-Pt5 516 pF/cm float 32 bits 2 Scan-C(f)-Pt6 518 pF/cm float 32 bits 2 Scan-C(f)-Pt7 520 pF/cm float 32 bits 2 Scan-C(f)-Pt8 522 pF/cm float 32 bits 2 Scan-C(f)-Pt9 524 pF/cm float 32 bits 2 Scan-C(f)-Pt10 526 pF/cm float 32 bits 2 Scan-C(f)-Pt11 528 pF/cm float 32 bits 2 Scan-C(f)-Pt12 530 pF/cm float 32 bits 2 Scan-C(f)-Pt13 532 pF/cm float 32 bits 2 Scan-C(f)-Pt14 534 pF/cm float 32 bits 2 Scan-C(f)-Pt15 536 pF/cm float 32 bits 2 Scan-C(f)-Pt16 538 pF/cm float 32 bits 2 Scan-C(f)-Pt17 540 pF/cm float 32 bits 2 Scan-C(f)-Pt18 542 pF/cm float 32 bits 2 Scan-C(f)-Pt19 544 pF/cm float 32 bits 2 Scan-C(f)-Pt20 546 pF/cm float 32 bits 2 Scan-C(f)-Pt21 548 pF/cm float 32 bits 2 Scan-C(f)-Pt22 550 pF/cm float 32 bits 2 Scan-C(f)-Pt23 552 pF/cm float 32 bits 2 Scan-C(f)-Pt24 554 pF/cm float 32 bits 2 Scan-C(f)-Pt25 556 pF/cm float 32 bits 2 Scan-C(f)-Pt26 558 pF/cm float 32 bits 2 Scan-C(f)-Pt27 560 pF/cm float 32 bits 2 Scan-C(f)-Pt28 562 pF/cm float 32 bits 2 Scan-C(f)-Pt29 564 pF/cm float 32 bits 2 Info-Fscan-Ch4     Reserved 566   word 16bits 1 Reserved 567   word 16bits 1 Reserved 568   word 16bits 1 index-Fscan-ch2 569   unsigned long 2 Freq-FSCAN-Ch4(30 points)       Scan-Freq-Pt0 571 Hz float 32 bits 2 Scan-Freq-Pt1 573 Hz float 32 bits 2 Scan-Freq-Pt2 575 Hz float 32 bits 2 Scan-Freq-Pt3 577 Hz float 32 bits 2 Scan-Freq-Pt4 579 Hz float 32 bits 2 Scan-Freq-Pt5 581 Hz float 32 bits 2 Scan-Freq-Pt6 583 Hz float 32 bits 2 Scan-Freq-Pt7 585 Hz float 32 bits 2 Scan-Freq-Pt8 587 Hz float 32 bits 2 Scan-Freq-Pt9 589 Hz float 32 bits 2 Scan-Freq-Pt10 591 Hz float 32 bits 2 Scan-Freq-Pt11 593 Hz float 32 bits 2 Scan-Freq-Pt12 595 Hz float 32 bits 2 Scan-Freq-Pt13 597 Hz float 32 bits 2 Scan-Freq-Pt14 599 Hz float 32 bits 2 Scan-Freq-Pt15 601 Hz float 32 bits 2 Scan-Freq-Pt16 603 Hz float 32 bits 2 Scan-Freq-Pt17 605 Hz float 32 bits 2 Scan-Freq-Pt18 607 Hz float 32 bits 2 Scan-Freq-Pt19 609 Hz float 32 bits 2 Scan-Freq-Pt20 611 Hz float 32 bits 2 Scan-Freq-Pt21 613 Hz float 32 bits 2 Scan-Freq-Pt22 615 Hz float 32 bits 2 Scan-Freq-Pt23 617 Hz float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 139Cell Density Monitoring System Operating Instructions Scan-Freq-Pt24 619 Hz float 32 bits 2 Scan-Freq-Pt25 621 Hz float 32 bits 2 Scan-Freq-Pt26 623 Hz float 32 bits 2 Scan-Freq-Pt27 625 Hz float 32 bits 2 Scan-Freq-Pt28 627 Hz float 32 bits 2 Scan-Freq-Pt29 629 Hz float 32 bits 2 C(f)-FSCAN-Ch4(30 points)       Scan-C(f)-Pt0 631 pF/cm float 32 bits 2 Scan-C(f)-Pt1 633 pF/cm float 32 bits 2 Scan-C(f)-Pt2 635 pF/cm float 32 bits 2 Scan-C(f)-Pt3 637 pF/cm float 32 bits 2 Scan-C(f)-Pt4 639 pF/cm float 32 bits 2 Scan-C(f)-Pt5 641 pF/cm float 32 bits 2 Scan-C(f)-Pt6 643 pF/cm float 32 bits 2 Scan-C(f)-Pt7 645 pF/cm float 32 bits 2 Scan-C(f)-Pt8 647 pF/cm float 32 bits 2 Scan-C(f)-Pt9 649 pF/cm float 32 bits 2 Scan-C(f)-Pt10 651 pF/cm float 32 bits 2 Scan-C(f)-Pt11 653 pF/cm float 32 bits 2 Scan-C(f)-Pt12 655 pF/cm float 32 bits 2 Scan-C(f)-Pt13 657 pF/cm float 32 bits 2 Scan-C(f)-Pt14 659 pF/cm float 32 bits 2 Scan-C(f)-Pt15 661 pF/cm float 32 bits 2 Scan-C(f)-Pt16 663 pF/cm float 32 bits 2 Scan-C(f)-Pt17 665 pF/cm float 32 bits 2 Scan-C(f)-Pt18 667 pF/cm float 32 bits 2 Scan-C(f)-Pt19 669 pF/cm float 32 bits 2 Scan-C(f)-Pt20 671 pF/cm float 32 bits 2 Scan-C(f)-Pt21 673 pF/cm float 32 bits 2 Scan-C(f)-Pt22 675 pF/cm float 32 bits 2 Scan-C(f)-Pt23 677 pF/cm float 32 bits 2 Scan-C(f)-Pt24 679 pF/cm float 32 bits 2 Scan-C(f)-Pt25 681 pF/cm float 32 bits 2 Scan-C(f)-Pt26 683 pF/cm float 32 bits 2 Scan-C(f)-Pt27 685 pF/cm float 32 bits 2 Scan-C(f)-Pt28 687 pF/cm float 32 bits 2 Scan-C(f)-Pt29 689 pF/cm float 32 bits 2 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) Table 11: Incyte Input Registers. 12.1.2.8.1.3 Dencytee Input Registers Table Gateway serial number 693   string 6 Measures - Ch 1       Sensor serial number 699   string 6 Amplifier serial number 705 string 6 Amplifier status 711 word 16 bits 1 Reserved 712 float 32 2 Biomass 714 float 32 2 Measure Index 716 unsigned long 2 Amplifier Status – float format 718 float 32 2 Reserved 720 float 32 2 Reserved 722 float 32 2 Reserved 724 float 32 2 Reserved 726 float 32 2 Reserved 728 float 32 2 Acknowledge Val. Dencytee 730 word 16 bits 1 State Val. Dencytee 731 word 16 bits 1 Acknowledge Cal. Dencytee 732 word 16 bits 1 State Cal. Dencytee 733 word 16 bits 1 Reserved 734 float32 2 Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 141Cell Density Monitoring System Operating Instructions Reserved 736 float 32 2 Reserved 738 float 32 2 Reserved 740 float 32 2 Reserved 742 float 32 2 Reserved 744 float 32 2 Measures - Ch 2       Sensor serial number 746   string 6 Amplifier serial number 752 string 6 Amplifier status 758 word 16 bits 1 Reserved 759 float 32 2 Biomass 761 float 32 2 Measure Index 763 unsigned long 2 Amplifier Status – float format 765 float 32 2 Reserved 767 float 32 2 Reserved 769 float 32 2 Reserved 771 float 32 2 Reserved 773 float 32 2 Reserved 775 float 32 2 Acknowledge Val. Dencytee 777 word 16 bits 1 State Val. Dencytee 778 word 16 bits 1 Acknowledge Cal. Dencytee 779 word 16 bits 1 State Cal. Dencytee 780 word 16 bits 1 Reserved 781 float32 2 Reserved 783 float 32 2 Reserved 785 float 32 2 Reserved 787 float 32 2 Reserved 789 float 32 2 Reserved 791 float 32 2 Measures - Ch 3       Sensor serial number 793   string 6 Amplifier serial number 799 string 6 Amplifier status 805 word 16 bits 1 Reserved 806 float 32 2 Biomass 808 float 32 2 Measure Index 810 unsigned long 2 Amplifier Status – float format 812 float 32 2 Reserved 814 float 32 2 Reserved 816 float 32 2 Reserved 818 float 32 2 Reserved 820 float 32 2 Reserved 822 float 32 2 Acknowledge Val. Dencytee 824 word 16 bits 1 State Val. Dencytee 825 word 16 bits 1 Acknowledge Cal. Dencytee 826 word 16 bits 1 State Cal. Dencytee 827 word 16 bits 1 Reserved 828 float32 2 Reserved 830 float 32 2 Reserved 832 float 32 2 Reserved 834 float 32 2 Reserved 836 float 32 2 Reserved 838 float 32 2 Measures - Ch 4       Sensor serial number 840   string 6 Amplifier serial number 846 string 6 Amplifier status 852 word 16 bits 1 Reserved 853 float 32 2 Biomass 855 float 32 2 Measure Index 857 unsigned long 2 Amplifier Status – float format 859 float 32 2 Reserved 861 float 32 2 Reserved 863 float 32 2 Reserved 865 float 32 2 Reserved 867 float 32 2 Reserved 869 float 32 2 Acknowledge Val. Dencytee 871 word 16 bits 1 State Val. Dencytee 872 word 16 bits 1 Function Register Unit Data type Data Size address (registers) Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 143Cell Density Monitoring System Operating Instructions Function Register Unit Data type Data Size address (registers) Acknowledge Cal. Dencytee 873 word 16 bits 1 State Cal. Dencytee 874 word 16 bits 1 Reserved 875 float32 2 Reserved 877 float 32 2 Reserved 879 float 32 2 Reserved 881 float 32 2 Reserved 883 float 32 2 Reserved 885 float 32 2 Function Register Unit Data type Data Size address (registers) Table 12: Dencytee input register. 12.1.2.8.2 Holding Registers (read and write) 12.1.2.8.2.1 Read and Write Function Code Writing holding registers allow changing configuration and performs several actions. • Use 0x06 standard function code to write single holding register. This function code is used to write a single holding register in the Arc View Controller. The Request PDU specifies the address of the register to be written. Registers are addressed starting at zero. Therefore register numbered 1 is addressed as 0. The normal response is an echo of the request, returned after the register contents have been written. • Use 0x10 standard function code to write holding registers. This function code is used to write a block of contiguous registers (1 to 123 registers) in the Arc View Controller. The data to be written is specified in the request data field. Data is packed as two bytes per register. The normal response returns the function code, starting address, and quantity of registers written. • Use 0x03 standard function code to read Holding Registers This function code is used to read the contents of a contiguous block of holding registers in the Arc View Controller. The Request PDU specifies the starting register address and the number of registers. In the PDU registers are addressed starting at zero. Therefore registers numbered 1-16 are addressed as 0-15. The register data in the response message are packed as two bytes per register, with the binary contents right justified within each byte. 12.1.2.8.2.2 Incyte Holding Registers Table Channel 1 Mark Zero auto 1 word 16bits 1 Clear Zero auto 2 word 16bits 1 Recepy number Unit Factor 3 word 16bits 1 Recepy number Measure Mode 4 word 16bits 1 Activate Manual Cleaning 5 word 16bits 1 Activate auto Cleaning 6 word 16bits 1 Recepy number Cleaning 7 word 16bits 1 Reserved 8 word 16bits 1 Offset zero 9 pF/cm Float 32bits 2 Command Val. Incyte 11 word 16bits 1 Parameter Val. Incyte 12 word 16bits 1 Command Cal. Incyte 13 word 16bits 1 Parameter Cal. Incyte 14 word 16bits 1 Reserved 15 word 16bits 1 Reserved 16 word 16bits 1 Channel 2 Mark Zero auto 17 word 16bits 1 Clear Zero auto 18 word 16bits 1 Recepy number Unit Factor 19 word 16bits 1 Recepy number Measure Mode 20 word 16bits 1 Activate Manual Cleaning 21 word 16bits 1 Activate auto Cleaning 22 word 16bits 1 Recepy number Cleaning 23 word 16bits 1 Reserved 24 word 16bits 1 CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 145Cell Density Monitoring System Operating Instructions Offset zero 25 pF/cm Float 32bits 2 Command Val. Incyte 27 word 16bits 1 Parameter Val. Incyte 28 word 16bits 1 Command Cal. Incyte 29 word 16bits 1 Parameter Cal. Incyte 30 word 16bits 1 Reserved 31 word 16bits 1 Reserved 32 word 16bits 1 Channel 3 Mark Zero auto 33 word 16bits 1 Clear Zero auto 34 word 16bits 1 Recepy numbrer Unit Factor 35 word 16bits 1 Recepy number Measure Mode 36 word 16bits 1 Activate Manual Cleaning 37 word 16bits 1 Activate auto Cleaning 38 word 16bits 1 Recepy number Cleaning 39 word 16bits 1 Reserved 40 word 16bits 1 Offset zero 41 pF/cm Float 32bits 2 Command Val. Incyte 43 word 16bits 1 Parameter Val. Incyte 44 word 16bits 1 Command Cal. Incyte 45 word 16bits 1 Parameter Cal. Incyte 46 word 16bits 1 Reserved 47 word 16bits 1 Reserved 48 word 16bits 1 Channel 4 Mark Zero auto 49 word 16bits 1 Clear Zero auto 50 word 16bits 1 Recepy numbrer Unit Factor 51 word 16bits 1 Recepy number Measure Mode 52 word 16bits 1 Activate Manual Cleaning 53 word 16bits 1 Activate auto Cleaning 54 word 16bits 1 Recepy number Cleaning 55 word 16bits 1 Reserved 56 word 16bits 1 Offset zero 57 pF/cm Float 32bits 2 Function Register Unit Data type Data Size address (registers) Table 13: Incyte holding registers table. 12.1.2.8.2.3 Dencytee Holding Registers Table Function Register Unit Data type Data Size address (registers) Channel 1 Mark Zero auto 71 word 16bits 1 Clear Zero auto 72 word 16bits 1 Recepy number Unit Factor 73 word 16bits 1 Recepy number Measure Mode 74 word 16bits 1 Reserved 75 word 16bits 1 Reserved 76 word 16bits 1 Reserved 77 word 16bits 1 Reserved 78 word 16bits 1 Offset zero 79 pF/cm Float 32bits 2 Command Val. Dencytee 81 word 16bits 1 Parameter Val. Dencytee 82 word 16bits 1 Command Cal. Dencytee 83 word 16bits 1 Parameter Cal. Dencytee 84 word 16bits 1 reserved 85 word 16bits 1 reserved 86 word 16bits 1 Channel 2 Mark Zero auto 87 word 16bits 1 Command Val. Incyte 59 word 16bits 1 Parameter Val. Incyte 60 word 16bits 1 Command Cal. Incyte 61 word 16bits 1 Parameter Cal. Incyte 62 word 16bits 1 Reserved 63 word 16bits 1 Reserved 64 word 16bits 1 Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 147Cell Density Monitoring System Operating Instructions Clear Zero auto 88 word 16bits 1 Recepy number Unit Factor 89 word 16bits 1 Recepy number Measure Mode 90 word 16bits 1 Reserved 91 word 16bits 1 Reserved 92 word 16bits 1 Reserved 93 word 16bits 1 Reserved 94 word 16bits 1 Offset zero 95 pF/cm Float 32bits 2 Command Val. Dencytee 97 word 16bits 1 Parameter Val. Dencytee 98 word 16bits 1 Command Cal. Dencytee 99 word 16bits 1 Parameter Cal. Dencytee 100 word 16bits 1 Reserved 101 word 16bits 1 Reserved 102 word 16bits 1 Channel 3 Mark Zero auto 103 word 16bits 1 Clear Zero auto 104 word 16bits 1 Recepy number Unit Factor 105 word 16bits 1 Recepy number Measure Mode 106 word 16bits 1 Reserved 107 word 16bits 1 Reserved 108 word 16bits 1 Reserved 109 word 16bits 1 Reserved 110 word 16bits 1 Offset zero 111 pF/cm Float 32bits 2 Command Val. Dencytee 113 word 16bits 1 Parameter Val. Dencytee 114 word 16bits 1 Command Cal. Dencytee 115 word 16bits 1 Parameter Cal. Dencytee 116 word 16bits 1 Reserved 117 word 16bits 1 Reserved 118 word 16bits 1 Channel 4 Mark Zero auto 119 word 16bits 1 Clear Zero auto 120 word 16bits 1 Function Register Unit Data type Data Size address (registers) Table 14: Dencytee holding registers table. 12.1.2.8.2.4 Allowed Values for Holding Registers Only the following values may be written in the holding registers: Register Name Value(s) Mark Zero 1 Clear Zero 1 Activate Auto Cleaning 0 to 1 Recipe Number Unit Factor 0 to 8 Recipe Number Measure Mode 0 to 3 Recipe Number Cleaning 0 to 1 Table 15: Allowed values for holding registers. Recepy number Unit Factor 121 word 16bits 1 Recepy number Measure Mode 122 word 16bits 1 Reserved 123 word 16bits 1 Reserved 124 word 16bits 1 Reserved 125 word 16bits 1 Reserved 126 word 16bits 1 Offset zero 127 pF/cm Float 32bits 2 Command Val. Dencytee 129 word 16bits 1 Parameter Val. Dencytee 130 word 16bits 1 Command Cal. Dencytee 131 word 16bits 1 Parameter Cal. Dencytee 132 word 16bits 1 Reserved 133 word 16bits 1 Reserved 134 word 16bits 1 Function Register Unit Data type Data Size address (registers) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 149Cell Density Monitoring System Operating Instructions 12.2 Connect by OPC Communication NOTE: To connect the Arc View Controller or ComBox with the process control system, the OPC License (Ref 243820) is needed. The Cell Density Integration Kit is equipped with all licenses. 12.2.1 Introduction to OPC OPC is a standard which specifies the communication of real-time data between control devices from different manufacturers. The Arc View Controller offers an OPC XML-DA server. OPC XMLDA is the OPC Foundation’s adoption of the XML (eXtensible Markup Language) set of technologies. An OPC XML-DA client is used to communicate with the Arc View Controller. The Arc View Controller takes advantage of the “OPC-XML-DA-SDK-LIN” library supplied by Softing. The library is OPC compliant. 12.2.2 Configuration of the Control Unit Enter the Administrator Settings (Figure 47), select the Network Applications and activate OPC (Figure 62). Figure 62: Activation of the OPC protocol. The Arc View Controller OPC server provides the following data: Arc View Controller: • Controller name (with serial number): character string. Sensor Unit: • Amplifier serial number: character string. • Sensor serial number: character string. • Probe status: character string. • Conductivity: floating point value. • Permittivity: floating point value. • Cell Density: floating point value. • Fc (IncyteScan cutting frequency): floating point value. • DeltaEps (IncyteScan Δε): floating point value. • Alpha (IncyteScan α): floating point value. • IncyteScanFrequencies: list of floating point values. • C(f) (IncyteScan Permittivity values): list of floating point values. • IncyteScanPermittivity (Permittivity measured by FScan method): floating point value. NOTE: The refresh rate is about 30s for the Incyte Scan parameters and 3s for the other parameters. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 151Cell Density Monitoring System Operating Instructions 12.2.3 Configuration of the OPC Client The connection of the OPC client to the Control Unit OPC server is made using the following URL: http://ArcViewControllerName:8079/EVOOPC/XMLDA, where «ArcViewControllerName» (ex: E265-XXX-XXX) is the name of the Control Unit that is on the sticker on the left side panel and in the about menu (Figure 13) of the software. Alternatively, you can use the IP address. NOTE: The Control Unit OPC server has been tested and validated with the OPC Classic Demo client, from Softing and the OPC Viewer client from CommServer. Further patches, to ensure the newest version of the bridge may be required. 12.2.4 OPC Classic Demo Client Start Guide 12.2.4.1 System Overview Figure 63: OPC Classic Demo System Overview. 12.2.4.2 Download and Installation 1. Download the OPC-Classic Demo installer for Windows from the Softing homepage. The OPC client must be installed on a user PC connected to the local network. 2. After downloading OPC-Classic Demo, run the setup and follow the instructions. 3. Once the installation is completed, run the OPC Demo Client. If the installation was successful, the following screen is displayed (Figure 64). Figure 64: OPC Classic Demo client. 12.2.4.3 Connecting the OPC Classic Demo Client to the Control Unit 1. Connect the Control Unit and the PC (with the OPC client) to the local network. 2. Connection of the OPC client on the PC to the integrated OPC server in the Control Unit using the following URL: http://ArcViewControllerName:8079/EVOOPC/XMLDA, where “ArcViewControllerName” (ex: E265-XXX-XXX) is the name of the Control Unit that is on the sticker on the left side panel and in the About Menu (Figure 13) of the software. Alternatively, the IP address of the device may be used. In the example below, the IP address 192.168.128.181 was used. Enter the OPC Server URL in the associated text control and press add server (Figure 65). CONTROL UNIT OPC CLIENT LOCAL NETWORK OPC XML-DA USER PC CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 153Cell Density Monitoring System Operating Instructions Figure 65: Insertion of the Arc View Controller URL. 3. If the OPC Server URL is correct, the following screen (Figure 66) is displayed with a green light beside the OPC server: Figure 66: Connected Arc View Controller in the Classic Demo. 4. Select the DA Browse tab to do the data selection (Figure 67). Figure 67: Data selection in the DA Browse tab. 5. Double click on each item that needs to be monitored. Each selected item is added to the list (Figure 68). Figure 68: Item selection in the DA Browse tab. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 155Cell Density Monitoring System Operating Instructions 6. To observe the data evolution select the DA Items tab to see the evolution of the data selection (Figure 69). Figure 69: Connected Arc View Controller with the Tool Box Demo Client. ATTENTION! If there is a red light besides the items, activate the OPC server on your Arc View Controller (see Chapter 9.4). 12.2.5 Softing OPC Bridge Start Guide 12.2.5.1 System Overview Figure 70: System Overview of the Softing OPC bridge. 12.2.5.2 Download and Installation 1. Purchase the dataFEED OPC Suite for Windows, available from Softing. 2. After downloading dataFEED OPC Suite, run the setup and follow the instructions. 3. Start the dataFEED OPC Suite Configurator on the Windows start menu (Start => Programs => Softing OPC Easy Connect => dataFEED OPC Suite => Configurator). 4. Select the Tab Data Source on the left side and click on OPC Server (Figure 71). CONTROL UNIT OPC BRIDGE LOCAL NETWORK OPC XML-DA USER PC OPC CLIENT OPC DA CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 157Cell Density Monitoring System Operating Instructions Figure 71: Starting with the OPC Bridge. 5. Click on the add a new data source button. The OPC Server Wizard starts on the widget (Figure 72). 6. Click next. Figure 72: OPC Server Wizard. 7. Enter a name for the data source and select Server Type XML-DA (Figure 73). Figure 73: Definition of the data source and the server type. 8. Connect the Control Unit to the local network. Connect the PC to the local network, the connection of the OPC client to the Control Unit OPC server is made using the following URL: http://ArcViewControllerName:8079/EVOOPC/XMLDA, where “ArcViewControllerName” (ex: E265-XXX-XXX) is the name of the Control Unit that is on the sticker on the left side panel and in the About Menu (Figure 13) in the Controller software. Alternatively, the IP address of the device may be used. In the example below, the IP address 172.16.125.102 was used (Figure 74). Enter the URL and proceed. Figure 74: Definition of the Control Unit URL. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 159Cell Density Monitoring System Operating Instructions 9. Test the connection (Figure 75). Figure 75: Validating the server connection. 10. If the configuration was successful the following text (Figure 76) is written on the terminal. Figure 76: Successful validation of the server connection. 11. Finish the valid bridge configuration without further adjustments (Figure 77). Figure 77: ending the valid bridge configuration. 12. Confirm the new settings to save them. If the configuration was successful, the screen should look like the following (Figure 78). Figure 78: Data Source Window of the OPC connection. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 161Cell Density Monitoring System Operating Instructions 13. Select Data Destination and choose OPC Client. 14. Copy the Class ID of the OPC Data Access (Figure 79). Figure 79: Choose the OPC Client Data Destination. 15. Start the OPC Bridge (local application => start). The settings are disabled and the programm icon (upper left corner) is green (Figure 80). Figure 80: Starting the OPC Bridge. 16. In order to test the bridge start the Softing OPC Classic Demo Client. Select the OPC Server Tab and select the DA server (Figure 81). Figure 81: Selection of the DA Server. 17. The Server is added to the left terminal and with an green status indication (Figure 82). Figure 82: Activation of the DA Server. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 163Cell Density Monitoring System Operating Instructions 18. Select the available DA items (types of information) in the tab DA Browse with right click and the selection add item (Figure 83). Figure 83: Selection of the DA Items. 19. Display the selected DA Item values in the Tab DA Items (Figure 84). Figure 84: Display of the selected DA Item values. The OPC DA bridge is now functional and communicates with the OPC server of the Control Unit. It may be connected with the favorite OPC DA client. 12.3 Connecting the 4-20 mA Output Box NOTE: Please be aware that the 4-20 mA Output Box (Ref 243850) is needed. The Analog Output Box is specially designed to connect the Control Unit to the Process Control Systems with 4-20 mA input cards. It has 4 outputs for cell density, 4 outputs for conductivity (both 4-20 mA current sources) and 4 alarms (open drain). The Analog Output Box is designed in a DIN rail for being installed in the cabinet. Refer to figure 61 to see the M12 Pin assignment. 12.3.1 Electrical Installation 1. Connect the Arc 4-20 mA Output Box to the AUX M12 connector of the Control Unit (Figure 85). VCD TCD ON/OFF Channel 1 Channel 2 ModbusAUX DC 24V ModbusAUXChannel 2Channel 1 Figure 85: Connecting the 4-20mA Analog Box with the Arc View and ComBox. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 165Cell Density Monitoring System Operating Instructions 2. Connect the outputs of the 4-20mA Output Box to the passive inputs of the process control system (process control system is a current sink) (Figure 86, Figure 87, Figure 88). • X (1/2/3/4) represents the cell density on Channel 1/2/3/4 • Sigma (1/2/3/4) represents the conductivity on Channel 1/2/3/4 • Alarm (1/2/3/4) represents the alarm of Channel 1/2/3/4 4-20 mA OUTPUTS CONNECTIONS ALARM OUTPUTS CONNECTIONS 4-20 mA OUTPUTS CONNECTIONS ALARM OUTPUTS CONNECTIONS Figure 86: Port Configuration of the 4-20 Box. +5V X / Sigma External load, max 150 Ω 4-20 mA Box Figure 87: Schematic diagram of an analog output. 3. To use the alarm output, which is an open drain output, connect an external relay to the Alarm output. Vdd Alarm X 4-20mA Box imax = 1A Figure 88: schematic diagram to connect an external relay to the Alarm output. The channel and port overview is shown in table 17. Channel Pin Description Measurement 1 1 X1 Cell Density Channel 1 2 Ground Cell Density Channel 1 3 Sigma 1 Conductivity Channel 1 4 Ground Condcutivity Channel 1 2 1 X2 Cell Density Channel 2 2 Ground Cell Density Channel 2 3 Sigma 2 Conductivity Channel 2 4 Ground Condcutivity Channel 2 3 1 X3 Cell Density Channel 3 2 Ground Cell Density Channel 3 3 Sigma 3 Conductivity Channel 3 4 Ground Condcutivity Channel 3 4 1 X4 Cell Density Channel 4 2 Ground Cell Density Channel 4 3 Sigma 4 Conductivity Channel 4 4 Ground Condcutivity Channel 4 Table 17: Channel and port overview. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 167Cell Density Monitoring System Operating Instructions 12.3.2 Software Configuration 1. Go to the Advanced settings (Figure 13) and select the 4-20 mA Configuration (Figure 85). 2. Select the Channel to configure the output range after 4-20 mA current output. Enter the cell density measurement values for 4 mA and 20 mA output (Figure 89). Figure 89: Software configuration of the 4-20 mA Box for Cell Density measurements. 3. Enter the conductivity measurement values for 4 mA and 20 mA output (Figure 90). NOTE: Conductivity measurement is just available for Incyte Sensors. Figure 90: Software Configuration of the 4-20 mA Box for conductivity. 4. Select a trigger for the alarm output. Choose from cell density, conductivity or measurement status. and define the minimal and maximal threshold, where an alarm should be triggered (Figure 91). Figure 91: Set Up of the Alarms. 12.4 Connect by Profibus Communication 12.4.1 Profibus General Information The Arc View Controller Modbus-Profibus DP Module is already pre-configured. The ID for the Card must be set correctly. The GSD file provided must be filed to the process control system for it to recognize the Module. The GSD file is available at: www.hamiltoncompany.com/ arc_view_update ATTENTION! The Arc View Controller Modbus-Profibus DP Module is only available for Arc View Controller 465 and Arc View Controller 465 XL. NOTE: Please be aware that the Modbus-Profibus module(Ref 243889) and M12 / open end cables (5 m, Ref 243851, 10 m, Ref 243852) are needed. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 169Cell Density Monitoring System Operating Instructions 12.4.1.1 Ground the Modbus-Profibus module NOTE: If the terminal 3 is not connected, the Modbus is in a floating state. Normally (with short cables)  this is not an issue but if EMI (EMC) issues are present it may cause  a problem. IIn this case terminal 3 should be connected together with the PWR supply GND to hold the bus line to a defined potential (Figure 92). Modbus line brown yellow Rx232 Tx232 AP-GND Rx422+ Rx422- Tx422+ Tx422- RS485-Termination: Rx422 = off Tx422 = on *) 1 2 3 4 5 6 7 RS485+ RS485I/O 1 I/O 3 I/O 5 I/O 7 I/O 2 I/O 4 I/O 6 I/O 8 X6 X1 Rx422*TerminationTx422* 1 3 5 7 2 4 6 8 Deutschmann Automation CL-Profibus OnOffOnOff Power supply 10...33 VDCX2 Tx-D ebug R x-D ebug 0V (Pw r) U B (Pw r) 4 3 2 1 Figure 92: Overview of the Modbus-Profibus module 12.4.2 Profibus Interface 12.4.2.1 Overview The following information is always available in the Profibus output buffer: Incyte: • Amplifier Status Incyte (word, 2 Bytes) • Conductivity (float, 4 Bytes) • Permittivity (float, 4 Bytes) • Measure Index Incyte (unsigned long, 4 Bytes) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM • Delta Epsilon (float, 4 Bytes) • Fc (float, 4 Bytes) • Alpha (float, 4 Bytes) • FScanEpsilon (float, 4 Bytes) • SpectroIndex (unsigned long, 4 Bytes) Dencytee: • Amplifier Status Dencytee (word, 2 Bytes) • Biomass (float, 4 Bytes) • Measure Index Dencytee (unsigned long, 4 Bytes) 12.4.2.2 Input Buffer The input buffer must be at least 8 byte and contain the function code and the sensor type (see page 9, Function codes) for every channel: Byte Number Type Content 0 byte Sensor type Channel 1 1 byte Function code Channel 1 2 byte Sensor type Channel 2 3 byte Function code Channel 2 4 byte Sensor type Channel 3 5 byte Function code Channel 3 6 byte Sensor type Channel 4 7 byte Function code Channel 4 Table 18: Input buffer table 12.4.2.3 Output Buffer The following table defines the content of the output buffer for each channel. The data is refreshed continously and depends on the defined sensor type (defined in the input buffer). The corresponding Modbus Registers and lengths in the Arc View Controller are defined in brackets. For data requested by a function code one byte-array per channel is reserved in the output. 171Cell Density Monitoring System Operating Instructions Byte Type Sensor type Sensor type Sensor type Number Channel 1 = 0 (none) Channel 1 = 1 Incyte Channel 1 = 2 Dencytee 0…1 word 0x0000 Amplifier Status Capa Amplifier Status OD 16bit (19/1) (711/1) 2…5 float 0.0f Conductivity (20/2) 0.0f 6…9 float 0.0f Permittivity (22/2) Biomass (714/2) 10…13 uint32 0 Measure Index Capa (24/2) Measure Index OD (716/2) 14…17 float 0.0f Delta Epsilon (42/2) 0.0f 18…21 float 0.0f Fc (44/2) 0.0f 22…25 float 0.0f Alpha (46/2) 0.0f 26…29 float 0.0f FScanEpsilon (48/2) 0.0f 30…33 uint32 0 SpectroIndex (50/2) 0 34…45 byte[] 0x00 Data requested by the function code for the channel Byte Type Sensor type Sensor type Sensor type Number Channel 2 = 0 (none) Channel 2 = 1 Incyte Channel 2 = 2 Dencytee 46…47 word 0x0000 Amplifier Status Capa Amplifier Status OD 16bit (65/1) (758/1) 48…51 float 0.0f Conductivity (66/2) 0.0f 52…55 float 0.0f Permittivity (68/2) Biomass (761/2) 56…59 uint32 0 Measure Index Capa Measure Index OD (70/2) (763/2) 60…63 float 0.0f Delta Epsilon (88/2) 0.0f 64…67 float 0.0f Fc (90/2) 0.0f 68…71 float 0.0f Alpha (92/2) 0.0f 72…75 float 0.0f FScanEpsilon (94/2) 0.0f 76…79 uint32 0 SpectroIndex (96/2) 0 80…91 byte[] 0x00 Data requested by the function code for the channel Byte Type Sensor type Sensor type Sensor type Number Channel 3 = 0 (none) Channel 3 = 1 Incyte Channel 3 = 2 Dencytee 92…93 word 0x0000 Amplifier Status Capa Amplifier Status OD 16bit (111/1) (805/1) 94…97 float 0.0f Conductivity (112/2) 0.0f 98…101 float 0.0f Permittivity (114/2) Biomass (808/2) 102…105 uint32 0 Measure Index Capa Measure Index OD (116/2) (810/2) 106…109 float 0.0f Delta Epsilon (134/2) 0.0f 110…113 float 0.0f Fc (136/2) 0.0f 114…117 float 0.0f Alpha (138/2) 0.0f 118…121 float 0.0f FScanEpsilon (140/2) 0.0f 122…125 uint32 0 SpectroIndex (142/2) 0 126…137 byte[] 0x00 Data requested by the function code for the channel Byte Type Sensor type Sensor type Sensor type Number Channel 4 = 0 (none) Channel 4 = 1 Incyte Channel 4 = 2 Dencytee 138…139 word 0x0000 Amplifier Status Capa Amplifier Status OD 16bit (157/1) (852/1) 140…143 float 0.0f Conductivity (158/2) 0.0f 144…147 float 0.0f Permittivity (160/2) Biomass (855/2) 148…151 uint32 0 Measure Index Capa Measure Index OD (162/2) (857/2) 152…155 float 0.0f Delta Epsilon (180/2) 0.0f 156…159 float 0.0f Fc (182/2) 0.0f 160…163 float 0.0f Alpha (184/2) 0.0f 164…167 float 0.0f FScanEpsilon (186/2) 0.0f 168…171 uint32 0 SpectroIndex (188/2) 0 172…183 byte[] 0x00 Data requested by the function code for the channel Byte Number Type Content 184 byte Error Table 19: Input buffer table The output data is refreshed every second. CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 173Cell Density Monitoring System Operating Instructions 12.4.2.4 Decoding the Return Values Characters According to the Modbus convention, the characters are interchanged byte-wise. The character «01234567» will be transmitted as «10325476.» Float, Integers According to the Modbus convention, the 4 byte values are interchanged word/2-byte-wise. For example: the float value 2.54 has the hex representation of 0x40228f5c. The Arc View Controller puts this value as 0x8f5c 4022 on the line. Bit Masks When an interface error to the Arc View Controller occurs, the module makes a retry. If this retry fails, the module puts an error on byte 184 and the all other fields of the output buffer are set to 0x00. Byte 184 Description 0x00 Everything ok 0x01 Arc View Controller replies with an error (Arc View Controller replies correctly, but the answer contains an error code) 0x02 Wrong answer (no Arc View Controller available, Arc View Controller sends an invalid message or no message) Table 20: Return values of the Arc View Controller The pre-amplifier status looks like this: Hex Code Description 0x00 SENSOR: not found No sensor connected to the preamplifier. 0x01 SENSOR: sterilization count overflow The sensor has been sterilized more than 100 times. Please contact your local representative. Hex Code Description 0x04 AMPLIFIER : total duration of use overflow The preamplifier has been used more than 1 year continuously. Please perform a preamplifier check 0x05 SENSOR: invalid sensor calibration The current sensor has not been calibrated with the preamplifier 0x06 SENSOR: invalid calibration The current sensor has not been calibrated with the preamplifier. The sensor serial number is not valid. 0x07 SENSOR: total duration overflow The sensor has been used more than 6 months continuously. Please perform a probe check 0x08 SENSOR: out of calibration range The measurement is out of the calibration range. 0x09 SENSOR: cleaning state The preamplifier is applying specific voltage to clean the sensor, no measurement available. 0x0a AMPLIFIER: temperature security activated The preamplifier temperature is too high, the security fuse is activated, no measurement available. 0x0b SENSOR: Out of range measurement The measurement is out of range. 0x0c SENSOR: ok The probe is working as expected. 0x0d Biomass simulator connected The calibration cap is connected to the preamplifier. 0x0e SENSOR: total sterilization time overflow The automatic sterilization counter has reached the maximum value (100) 0x0f SENSOR: overtime reached-calib/check needed The sensor has been used more than 1 year since the last calibration. Please perform a calibration. 0x10 AMPLIFIER: overtime reached/check needed The amplifier has been used more than 1 year since the last calibration. Please perform a calibration. 0x11 AMPLIFIER: not found No preamplifier is connected to the channel. 0x12 bootloader Amplifier firmware updating Table 21: Return values of the pre-amplifier CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM 175Cell Density Monitoring System Operating Instructions 9 FScan Frequency Byte 34…37: Frequency 15 (226/2) Byte 34…45: 0x00 15-17 reading Byte 38…41: Frequency 16 (228/2) Byte 42…45: Frequency 17 (230/2) 10 FScan Frequency Byte 34…37: Frequency 18 (232/2) Byte 34…45: 0x00 18-20 reading Byte 38…41: Frequency 19 (234/2) Byte 42…45: Frequency 20 (236/2) 11 FScan Frequency Byte 34…37: Frequency 21 (238/2) Byte 34…45: 0x00 21-23 reading Byte 38…41: Frequency 22 (240/2) Byte 42…45: Frequency 23 (242/2) 12 FScan Frequency Byte 34…37: Frequency 24 (244/2) Byte 34…45: 0x00 24-26 reading Byte 38…41: Frequency 25 (246/2) Byte 42…45: Frequency 26 (248/2) 13 FScan Frequency Byte 34…37: Frequency 27 (250/2) Byte 34…45: 0x00 27-29 reading Byte 38…41: Frequency 28 (252/2) Byte 42…45: Frequency 29 (254/2) 14 FScan-Values Byte 34…37: FScan 0 (256/2) Byte 34…45: 0x00 0-2 reading Byte 38…41: FScan 1 (258/2) Byte 42…45: FScan 2 (260/2) 15 FScan-Values Byte 34…37: FScan 3 (262/2) Byte 34…45: 0x00 3-5 reading Byte 38…41: FScan 4 (264/2) Byte 42…45: FScan 5 (266/2) 16 FScan-Values Byte 34…37: FScan 6 (268/2) Byte 34…45: 0x00 6-8 reading Byte 38…41: FScan 7 (270/2) Byte 42…45: FScan 8 (272/2) 17 FScan-Values Byte 34…37: FScan 9 (274/2) Byte 34…45: 0x00 9-11 reading Byte 38…41: FScan 10 (276/2) Byte 42…45: FScan 11 (278/2) 18 FScan-Values Byte 34…37: FScan 12 (280/2) Byte 34…45: 0x00 12-14 reading Byte 38…41: FScan 13 (282/2) Byte 42…45: FScan 14 (284/2) 19 FScan-Values Byte 34…37: FScan 15 (286/2) Byte 34…45: 0x00 15-17 reading Byte 38…41: FScan 16 (288/2) Byte 42…45: FScan 17 (290/2) 20 FScan-Values Byte 34…37: FScan 18 (292/2) Byte 34…45: 0x00 18-20 reading Byte 38…41: FScan 19 (294/2) Byte 42…45: FScan 20 (296/2) 21 FScan-Values Byte 34…37: FScan 21 (298/2) Byte 34…45: 0x00 21-23 reading Byte 38…41: FScan 22 (300/2) Byte 42…45: FScan 23 (302/2) 22 FScan-Values Byte 34…37: FScan 24 (304/2) Byte 34…45: 0x00 24-26 reading Byte 38…41: FScan 25 (306/2) Byte 42…45: FScan 26 (308/2) 23 FScan-Values Byte 34…37: FScan 27 (310/2) Byte 34…45: 0x00 27-29 reading Byte 38…41: FScan 28 (312/2) Byte 42…45: FScan 29 (314/2) NOTE: Some status codes are only available for Incyte pre-amplifiers. For more details about the pre-amplifier status, see the Operating Instructions Manual available at www.hamiltoncompany.com 12.4.2.5 Function codes In the following tables, the integration of each channel with the function code in the output buffer is defined. The Modbus registers and lengths are written in brackets. Table for Channel 1: CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM Code Function Sensor type Sensor type Channel 1 = 1 (Incyte) Channel 1 = 2 (Dencytee) 0 No Function Byte 34…45: 0x00 1 Serial number Byte 34…45: Gateway serial number (1/6) Arc View reading 2 Serial number Byte 34…45: Sensor serial number (7/6) Byte 34…45: Sensor serial Sensor reading number (699/6) 3 Serial number Byte 34…45: Amplifier serial number (13/6) Byte 34…45: Amplifier Preamp reading serial number (705/6) 4 FScan Frequency Byte 34…37: Frequency 0 (196/2) Byte 34…45: 0x00 0-2 reading Byte 38…41: Frequency 1 (198/2) Byte 42…45: Frequency 2 (200/2) 5 FScan Frequency Byte 34…37: Frequency 3 (202/2) Byte 34…45: 0x00 3-5 reading Byte 38…41: Frequency 4 (204/2) Byte 42…45: Frequency 5 (206/2) 6 FScan Frequency Byte 34…37: Frequency 6 (208/2) Byte 34…45: 0x00 6-8 reading Byte 38…41: Frequency 7 (210/2) Byte 42…45: Frequency 8 (212/2) 7 FScan Frequency Byte 34…37: Frequency 9 (214/2) Byte 34…45: 0x00 9-11 reading Byte 38…41: Frequency 10 (216/2) Byte 42…45: Frequency 11 (218/2) 8 FScan Frequency Byte 34…37: Frequency 12 (220/2) Byte 34…45: 0x00 12-14 reading Byte 38…41: Frequency 13 (222/2) Byte 42…45: Frequency 14 (224/2) Table 22: Function Codes Channel 1 177Cell Density Monitoring System Operating Instructions Code Function Sensor type Sensor type Channel 2 = 1 (Incyte) Channel 2 = 2 (Dencytee) 0 No Function Byte 80…91: 0x00 1 Serial number Byte 80…91: Gateway serial number (1/6) Arc View reading 2 Serial number Byte 80…91: Sensor serial number (53/6) Byte 80…91: Sensor serial Sensor reading number (746/6) 3 Serial number Byte 80…91: Amplifier serial number (59/6) Byte 80…91: Amplifier Preamp reading serial number (752/6) 4 FScan Frequency Byte 80…83: Frequency 0 (321/2) Byte 80…91: 0x00 0-2 reading Byte 84…87: Frequency 1 (323/2) Byte 88…91: Frequency 2 (325/2) 5 FScan Frequency Byte 80…83: Frequency 3 (327/2) Byte 80…91: 0x00 3-5 reading Byte 84…87: Frequency 4 (329/2) Byte 88…91: Frequency 5 (331/2) 6 FScan Frequency Byte 80…83: Frequency 6 (333/2) Byte 80…91: 0x00 6-8 reading Byte 84…87: Frequency 7 (335/2) Byte 88…91: Frequency 8 (337/2) 7 FScan Frequency Byte 80…83: Frequency 9 (339/2) Byte 80…91: 0x00 9-11 reading Byte 84…87: Frequency 10 (341/2) Byte 88…91: Frequency 11 (343/2) 8 FScan Frequency Byte 80…83: Frequency 12 (345/2) Byte 80…91: 0x00 12-14 reading Byte 84…87: Frequency 13 (347/2) Byte 88…91: Frequency 14 (349/2) 9 FScan Frequency Byte 80…83: Frequency 15 (351/2) Byte 80…91: 0x00 15-17 reading Byte 84…87: Frequency 16 (353/2) Byte 88…91: Frequency 17 (355/2) 10 FScan Frequency Byte 80…83: Frequency 18 (357/2) Byte 80…91: 0x00 18-20 reading Byte 84…87: Frequency 19 (359/2) Byte 88…91: Frequency 20 (361/2) 11 FScan Frequency Byte 80…83: Frequency 21 (363/2) Byte 80…91: 0x00 21-23 reading Byte 84…87: Frequency 22 (365/2) Byte 88…91: Frequency 23 (367/2) 12 FScan Frequency Byte 80…83: Frequency 24 (369/2) Byte 80…91: 0x00 24-26 reading Byte 84…87: Frequency 25 (371/2) Byte 88…91: Frequency 26 (373/2) 13 FScan Frequency Byte 80…83: Frequency 27 (375/2) Byte 80…91: 0x00 27-29 reading Byte 84…87: Frequency 28 (377/2) Byte 88…91: Frequency 29 (379/2) Table for Channel 2: 14 FScan-Values Byte 80…83: FScan 0 (381/2) Byte 80…91: 0x00 0-2 reading Byte 84…87: FScan 1 (383/2) Byte 88…91: FScan 2 (385/2) 15 FScan-Values Byte 80…83: FScan 3 (387/2) Byte 80…91: 0x00 3-5 reading Byte 84…87: FScan 4 (389/2) Byte 88…91: FScan 5 (391/2) 16 FScan-Values Byte 80…83: FScan 6 (393/2) Byte 80…91: 0x00 6-8 reading Byte 84…87: FScan 7 (395/2) Byte 88…91: FScan 8 (397/2) 17 FScan-Values Byte 80…83: FScan 9 (399/2) Byte 80…91: 0x00 9-11 reading Byte 84…87: FScan 10 (401/2) Byte 88…91: FScan 11 (403/2) 18 FScan-Values Byte 80…83: FScan 12 (405/2) Byte 80…91: 0x00 12-14 reading Byte 84…87: FScan 13 (407/2) Byte 88…91: FScan 14 (409/2) 19 FScan-Values Byte 80…83: FScan 15 (411/2) Byte 80…91: 0x00 15-17 reading Byte 84…87: FScan 16 (413/2) Byte 88…91: FScan 17 (415/2) 20 FScan-Values Byte 80…83: FScan 18 (417/2) Byte 80…91: 0x00 18-20 reading Byte 84…87: FScan 19 (419/2) Byte 88…91: FScan 20 (421/2) 21 FScan-Values Byte 80…83: FScan 21 (423/2) Byte 80…91: 0x00 21-23 reading Byte 84…87: FScan 22 (425/2) Byte 88…91: FScan 23 (427/2) 22 FScan-Values Byte 80…83: FScan 24 (429/2) Byte 80…91: 0x00 24-26 reading Byte 84…87: FScan 25 (431/2) Byte 88…91: FScan 26 (433/2) 23 FScan-Values Byte 80…83: FScan 27 (435/2) Byte 80…91: 0x00 27-29 reading Byte 84…87: FScan 28 (437/2) Byte 88…91: FScan 29 (439/2) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM Table 23: Function Codes Channel 2 179Cell Density Monitoring System Operating Instructions Code Function Sensor type Sensor type Channel 3 = 1 (Incyte) Channel 3 = 2 (Dencytee) 0 No Function Byte 126…137: 0x00 1 Serial number Byte 126…137: Gateway serial number (1/6) Arc View reading 2 Serial number Byte 126…137: Sensor serial number (99/6) Byte 126…137: Sensor Sensor reading serial number (793/6) 3 Serial number Byte 126…137: Amplifier serial number Byte 126…137: Amplifier Preamp reading (105/6) serial number (799/6) 4 FScan Frequency Byte 126…129: Frequency 0 (446/2) Byte 126…137: 0x00 0-2 reading Byte 130…133: Frequency 1 (448/2) Byte 134…137: Frequency 2 (450/2) 5 FScan Frequency Byte 126…129: Frequency 3 (452/2) Byte 126…137: 0x00 3-5 reading Byte 130…133: Frequency 4 (454/2) Byte 134…137: Frequency 5 (456/2) 6 FScan Frequency Byte 126…129: Frequency 6 (458/2) Byte 126…137: 0x00 6-8 reading Byte 130…133: Frequency 7 (460/2) Byte 134…137: Frequency 8 (462/2) 7 FScan Frequency Byte 126…129: Frequency 9 (464/2) Byte 126…137: 0x00 9-11 reading Byte 130…133: Frequency 10 (466/2) Byte 134…137: Frequency 11 (468/2) 8 FScan Frequency Byte 126…129: Frequency 12 (470/2) Byte 126…137: 0x00 12-14 reading Byte 130…133: Frequency 13 (472/2) Byte 134…137: Frequency 14 (474/2) 9 FScan Frequency Byte 126…129: Frequency 15 (476/2) Byte 126…137: 0x00 15-17 reading Byte 130…133: Frequency 16 (478/2) Byte 134…137: Frequency 17 (480/2) 10 FScan Frequency Byte 126…129: Frequency 18 (482/2) Byte 126…137: 0x00 18-20 reading Byte 130…133: Frequency 19 (484/2) Byte 134…137: Frequency 20 (486/2) 11 FScan Frequency Byte 126…129: Frequency 21 (488/2) Byte 126…137: 0x00 21-23 reading Byte 130…133: Frequency 22 (490/2) Byte 134…137: Frequency 23 (492/2) 12 FScan Frequency Byte 126…129: Frequency 24 (494/2) Byte 126…137: 0x00 24-26 reading Byte 130…133: Frequency 25 (496/2) Byte 134…137: Frequency 26 (498/2) 13 FScan Frequency Byte 126…129: Frequency 27 (500/2) Byte 126…137: 0x00 27-29 reading Byte 130…133: Frequency 28 (502/2) Byte 134…137: Frequency 29 (504/2) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM Table for Channel 3: 14 FScan-Values Byte 126…129: FScan 0 (506/2) Byte 126…137: 0x00 0-2 reading Byte 130…133: FScan 1 (508/2) Byte 134…137: FScan 2 (510/2) 15 FScan-Values Byte 126…129: FScan 3 (512/2) Byte 126…137: 0x00 3-5 reading Byte 130…133: FScan 4 (514/2) Byte 134…137: FScan 5 (516/2) 16 FScan-Values Byte 126…129: FScan 6 (518/2) Byte 126…137: 0x00 6-8 reading Byte 130…133: FScan 7 (520/2) Byte 134…137: FScan 8 (522/2) 17 FScan-Values Byte 126…129: FScan 9 (524/2) Byte 126…137: 0x00 9-11 reading Byte 130…133: FScan 10 (526/2) Byte 134…137: FScan 11 (528/2) 18 FScan-Values Byte 126…129: FScan 12 (530/2) Byte 126…137: 0x00 12-14 reading Byte 130…133: FScan 13 (532/2) Byte 134…137: FScan 14 (534/2) 19 FScan-Values Byte 126…129: FScan 15 (536/2) Byte 126…137: 0x00 15-17 reading Byte 130…133: FScan 16 (538/2) Byte 134…137: FScan 17 (540/2) 20 FScan-Values Byte 126…129: FScan 18 (542/2) Byte 126…137: 0x00 18-20 reading Byte 130…133: FScan 19 (544/2) Byte 134…137: FScan 20 (546/2) 21 FScan-Values Byte 126…129: FScan 21 (548/2) Byte 126…137: 0x00 21-23 reading Byte 130…133: FScan 22 (550/2) Byte 134…137: FScan 23 (552/2) 22 FScan-Values Byte 126…129: FScan 24 (554/2) Byte 126…137: 0x00 24-26 reading Byte 130…133: FScan 25 (556/2) Byte 134…137: FScan 26 (558/2) 23 FScan-Values Byte 126…129: FScan 27 (560/2) Byte 126…137: 0x00 27-29 reading Byte 130…133: FScan 28 (562/2) Byte 134…137: FScan 29 (564/2) Table 24: Function Codes Channel 3 181Cell Density Monitoring System Operating Instructions Code Function Sensor type Sensor type Channel 4 = 1 (Incyte) Channel 4 = 2 (Dencytee) 0 No Function Byte 172…183: 0x00 1 Serial number Byte 172…183: Gateway serial number (1/6) Arc View reading 2 Serial number Byte 172…183: Sensor serial number Byte 172…183: Sensor Sensor reading (145/6) serial number (840/6) 3 Serial number Byte 172…183: Amplifier serial number Byte 172…183: Amplifier Preamp reading (151/6) serial number (846/6) 4 FScan Frequency Byte 172…175: Frequency 0 (571/2) Byte 172…183: 0x00 0-2 reading Byte 176…179: Frequency 1 (573/2) Byte 180…183: Frequency 2 (575/2) 5 FScan Frequency Byte 172…175: Frequency 3 (577/2) Byte 172…183: 0x00 3-5 reading Byte 176…179: Frequency 4 (579/2) Byte 180…183: Frequency 5 (581/2) 6 FScan Frequency Byte 172…175: Frequency 6 (583/2) Byte 172…183: 0x00 6-8 reading Byte 176…179: Frequency 7 (585/2) Byte 180…183: Frequency 8 (587/2) 7 FScan Frequency Byte 172…175: Frequency 9 (589/2) Byte 172…183: 0x00 9-11 reading Byte 176…179: Frequency 10 (591/2) Byte 180…183: Frequency 11 (593/2) 8 FScan Frequency Byte 172…175: Frequency 12 (595/2) Byte 172…183: 0x00 12-14 reading Byte 176…179: Frequency 13 (597/2) Byte 180…183: Frequency 14 (599/2) 9 FScan Frequency Byte 172…175: Frequency 15 (601/2) Byte 172…183: 0x00 15-17 reading Byte 176…179: Frequency 16 (603/2) Byte 180…183: Frequency 17 (605/2) 10 FScan Frequency Byte 172…175: Frequency 18 (607/2) Byte 172…183: 0x00 18-20 reading Byte 176…179: Frequency 19 (609/2) Byte 180…183: Frequency 20 (611/2) 11 FScan Frequency Byte 172…175: Frequency 21 (613/2) Byte 172…183: 0x00 21-23 reading Byte 176…179: Frequency 22 (615/2) Byte 180…183: Frequency 23 (617/2) 12 FScan Frequency Byte 172…175: Frequency 24 (619/2) Byte 172…183: 0x00 24-26 reading Byte 176…179: Frequency 25 (621/2) Byte 180…183: Frequency 26 (623/2) 13 FScan Frequency Byte 172…175: Frequency 27 (625/2) Byte 172…183: 0x00 27-29 reading Byte 176…179: Frequency 28 (627/2) Byte 180…183: Frequency 29 (629/2) CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM CONNECTION OF THE CONTROL UNIT TO THE PROCESS CONTROL SYSTEM Table for Channel 4: 14 FScan-Values Byte 172…175: FScan 0 (631/2) Byte 172…183: 0x00 0-2 reading Byte 176…179: FScan 1 (633/2) Byte 180…183: FScan 2 (635/2) 15 FScan-Values Byte 172…175: FScan 3 (637/2) Byte 172…183: 0x00 3-5 reading Byte 176…179: FScan 4 (639/2) Byte 180…183: FScan 5 (641/2) 16 FScan-Values Byte 172…175: FScan 6 (643/2) Byte 172…183: 0x00 6-8 reading Byte 176…179: FScan 7 (645/2) Byte 180…183: FScan 8 (647/2) 17 FScan-Values Byte 172…175: FScan 9 (649/2) Byte 172…183: 0x00 9-11 reading Byte 176…179: FScan 10 (651/2) Byte 180…183: FScan 11 (653/2) 18 FScan-Values Byte 172…175: FScan 12 (655/2) Byte 172…183: 0x00 12-14 reading Byte 176…179: FScan 13 (657/2) Byte 180…183: FScan 14 (659/2) 19 FScan-Values Byte 172…175: FScan 15 (661/2) Byte 172…183: 0x00 15-17 reading Byte 176…179: FScan 16 (663/2) Byte 180…183: FScan 17 (665/2) 20 FScan-Values Byte 172…175: FScan 18 (667/2) Byte 172…183: 0x00 18-20 reading Byte 176…179: FScan 19 (669/2) Byte 180…183: FScan 20 (671/2) 21 FScan-Values Byte 172…175: FScan 21 (673/2) Byte 172…183: 0x00 21-23 reading Byte 176…179: FScan 22 (675/2) Byte 180…183: FScan 23 (677/2) 22 FScan-Values Byte 172…175: FScan 24 (679/2) Byte 172…183: 0x00 24-26 reading Byte 176…179: FScan 25 (681/2) Byte 180…183: FScan 26 (683/2) 23 FScan-Values Byte 172…175: FScan 27 (685/2) Byte 172…183: 0x00 27-29 reading Byte 176…179: FScan 28 (687/2) Byte 180…183: FScan 29 (689/2) Table 25: Function Codes Channel 4 183Cell Density Monitoring System Operating Instructions SOFTWARE VERSIONS HARDWARE COMPATIBILITY 12.4.3 Modbus Interface 12.4.3.1 Configuration The Arc View Controller Modbus Communication needs to be configured as described in Chapter 12.1. 13 Disposal The design of Hamilton Sensors optimally considers environmental compatibility. In accordance with the EC guideline 2012/19/EU Hamilton Sensors, pre-amplifier and Controller that are worn out or no longer required must be sent to a dedicated collection point for electrical and electronic devices, alternatively, must be sent to Hamilton for disposal. Sensors must not be sent to an unsorted waste disposal point. 有害物質表,請參閱www.hamiltoncompany.com, 章節過程分析,符合性聲明 DISPOSAL 14 Software Versions Version Description 3.1 Hamilton branding and basic Arc Sensor integration (MB) 3.2 Graph and Data Logging for Arc Sensors (MB) 3.3 Manual configuration of "Network Settings", allowing Cell Factor < 1.0, set time-zone offset, Graph and Data Logging for Arc Sensors (MB and BT) 3.4 Enabling Incyte (LC) low conductivity calibration, OPC connection improved, Dencytee external light detection improved 3.5 Enabling Cell Density Integration Kit, Dencytee external light detection improved 3.6 Improved error analysis and sterilization cycle counter. Dencytee external light detection adjusted 3.7 Implementation of an overall software stabilization, including: - Protection against corruption of initialization and culture files - Disabled function to download running culture files - Possibility to delete corrupt files from the Arc View Controller - Fix of export data procedure - Possibility to perform a script recovery - Implementation of an error-logger to collect status-data in case of software malfunction 3.7.1 General improvements to increase system stability based on the error-logger, which was implemented in version 3.7. Corrections of the alarm output for the 4-20 mA Output Box 3.7.2 General improvements to increase system stability. Tolerances and measured value of verification are now available in verification report 185Cell Density Monitoring System Operating Instructions ORDERING INFORMATION ORDERING INFORMATION 15 Hardware Compatibility ATTENTION! Fogale devices are only partially compatible with Hamilton products. Please do not combine Hamilton labelled Pre-Amplifier and sensors with a Control Unit having a software version below 3.0. PC Box Software (V 2.8) and licenses are not compatible with the ComBox and vice versa. Evo200 and ibiomass cannot be updated with the current Hamilton Software. 16 Ordering Information 16.1 Controller Choose Controller Choose License* 243800 Arc View 265 Code Incyte License 243801 Arc View 465 1 yes 243802 Arc View 465 XL 0 no 243810 Cell Density ComBox Code Incyte Scan License** 1 yes 0 no Code Dencytee License 1 yes 0 no Code OPC License 1 yes 0 no 2438xy Order Code * Licenses are specific to the Control Unit / ** Requires Incyte License 16.2 Cell Density Integration Kit Ref Description 243809 Cell Density Integration Kit 16.3 Units Ref Description 243710 Incyte Unit DN25 - SG 243711 Incyte Unit DN25 - DG 243712 Incyte Unit DN25 - DG BE 243700 Incyte Unit DN12 - 120 243701 Incyte Unit DN12 - 220 243702 Incyte Unit DN12 - 320 243703 Incyte Unit DN12 - 420 243704 Incyte Unit - 120 LC 243705 Incyte Unit - 220 LC 243755 Dencytee Unit - 120 243756 Dencytee Unit - 225 243757 Dencytee Unit - 325 243758 Dencytee Unit - 425 187Cell Density Monitoring System Operating Instructions ORDERING INFORMATION GLOSSARY 16.4 Spare Parts and Accessories Ref Description 242379 I-PC for CDIK 242380 CDIK ComBox 243850 4-20 mA Output Box 243851 5 m cable M12 / open end 243852 10 m cable M12 / open end 243720 Incyte Pre-Amp 243730 Incyte Sensor DN25 - SG 243731 Incyte Sensor DN25 - DG 243732 Incyte Sensor DN12 - 120 243733 Incyte Sensor DN12 - 220 243734 Incyte Sensor DN12 - 320 243735 Incyte Sensor DN12 - 420 243736 Incyte Sensor DN25 - DG BE 243716 Incyte Sensor - 120 LC 243717 Incyte Sensor - 220 LC 243760 Dencytee Pre-Amp 243750 Dencytee Sensor - 120 243751 Dencytee Sensor - 225 243752 Dencytee Sensor - 325 243753 Dencytee Sensor - 425 243870 5 m cable M12 / M12 243871 10 m cable M12 / M12 243872 20 m cable M12 / M12 243873 40 m cable M12 / M12 243820 OPC License (OPC XML-DA) 243822 Incyte License 243823 Incyte Scan License 243824 Dencytee License 243883 Arc View Controller Power Supply 243889 Arc View Controller Modbus-Profibus DP module 16.5 Calibration Tools Ref Description 243740 Incyte Val/Cal Kit Sol A, Sol B, Capacitance Simulator 243713 Incyte Val/Cal Kit LC Sol A, Sol B (LC), Capacitance Simulator 238988 Sol A for Incyte Calibration 243742 Sol B for Incyte Calibration 243708 Sol B for Incyte Calibration LC 243743 Capacitance Simulator 243886 Dencytee Val/Cal Sol 17 Glossary Name Description Unit Refers to sensor and pre-amplifier calibrated onto each other Probe Equivalent to sensor Capacitance Measurement of the viable cell density with the Incyte sensor, measurement based on an alternating electric field Optical density/ Measurement of the total cell density with the Dencytee sensor, turbidity measurement based on the turbidity of the medium Incyte Scan/FScan/Scan Multifrequency scanning during capacitance measurements Calibration Calibration of a sensor and a pre-amplifier, meaning not a correlation to biological offline measurements ComBox Refers to Cell Density ComBox CDIK Refers to Cell Density Integration Kit Table 21: Overview of relevant terms. 189Cell Density Monitoring System Operating Instructions FAQ FAQ 18 FAQ 18.1 General Questions What is the Administrator Password? The default Password is 14147125 How Should the Sensors be stored? The recommended storage conditions are at 4 to 50°C and 80 % relative humidity. The Sensors are very robust; they do not have to be stored in a specific container or liquid. How often do I need to calibrate the Unit? The Sensor comes factory-calibrated. A calibration procedure must be performed when the Sensor or the Pre-amplifier is replaced. What is the max. cable length between Unit and Arc View Controller? The distance is theoretically unlimited because the signal is digital. What is the effect of gas bubbles on the measurement? Gas bubbles have two main impacts: 1. Decrease in the signal intensity, since the Sensor is not measuring in air 2. Increase of background noise What is the effect of agitation variations on the measurement? The agitation can influence the size of air bubbles. This number and size of air bubbles can affect the Cell Density measurement Can I detect microorganism contamination with the Sensors? Yes, indirectly by the effect that the contamination has on the culture, for example a faster, or slower growths, but a direct measure of the contaminants is not possible. Do you measure % viability? No. We measure the viable cell density with Incyte and total cell density with Dencytee. You can graph both signals on one screen to analyze the deviation of viable and total cell density. What is the effect of anti-foam agents on the measurement? Antifoam has no effect on Incyte Sensors. Anti-foam may interfere with the turbidity measurement of Dencytee, as it may scatter the light. Is it possible to validate the system? Yes, the system can be validated. Hamilton can provide IQ/OQ validation packages (Ref 243999-07). The systems have already been validated for use in GMP environment with several customers. How big is the memory for data storage in the Arc View Controller? The memory of the system is 16 GB which corresponds to one year of recording with a recording period of 1 second (recording of full dataset, including spectroscopy data), or 60 years of recording for a recording period of 1 minute. Is the system software 21 CFR Part11 compliant? No Can I use the cell density signal to trigger actuator(s)? Yes. You can configure the alarm outputs of the 4-20 mA Box to control pumps. 191Cell Density Monitoring System Operating Instructions 18.2 Incyte Questions How is the correlation Cell Factor calculated for the Incyte Sensor? The system gives you the possibility to correlate the permittivity signal of the Incyte with a reference measurement such as Viable Cell Count (cell/ml), OD or Dry Cell Weight (g/l). The correlated value is referred to as the “Cell Density” value and is linked to the permittivity according to the following equation: Cell Density = (Permittivity – Offset) x cell factor. In order to correlate the permittivity signal with your reference measurements, you must determine the cell factor during the exponential phase of a culture by simultaneously taking your offline reference measurements and the permittivity readings. You can then build the linear relationship between the capacitance measurements (x axis) and your reference measurements (y axis): the slope of the linear correlation corresponds to the cell factor. The cell factor is cell line specific. Can the electric field of Incyte damage/harm the cells? No. The frequency of the electric field applied by the system is within 300 kHz and 10 MHz. These frequencies do not damage the cells. What is the effect of cell aggregates, cell clumping on the Incyte measurement? The proximity of cells can induce a slight decrease in the capacitance signal, up to 3-4%. It is also known that cell aggregates can create some noise on the signal which can be minimized with the system integration function which is based on a moving average calculation. When should I use the Incyte Scan? All calculation done during the Incyte Scan are based on a spherical model. For this reason, this scan works best is for cell culture and yeast applications. The spectroscopy module provides information related to the physiological state of the cells. It may be used to monitor key phases of processes that are typically characterized by major physiological changes in the cell population. Can I use the Incyte Sensor with micro-carriers? Yes, the Sensor is insensitive to micro-carriers. Incyte measures the bio volume of each cell on the microcarrier. FAQ FAQ How is Incyte measurement influenced if the organisms are not round but cylindrical or long stretched? The measurement is possible but calculations are based on spherical cells. That means there are 2 beta dispersion curves, one for the small dimension and one for the long dimension. The signal is the combination of those 2 curves. Are spores detected by Incyte Sensor? Spores are not detectable because they do not have a plasma membrane, and are too small. Will oil drops influence the measurement of Incyte? Oil will most likely create a surface layer on the electrodes and that will affect calibration and probably disturb the signal stability during the run. The effect at higher concentrations is not significant. What is the effect of changes in conductivity on the Incyte measurement? Changes in conductivity during a culture do not impact the capacitance measurement. The Sensor measures in a wide range of conductivity (2 to 50 mS/cm) and changes within that range do not affect the capacitance measurements. What is the effect of changes in pH on the Incyte measurement? There is no effect from pH. Can I use the Incyte Sensor in very high cell concentrations? Yes, there is no limitation in the higher end of concentrations. Can I use the Incyte Sensors for offline measurements? Yes, the Sensor may be used in stirred beaker experiments. For optimal results, it is important to make sure that there is at least 1 cm of free space around the Sensor to avoid any interactions with the electric field. The integration/moving average function may be reduced to a low integration or no integration for real-time measurements. 193Cell Density Monitoring System Operating Instructions FAQ NOTES What can I do, if a constant high offset (> 3 pF/cm) is observed? Please make clear zero of the dual measurement and scan after the process or ensure to mark zero prior to every inoculation. What is the effect of solid particles on the Incyte measurement? A content of solid particles higher than 10% (w/v) may interfere with the Incyte measurement, as the particles disturb the electrical field and may create a noise level. 195Cell Density Monitoring System Operating Instructions NOTES NOTES Hamilton Americas & Pacific Rim 4970 Energy Way Reno, Nevada 89502 USA Tel: +1-775-858-3000 Fax: +1-775-856-7259 sales@hamiltoncompany.com Hamilton Europe, Asia & Africa Via Crusch 8 CH-7402 Bonaduz, Switzerland Tel: +41-58-610-10-10 contact.pa.ch@hamilton.ch Web: www.hamiltoncompany.com USA: 800-648-5950 Europe: +41-58-610-10-10 To find a representative in your area, please visit www.hamiltoncompany.com. © 2021 Hamilton Bonaduz AG. All rights reserved. 624702/12 — 07/2021